Specifying Different Standards for Each Frequency

Overview
Defining Standards for Each Subclass
Disabling Standards Defined for a Subclass
Possible Cases of Frequency Ranges Using Subclasses
Other topics about Advanced Calibration
Overview
This section demonstrates the procedure to define a different open
standard for each frequency band, based on the following information:
Standard
Label name
Open
Open 3G
Open
Open 6G
1. Press
Cal
2. Click Cal Kit.
3. Select a calibration kit you want to define.
4. Click Modify Kit > Define STDs.
5. Select #1, click Label, and enter Open 3G using the character input
pad that appears on the screen.
6. Click Open as the standard type (STD Type).
7. Set a necessary standard coefficient.
8. Enter 1G as the minimum frequency for Min Frequency.
9. Enter 3G as the maximum frequency for Max Frequency.
10.
Click Return.
11.
In the same way, select #2, press Label, and enter Open 6G
using the character input pad that appears on the screen.
12.
Select Open as the standard type (STD Type).
13.
Set a necessary standard coefficient.
14.
Enter 3G as the minimum frequency for Min Frequency.
15.
Enter 6G as the maximum frequency for Max Frequency.
16.
Click Return.
Defining Standards for Each Subclass
Frequencies to be defined
1 GHz - 3 GHz
3 GHz - 6 GHz
key.