Executing Power Calibration - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C

Executing Power Calibration

Overview
Procedure
Other Measurement Examples
Overview
This section shows an example of executing power calibration using the
E4418B power meter and the E4412A power sensor.
Power calibration may not be executed when the specified
frequency and level is out of range of the power sensor
specification.
Procedure
In this example, power calibration is executed according to the steps
shown in table below.
Step
1. Connecting Power
Meter
2. Selecting Power Meter
a. Setting Address of
GPIB Power Meter
b. Setting USB Power
Meter
3. Setting Stimulus
Condition
4. Executing Zero
Adjustment and
Calibration of Power
Sensor
5. Setting Calibration
Data Measurement
Conditions
6. Connecting Power
446
Description
Connects the power meter to the E5071C.
Configures the power meter's USB/GPIB
interface and GPIB address with the
E5071C.
Sets conditions such as the power level and
frequency.
Executes zero adjustment and calibration of
the power sensor.
Selects the port, selects the power sensor,
sets the number of measurements at one
point, and sets the tolerance during power
calibration.
Connects the power sensor.