4.
If you have selected User Span, use User Span Start and User
Span Stop to set a start value and a stop value.
5.
of the frequency range is set to Freq1; a frequency point at 3/4
of the frequency range is set to Freq2.
6.
standard measurement, it does not affect the calculation result.
3. Specifying a loss value as a calculation target
Specify whether you want to include a loss value in the calculation result.
1. Press
Channel Next/Channel Prev
which you want to set auto port extension.
2. Press
Cal
3. Click Port Extensions > Auto Port Extension > Select Ports > Include
Loss to turn it on.
If the setting is not made before starting the measurement of
the OPEN/SHORT standard, it does not affect the calculation
result.
4. Specifying a DC loss value as a calculation target
Specify whether you want to include a DC loss value in the calculation
result.
For Current Span and User Span, a frequency point at 1/4
If the setting is not made before starting OPEN/SHORT
key.
keys to activate the channel for
Measurement
377