-112
Program
mnemonic too
long
Q
No.
Message
-430
Query
DEADLOCKED
-400
Query error
-410
Query
INTERRUPTED
-420
Query
UNTERMINATED
":CALC:DEV SEND1, SEND2" is sent
instead of a correct program message
with a command ":CALC:DEV SEND1"
which requires a parameter, the
instrument receives the message as the
number of parameters is invalid. See the
command reference to confirm the
required number of parameters.
The length of the header exceeds 12
characters (see 7.6.1.4.1, IEEE488.2).
Description
The state that generates a
"DEADLOCKED" Query error (see
6.3.1.7, IEEE488.2). This error occurs
when both input and output buffers
have become full, preventing the
instrument from continuing processing,
for example.
A comprehensive query error has
occurred showing that this instrument
cannot detect a more detailed error.
This code simply shows that a query
error defined in 11.5.1.1.7 and 6.3,
IEEE488.2 has occurred.
The state that generates a
"INTERRUPTED" Query error (see
6.3.2.3, IEEE488.1). This error occurs
when data bytes (DAB) or GET are
received before the transmission of the
response after a query has not been
completed, for example.
The state that generates an
"UNTERMINATED" Query error (see
6.3.2, IEEE488.2). This error occurs
when this instrument is designated as