You can confirm the progress of error correction for each channel by
viewing the error calibration status. For the traces in which the scalar-
mixer is valid, the status [S2] is added.
In scalar-mixer calibration, the normal calibration coefficient is
invalid while the frequency-offset sweep is in progress; in this
case the scalar-mixer calibration coefficient is used instead.
Turning the frequency-offset sweep off switches over to the
normal calibration coefficient; however, the information on the
scalar-mixer calibration coefficient is retained.

Procedure Using Mechanical Calibration Kit

1. Before starting scalar-mixer calibration, verify that the frequency-
offset is activated by clicking
the frequency-offset is not valid, you cannot start scalar-mixer
calibration.
2. Press
Channel Next/Channel Prev
to calibrate.
3. Press
Cal
4. Click Mixer/Converter Calibration > Scalar Cal (Manual) > Select Ports.
5. Select the test port for starting scalar-mixer calibration.
240
key.
key > Frequency Offset. If
Sweep Setup
keys to select the channel you want