Response calibration (OPEN)
Response calibration (SHORT)
Response calibration (THRU)
Enhanced response calibration
1-port calibration
Full 2-port calibration
Full 3-port calibration
Full 4-port calibration
2-Port TRL calibration
3-Port TRL calibration
4-Port TRL calibration
If either an invalid calibration coefficient is specified for the
writing command or a nonexistent calibration coefficient is
specified for its reading command, the following error will occur:
23, Specified error term does not exist
Procedures for Writing Calibration Coefficient
You must follow the steps below to write the calibration coefficient.
1. Declare the calibration type to write.
Execute
:SENS{1-160}:CORR:COEF:METH:xxxx
2. Write any calibration coefficient.
Execute
:SENS{1-160}:CORR:COEF
calibration coefficients
3. Validate the calibration coefficients.
Execute
:SENS{1-160}:CORR:COEF:SAVE
Do not execute any other command while writing the calibration
coefficients.This may cause the system to function incorrectly.
If the fixture simulator is turned on, the calibration coefficient
writing will not function correctly. Make sure to turn off the
fixture simulator before execution.