Agilent Technologies E5071C Manual page 379

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
1. Press
Channel Next/Channel Prev
which you want to set auto port extension.
2. Press
Cal
3. Click Port Extensions > Auto Port Extension > Adjust Mismatch to turn
it on.
If the setting is not made before starting the measurement of
the OPEN/SHORT standard, it does not affect the calculation
result.
5. Measuring the OPEN/SHORT standard and executing calculation
Calculate port extensions and loss values based on the calculation results
using the OPEN/SHORT standard.
1. Press
Channel Next/Channel Prev
which you want to want to set auto port extension.
2. Press
Cal
3. Click Port Extensions > Auto Port Extension.
4. If you use the OPEN standard, click Measure OPEN, and select the
port(s) for which you want to execute measurement. Execution is
restricted to ports selected in Selecting a port(s)
5. If you use the SHORT standard, click Measure SHORT, and select the
port(s) for which you want to execute measurement. Execution is
restricted to ports selected in Selecting a port(s).
If a port extension value or loss value has been set, the value is
updated to the calculated result.
If you execute both open measurement and short
measurement, the average of the calculation results is reflected
to the port extension and loss value.
6. Deleting the result of open/short measurement
When you exit from the softkey menu in the same level after open/short
measurement, the measurement results are deleted. Note that you can use
a GPIB command.
Port extension and loss values that have been calculated are
not cleared.
378
key.
key.
keys to activate the channel for
keys to activate the channel for