smaller stimulus values and moves the marker to first
(Search
peak encountered.
Left)
Right
Execute the search from current marker position to the
search
larger stimulus values and moves the marker to first peak
(Search
encountered.
Right)
Procedure
1. Activate the marker you are using for the peak search.
2. Press
Marker Search
3. Click Peak > Peak Excursion.
4. Enter the lower limit for the peak excursion value. This sets the peak
search to be executed based on the definitions of the newly set lower
limit for the peak excursion value and the currently set peak polarity.
5. Click Peak Polarity.
6. Select a peak polarity. This sets the peak search to be executed
based on the definitions of the currently set lower limit for the peak
excursion value and the newly set peak polarity.
7. Click the corresponding softkey to move the marker to the peak.
8. When the data format is in Smith chart or polar format, execute
the search for the main response value of the two marker
response values.
key.