E5071C
To select the subclass, use the following command.
:SENS{1-160}:CORR:COLL:CKIT:ORD
To set the start frequency of a specified subclass, use the following
command.
:SENS{1-160}:CORR:COLL:CKIT:STAN{1-30}:FMIN
To set the stop frequency of a specified subclass, use the following
command.
:SENS{1-160}:CORR:COLL:CKIT:STAN{1-30}:FMAX
Defining a Standard Class Assignment
To select the standard to be applied to the OPEN measurement for each
port, use the following command:
:SENS{1-160}:CORR:COLL:CKIT:ORD:OPEN
To select the standard to be applied to the SHORT measurement for each
port, use the following command:
:SENS{1-160}:CORR:COLL:CKIT:ORD:SHOR
To select the standard to be applied to the LOAD measurement for each
port, use the following command:
:SENS{1-160}:CORR:COLL:CKIT:ORD:LOAD
To select the standard to be applied to the THRU measurement between
each pair of ports, use the following command:
:SENS{1-160}:CORR:COLL:CKIT:ORD:THRU
To select the standard to be applied to the THRU measurement for the TRL
calibration between each pair of ports, use the following command.
:SENS{1-160}:CORR:COLL:CKIT:ORD:TRLT
To select the standard to be applied to the Reflection measurement for the
TRL calibration between each pair of ports, use the following command.
:SENS{1-160}:CORR:COLL:CKIT:ORD:TRLR
To select the standard to be applied to the Line/Match measurement for
the TRL calibration between each pair of ports, use the following command.
:SENS{1-160}:CORR:COLL:CKIT:ORD:TRLL
Setting the Standard Media Type
To select the media type of the standard to be used, use the following
command.
:SENS{1-160}:CORR:COLL:CKIT:STAN{1-30}:CHAR