Internal Data Processing; E5071C's Data Processing Flow - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C

Internal Data Processing

Data Flow
Internal Data Arrays
Other topics about Reading-Writing Measurement Data
Data Flow
The following figure provides an overview of the E5071C's internal data
processing flow.

E5071C's data processing flow

Internal Data Arrays
Raw data arrays
Raw data array contains uncorrected data. The data is obtained by
performing IF range correction, ratio calculation, port characteristics
correction and sweep averaging. This data will be used in error correction.
To read one of the raw data arrays, use the following command:
:SENS{1-160}:DATA:RAWD
Calibration Coefficient Data Arrays
A calibration coefficient data array contains the calibration coefficients
calculated based on the results of measurement performed with standard
devices.