E5071C
Calibration
Overview
Performing Calibration
Defining Calibration Kits
Standard Definitions
Reading/Writing Calibration Coefficient Alone
Clearing Calibration Data and Calibration Coefficients
Other topics about Performing Calibration
Overview
You need to execute calibration to eliminate error elements related to
measurement, thus allowing you to perform accurate measurement.
Performing Calibration (Obtaining calibration coefficients)
Selecting a Calibration Kit
To select a calibration kit, use the following command:
:SENS{1-160}:CORR:COLL:CKIT
Selecting a Calibration Type
The calibration coefficients are calculated based on the selected calibration
type. Therefore, before you can calculate the calibration coefficients, you
must select the appropriate calibration type by using one of the following
commands.
Calibration type
Response
OPEN
SHORT
THRU
Enhanced Response
1-Port
Full 2-Port
Full 3-Port
Full 4-Port
1008
Command
:SENS{1-160}:CORR:COLL:METH:OPEN
:SENS{1-160}:CORR:COLL:METH:SHOR
:SENS{1-160}:CORR:COLL:METH:THRU
:SENS{1-160}:CORR:COLL:METH:ERES
:SENS{1-160}:CORR:COLL:METH:SOLT1
:SENS{1-160}:CORR:COLL:METH:SOLT2
:SENS{1-160}:CORR:COLL:METH:SOLT3
:SENS{1-160}:CORR:COLL:METH:SOLT4