Converting the Port Impedance of the Measurement Result

Overview

Converting Port Impedance

Other topics about Fixture Simulator
Overview
The measured value obtained by using a port impedance of 50 ohm can be
converted into a measured value at an arbitrary port impedance.
impedances of the two unbalanced ports equal to each other.

Port impedance conversion function

Converting Port Impedance
1. Press
Analysis
2. Specify the port reference impedance for the required port.
Select Port 1 Z0 Real, Port 2 Z0 Real, Port 3 Z0 Real, or Port 4 Z0
Real to set the port reference impedance in Real format.
When using balance conversion, be sure to set the
key, the click Fixture Simulator > Port Z Conversion.
Measurement