Faster Recall; Saving Instrument State - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
2. Click Channel/Trace and select the save target from all
channels/traces (All) or displayed channel/traces only (Disp Only).
If you specify the displayed channel/traces only as the save target, you
can reduce the file size. However, for channels/traces that are not
displayed, you cannot recall and reproduce the instrument state separately
held for each channel/trace at a later time.

Faster Recall

From Firmware revision A.08.0x or later, the recall speed has been
increased in the DISP only mode. In Firmware revision below A.08.0x,
preset was done for hidden channels. In Firmware revision A.08.0x or
later, hidden channels are not made preset and results in a faster recall.

Saving Instrument State

Follow the procedure below to save internal data from the E5071C.
1. Press
Save/Recall
2. Click Save State.
3. Click the softkey corresponding to the destination you want to save.
Softkey
State01 to
State08
Autorec
346
key.
Description
Save the instrument state into the state number.
Save the instrument state as the auto recall setting. The
E5071C is automatically configured with this state at the
startup.
This key saves the state into the "D:Autorec.sta".
When Autorec.sta file is found on the D: drive at startup,
the E5071C is automatically configured using the saved
settings. To disable the auto recall function, delete the

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents