E5071C
11.
Click Port x@Freq x Short to start measurement of the calibration
standard.
12.
Click Port x@Freq y Short to start measurement of the calibration
standard.
13.
Disconnect the SHORT calibration standard, then connect LOAD
calibration standard in its place.
14.
Click Port x@Freq x Load to start measurement of the calibration
standard.
15.
Click Port x@Freq y Load to start measurement of the calibration
standard.
16.
Repeat the procedure for port y.
17.
Click Return.
Transmission
18.
Click Transmission.
19.
Connect a THRU between the test ports x and y (between the
connectors for the DUT).
20.
Click Port y-x@Freq y Thru to start measurement of the
calibration standard.
21.
Click Port x-y@Freq x Thru to start measurement of the
calibration standard.
22.
Click Return.
Power Meter
23.
Click Power Meter. See Preparing to control the power meter for
setup the power meter.
24.
Click Use Sensor. Each time the key is pressed, Channels A and B
switch over alternately. If you are using a one-channel power meter,
select Channel A.
25.
Connect the power sensor for the selected channel to the
selected port.
26.
Click Port x@Freq x.
27.
Click Port x@Freq y.
28.
Click Port y@Freq x.
29.
Click Port y@Freq y.
30.
Click Return.
Error Correction Done