Saving/recalling instrument state for each channel into/from memory

Saving/recalling instrument state for each channel into/from
memory
The E5070A/E5071A allows you to save/recall the instrument state for each channel only.
This function lets you save the instrument state of the active channel only that is specified
independently for each channel into one of registers A to D (volatile memory) and recall
the instrument state saved in one of registers A to D to restore it as the state of the active
channel. Like when saving the entire state of the instrument into a file, you can select items
to be saved from 4 kinds (refer to Table 9-2).
Because you can call the instrument state for each channel saved with this function from a
channel different from the channel used when it was saved, you can use this function to
copy an instrument state between channels.
NOTE
Unlike when saving the entire instrument state, the instrument state for each channel is
saved into volatile memory instead of a file and therefore, if you turn off the power, it is
lost.

Operational procedure

Saving instrument state for each channel
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press one of
specified register.
NOTE
For registers having saved data, the * symbol is displayed to the right of their softkey label.
If you specify one of them, it is overwritten.
Recalling instrument state for each channel
Step 1. Press
restore.
Step 2. Press
Step 3. Press
Step 4. Press the softkey of the register in which the state you want to restore is saved. The
instrument state for the channel is recalled to the active channel.
Deleting the saved instrument states (clearing all the registers)
Step 1. Press
Step 2. Press
Step 3. Press
200
or
.
.
Save Channel
to
State A
State D
or
.
.
Recall Channel
.
.
Save Channel
. The contents of all the registers are deleted.
Clear States
to activate a channel whose state you want to save.
to save the instrument state of the active channel to the
to activate a channel whose state you want to recall and
Chapter 9