Agilent Technologies E5071C Manual page 147

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
1. Press
Channel Next/Channel Prev
you want to perform the calibration.
2. Press
Cal
3. Click Calibrate > 4-Port Cal.
Reflection
1. Click Reflection.
2. Connect an OPEN calibration standard to test port 1.
3. Click Port 1 Open to start the calibration measurement.
o
4. Disconnect the OPEN calibration standard and replace it with a
SHORT calibration standard.
5. Click Port 1 Short to start the calibration measurement.
6. Disconnect the SHORT calibration standard and replace it with a
LOAD standard.
7. Click Port 1 Load to start the calibration measurement.
8. Repeat the above procedure for ports 2, 3 and 4.
9. Click Return.
Transmission
1. Click Transmission.
2. Make a THRU connection between ports 1 and 2.
3. Click Port 1-2 Thru to start the calibration measurement.
4. Repeat above procedure ports 1 and 3.
5. Repeat the procedure on ports 1 and 4.
6. Repeat the procedure on ports 2 and 3.
7. Repeat the procedure on ports 2 and 4.
8. Repeat the procedure on ports 3 and 4.
9. Click Return.
10.
If an isolation calibration must be performed using a LOAD
standard, follow the procedure below.
11.
Click Isolation (Optional).
146
key.
If you select the calibration kit which has different
calibration definitions for each gender, (m) and (f) in the
name (label) of the standard displayed in the softkey
indicate male (m) and female (f) for the analyzer's
connector, respectively.
keys to select the channel for which