Executing Zero Adjustment And Calibration Of Power Sensor; Setting Calibration Data Measurement Conditions - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
Presetting
Sweep type: Power
Fixed frequency: 1 GHz
Start value: -40 dBm
Stop value: -10 dBm
Number of points: 61

4. Executing Zero Adjustment and Calibration of Power Sensor

Execute the zero adjustment and calibration of the power sensor according
to the E4418B Power Meter User's Guide.

5. Setting Calibration Data Measurement Conditions

Follow these steps to set the calibration data measurement conditions:
Setting Description
Selecting the test port: 1
Selection of power sensor: A
Number of measurements at one
measurement point: 4
Tolerance during power calibration:
5 dB
6. Connecting Power Sensor
Connect the power sensor as shown below.
448
> OK
Preset
> Sweep Type > Power Sweep
Sweep Setup
> Power > CW Freq >
Sweep Setup
>
>
>
Start
+/-
4
0
>
>
>
Stop
+/-
1
0
> Points >
Sweep Setup
Key Operation
> Power Calibration > Select
Cal
Port > 1
> Power Calibration > Use
Cal
Sensor [ A ]
> Power Calibration > Num
Cal
of Readings >
> Power Calibration >
Cal
Tolerance >
1
>
x1
>
x1
>
>
6
1
x1
>
4
x1
>
5
x1
>
G/n