1. Press
Channel Next/Channel Prev
which you want to set DC loss values.
2. Press
Cal
3. Click Port Extensions > Loss.
4. Click Select Port to select the port for which you want to set a DC loss
value.
5. Click Loss at DC, then enter a DC loss value.
6. If you want to set a DC loss value for other ports, repeat steps 4 to
5.

Enabling port extensions and loss values

1. Press
Channel Next/Channel Prev
which you want to enable port extension.
2. Press
Cal
3. Click Port Extensions.
4. Turn on Extensions.

Using the auto port extension function

The auto port extension function measures port extension and loss values
for each port using the OPEN/SHORT standard connected to the port,
automatically calculates them, and set them.
Softkey
Auto Port Extension
When the auto port extension function is completed, the port extensions
and loss values are updated to the calculated values.
You can use both open and short measurement values in the
auto port extension function. Note that in this case, the average
value of the calculation results is used for updating
key.
key.
Select Ports
Measure OPEN
Measure SHORT
Method
Include Loss
Adjust Mismatch
keys to activate the channel for
keys to activate the channel for
Function
Sets the Ports (1 to 4)
Performs a OPEN measurement
Performs a Short measurement
Sets the Span
Sets Loss ON/OFF
Sets Adjust Mismatch ON/OFF
Measurement
.