Agilent Technologies E5071C Manual page 427

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
3. Perform a transmission calibration.
Setting Description
Select transmission calibration
Perform a Port 1-to-Port 2
calibration
Perform a Port 1-to-Port 3
calibration
Perform a Port 2-to-Port 3
calibration
4. Finish the calibration.
Setting Description
Complete the calibration and then
calculate and store calibration
coefficients.
Calibration property display: ON
4. Setting a Balance Conversion Topology
Follow the procedure below to set the balanced conversion topology.
Setting Description
Set port 1 on the DUT to unbalanced
and port 2 on the DUT to balanced.
Set the connecting destination of
port 1 on the DUT (unbalanced) to
test port 1 of the analyzer.
Set the connecting destination of
port 2 on the DUT (balanced) to test
ports 2 and 3 of the analyzer.
5. Selecting Measurement Parameters
1. Display four traces.
426
Key Operation
Return > Reflection
(With thru connection) Port 1-2
Thru
(With thru connection) Port 1-3
Thru
(With thru connection) Port 2-3
Thru
Key Operation
Return > Done (This causes
Correction to turn ON.)
Return > Return > Property (Turns
it ON.)
Key Operation
> Fixture Simulator >
Analysis
Topology > Device > SE-Bal (check
only)
Port 1 (se) > 1 (check only)
Port 2 (bal) > 2-3 (check only)