3. Performing Calibration

Perform a full three-port calibration for the three ports to be used.
1. Set the type and conditions of calibration.
Setting Description
Calibration kit to use: 85033D
Type of calibration: Full three port
calibration
Test ports to calibrate: 1, 2, 3
2. Perform a reflection calibration.
Setting Description
Select reflection calibration
Perform Port 1 calibration
Perform Port 2 calibration
Perform Port 3 calibration
Key Operation
> Cal Kit > 85033D
Cal
Calibrate > 3-Port Cal
Select Ports > 1-2-3 (check
only)
Key Operation
Reflection
(With the OPEN connected) Port 1
OPEN
(With the SHORT connected) Port
1 SHORT
(With the LOAD connected) Port 1
LOAD
(With the OPEN connected) Port 2
OPEN
(With the SHORT connected) Port
2 SHORT
(With the LOAD connected) Port 2
LOAD
(With the OPEN connected) Port 3
OPEN
(With the SHORT connected) Port
3 SHORT
(With the LOAD connected) Port 3
LOAD