Obtaining Test Results

Overview
Test Results at each Measurement Point
Test Results for each Trace
Test Results for each Channel
Overall Test Results
Other topics about Limit Test
Overview
You can obtain test results by issuing a result retrieval command or
through the status register.
Test Results at each Measurement Point

Using commands that retrieve test results

You can obtain the test results at each measurement point by retrieving
the stimulus value at failed measurement points. To retrieve failed
measurement points, use the following command:
Stimulus value
Number of measurement points

Using the status register

You cannot use the status register to obtain the test results at each
measurement point.
Test Results for each Trace
Using commands that retrieve test results
You can retrieve the test result for each trace (i.e., the trace-wide result
that combines the results for all measurement points in a particular trace)
by issuing the following command:
:CALC{1-36}:LIM:FAIL?
Using the status register
The condition register and event register under the questionable limit
channel {1-16} status register provide 14 bits that correspond to traces 1
to 14 and contain the test results (0: Pass, 1: Fail) for the respective
traces; for example, you can obtain the test result for trace 1 from bit 1
and that for trace 14 from bit 14. The condition register and event register
1078
:CALC{1-36}:LIM:REP?
:CALC{1-36}:LIM:REP:POIN?

Hide quick links: