Measuring Unbalanced-Balanced Bandpass Filter - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C

Measuring Unbalanced-Balanced Bandpass Filter

Overview
Procedure
Other Measurement Examples
Overview
This section introduces an example of actually evaluating the unbalanced
and balanced SAW bandpass filter with a center frequency of 942.5 MHz.
The following figure shows the measurement circuit in the condition for
evaluating a DUT.
Procedure
Here, the DUT is evaluated by following the steps.
Step
1. Connecting the DUT
2. Setting the Measurement Conditions
3. Performing Calibration
4. Setting a Balance Conversion
Topology
422
Description
The DUT is connected.
The measurement conditions
are defined.
The full 3-port calibration is
executed.
The balance conversion
topology is specified.