When Power Sweep is selected as sweep type, Step attenuation is selected
with the start power condition and the maximum value of the Stop power
(>= Start Power) is selected with the step ATT value.
For example:
Case of Start Power= -26 dBm
Step ATT = 10 dB, Stop Power = -26 dBm ~ 0 dBm
o
Case of Start Power = -25 dBm
Step ATT = 0 dB, Stop Power = -25 dBm ~ + 10 dBm
o
If the source power range (start/stop) is located on the two ranges,
the Step ATT is set with the latest condition as:
After setting Start = -26 dBm, set stop = + 10 dBm, then start
o
is reset with - 25 dBm value.
Available power differs according to the measurement
frequency. For example, at 0 dBm Attenuator range, the
specified maximum available power level is as follows and
reduces in proportion to inserted Attenuator range:
Frequency
300 kHz to 1 MHz
1MHz to 6 GHz
6GHz to 8 GHz
8GHz to 10.5 GHz
10.5GHz to 15 GHz
15 GHz to 20 GHz
Correcting attenuation of power level (using power slope feature)
You can use the power slope feature to correct the attenuation of a power
level so that it is simply proportional to the frequency (attenuation due to
cables and so on), which improves the accuracy of the level actually
applied to the DUT.
Turning power slope feature ON/OFF
1. Press
Channel Prev/Channel Next
power slope feature will be turned OFF.
2. Press
Sweep Setup
ON/OFF setting of the power level.

Setting correction coefficient (correction amount for 1 GHz)

Power Level
+8 dBm
+10 dBm
+9 dBm
+7 dBm
+3 dBm
0 dBm
to select the channel for which the
key, then click Power > Slope and select the
Measurement
91