channels. The channel allows up to maximum 16 traces to be defined
depending on the number of channel.
Measurement Parameters
With the E5071C, you can perform measurements of S parameter and
absolute value of the power input into ports. In addition to these, the
8753ES supports measurement parameters such as A, B, R, A/R, B/R, A/B,
etc.

Test Port Output Ranges

The 8753ES comes standard with test sets for two ports, while the E5071C
comes optionally with test sets for two ports and four ports. Furthermore,
the 8753ES comes optionally with a 75 ohm test set (Option 075), while
the E5071C does not.
For more about the measurement frequency ranges of the 8753ES and
E5071C, see the following table.

Measurement Frequency Ranges

Function
Measurement
frequency range
For more about the output power levels and output power ranges of the
8753ES and E5071C, see the following table.

Test Port Output Power Levels

Function
Output power levels
Output power ranges
8753ES
30 kHz to 3 GHz
(std.)
30 kHz to 6 GHz
(Option 006)
8753ES
- 85 dBm to 10
dBm (std.)
- 85 dBm to 8
dBm
(Options 014
and 075)
- 15 dBm to 10
dBm (std.) or -
Product Information
E5071C
See Test Port and
Frequency Options
E5071C
- 55 dBm to 10
dBm (3 GHz,
4.5 GHz, 6.5
GHz, 8.5 GHz
options)
-85 dBm to
+10 dBm (14
GHz and 20
GHz options)
None