Agilent Technologies E5071C Manual page 217

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
on how to execute these operations, see the manual of the
power meter you are using.
When using the power sensor calibration factor table of the E5071C, set
the calibration factor to 100% and execute the calibration of the power
sensor.
1. Press
key.
Cal
2. Click Power Calibration.
3. Select a port (refer to Selecting target port of error correction).
4. Click Use Sensor. Each press toggles between channel A and channel
B. If you use a power meter with one channel, channel A is always
selected.
5. Click Num of Readings.
6. Enter the number of power level measurements at one measurement
point (averaging factor).
7. Connect the power sensor for the selected channel to the selected
port.
8. Click Take Cal Sweep to start the measurement of calibration data.
9. You can abort the measurement by pressing Abort during
measurement.
10.
When the measurement is complete, the power level error
correction is automatically turned on.
11.
If the power meter GPIB address is not set correctly or if
the power sensor is not connected to the specified channel, an
error occurs and calibration data are not measured.
If appropriate calibration data cannot be obtained, an error may
occur for each sweep after the measurement of calibration data
is complete. In this event, turn off the power level error
correction, check the connection and setting, and then measure
the calibration data again.
216