E5071C
4. Click Edit Ripple Limit to display the ripple limit table shown below.
5. Create or edit the bands by using softkeys. Note that no frequency
band is provided in the ripple limit table by default. The Edit Ripple
Limit menu is also displayed, which can be used for creating or
editing the frequency band.
6. Click Add to add a frequency band to the ripple limit table and then
specify the following parameters for the frequency band:
Parameter
Type
Begin
Stimulus
End Stimulus
Max Ripple
Acceptable range for the stimulus value: -500G to +500G. If
any outranging value is specified, it will be reset to fall within
the range.
322
Description
Selects a frequency type, either ON or OFF.
ON
Band used for the ripple test.
OFF
Band not used for the ripple test.
Specifies the start point for the stimulus value in the
ripple test.
Specifies the end point for the stimulus value in the
ripple test.
Specifies the ripple limit value.