Setup Procedure For Trigger
1. Selecting a Trigger Source
Follow the procedure below to select a trigger source.
1. Press Trigger.
2. Click Trigger Source.
3. Click the softkey that corresponds to the desired trigger source.
4. When External is selected as a trigger source, click Ext Trig Input to
select trigger polarity.
Softkey
Negative
Edge
Positive
Edge
5. The setting for trigger polarity is not valid for the external
trigger from Hander I/O.
2. Selecting a Trigger Scope
Follow the procedure below to select a trigger mode.
1. Press
Trigger
2. Click Trigger Scope.
sweep is not performed.
An "Initiate" status. When a trigger signal is
detected, a sweep is performed. After completion of
the sweep, the "Idle" status is activated.
An "Initiate" status. When a trigger signal is
detected, a sweep is performed. After completion of
the sweep, the "Initiate" status is maintained. The
sweep is repeated each time a trigger signal is
detected.
Sets all channel trigger modes to hold sweep mode
Sets trigger modes of all displayed channels
(Display
> Allocate Channels) to continuous sweep
mode
Function
Triggered at negative edge of an external trigger
input signal.
Triggered at positive edge of an external trigger
input signal.
key.
Measurement
267