Agilent Technologies E5071C Manual page 533

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
signal
generator
102
Failed to
copy file
104
Failed to
create
directory
103
Failed to
delete file
600
Failed to
simulate
eye
100
Failed to
read file
601
Failed to
read bit
file
101
Failed to
write file
-
File name
760
enabled (SENS:OFFS:LOC:STAT ON and
SENS:OFFS:LOC:CONT ON).
Ensure the connection of the external signal
source.
This error occurs when copying a file
(MMEM:COPY command) fails.
This error occurs when creating a directory
(MMEM:MDIR command) fails.
This error occurs when deleting a file (MMEM:DEL
command) fails.
This error occurs when eye diagram simulation
fails.
This error occurs when a 2-port touchstone file
(CALC:FSIM:SEND:PMC:PORT:USER:FIL
command), the formatted data array
(MMEM:LOAD:FDAT command) and limit table
(MMEM:STOR:LIM command) for the active trace
on the active channel, segment sweep table
(MMEM:LOAD:SEGM command) for the active
channel, a VBA project file (MMEM:LOAD:PROG
command), etc. cannot be read normally.
This error occurs when reading bit file fails.
This error occurs when the formatted data array
(MMEM:STOR:FDATcommand) and limit table
(MMEM:STOR:LIM command) for the active trace
on the active channel, segment sweep table
(MMEM:STOR:SEGM command) for the active
channel, display image (MMEM:STOR:IMAG
command) for the LCD screen, a VBA project file
(MMEM:STOR:PROG command), etc. cannot be
written normally.
A file name error. This message appears when an
error exists in the file name and hence a

Hide quick links: