E5071C
Response calibration (OPEN)
Response calibration (SHORT)
Response calibration (THRU)
Enhanced response calibration
1-port calibration
Full 2-port calibration
Full 3-port calibration
Full 4-port calibration
2-Port TRL calibration
3-Port TRL calibration
4-Port TRL calibration
If either an invalid calibration coefficient is specified for the
•
writing command or a nonexistent calibration coefficient is
specified for its reading command, the following error will occur:
23, Specified error term does not exist
Procedures for Writing Calibration Coefficient
You must follow the steps below to write the calibration coefficient.
1. Declare the calibration type to write.
Execute
:SENS{1-160}:CORR:COEF:METH:xxxx
2. Write any calibration coefficient.
Execute
:SENS{1-160}:CORR:COEF
calibration coefficients
3. Validate the calibration coefficients.
Execute
:SENS{1-160}:CORR:COEF:SAVE
Do not execute any other command while writing the calibration
•
coefficients.This may cause the system to function incorrectly.
If the fixture simulator is turned on, the calibration coefficient
•
writing will not function correctly. Make sure to turn off the
fixture simulator before execution.
1020
ES
ER
ED
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
command
command as needed for the written
command
EL
ET
EX
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*