Basic Calibrations
Selecting Calibration Kit
Overview
•
Procedure
•
Other topics about Basic Calibration
Overview
Before executing calibration, you need to select a calibration kit.
If you use a calibration kit other than a predefined one, you need to define
it. If the connector type of the standard of the calibration kit you use has
polarity (the distinction between male and female), you need to change the
standard class definition of the calibration kit depending on the standard
you actually use. For more information, see Modifying Calibration Kit
Definition.
If you select a predefined calibration kit, (m) and (f) in the
•
name (label) of the standard displayed in the softkey indicate
male (m) and female (f) for the analyzer's connector,
respectively.
System Z0 should be changed to 1 ohm before calibration when
•
using waveguide calibration kit and measuring waveguide
devices.
Procedure
1. Press
Channel Next/Channel Prev
you want to select the calibration kit.
2. Press
Cal
3. Click Cal Kit, then select the calibration kit.
4. If the name (label) of the calibration kit has been changed, the
label is displayed as the softkey.
5. An asterisk (*) on the upper right of the softkey corresponding
to a predefined calibration kit indicates that its definition value
has been changed from the factory setting by the user.
Key.
keys to select the channel for which
Measurement
127