Agilent Technologies E5071C Manual page 405

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
3. Select the Segment Sweep as the Sweep Type
The segment sweep is selected as the sweep type.
Setup Description
Sweep type: Segment
sweep
4. Execute the Calibration
In this step, a 2-port ECal module and 3-/4-port module installed in the
E5071C are used to execute calibration on the three ports used in the
measurement.
1. Connect the USB port of the 2-port ECal module and the USB port of
the E5071C with a USB cable. The connection may be made while
the unit is powered.
2. Load and execute the 3-/4-port ECal programs.
Setup Description
Opening the VBA Project
Open dialog box
Loading ECalAssistant.VBA
404
Key Operation
(Moves the focus to the softkey menu) >
Focus
Return > Sweep Type > Segment
(or
> Sweep Type > Segment)
Sweep Setup
Key Operation
Macro Setup
Select D:\Agilent\ECalAssistant.VBA and
press the Open button.
> Load Project