Agilent Technologies E5071C Manual page 538

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

52
No valid
memory trace
-120
Numeric data
error
-128
Numeric data
not allowed
O
No.
Message
200
Option not
installed
-225
Out of memory
P
No.
Message
Ensure the connection of USB/GPIB
interface and its setting.
This error occurs when you have executed
either DISP:WIND:TRAC:MEM ON
command to display memory trace, or any
other command to enable data calculation
using memory trace (CALC:MATH:FUNC
command with other than NORM is
specified), though no valid data exists in
memory trace. This error is not generated
by front key operations.
An error resulting from the numeric value
data (including numeric value data having
no decimal point representation) has
occurred. A numeric value error other than
Errors -121 through -129 has occurred.
An numeric-value-data element (not
violating the standard) has been received
at a position where this instrument does
not accept one.
Description
The command received has been ignored
because of the mismatch between the
contents of an option for this instrument
and the command.
It also occurs when you try to enable the
time domain function in a model not
having the time domain option. This holds
true for the frequency offset option.
This error is not generated by front key
operations.
Insufficient memory is available in this
instrument to perform the required
operation.
Description
Product Information
765

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents