OPEN/SHORT Response Calibration (reflection test)

Overview
Procedure
Other topics about Basic Calibration
Overview
In OPEN or SHORT response calibration, calibration data are measured by
connecting an OPEN or SHORT standard, respectively, to the desired test
port. For frequency response, these calibrations effectively eliminate the
reflection tracking error from the test setup reflection test using that port.
1-Port error model (OPEN/SHORT response)
It is also possible to carry out isolation calibration with a LOAD standard
during OPEN/SHORT response calibration. An isolation calibration will
eliminate the directivity error from the test setup in a reflection test using
that port
1-Port error model (OPEN/SHORT response + isolation)
1. System Z0 should be changed to 1 ohm before calibration when
using waveguide calibration kit and measuring waveguide
devices.
Procedure
1. Press
Channel Next/Channel Prev
you want to perform the calibration.
128
keys to select the channel for which