error
-
Execution
200
error
-
Exponent
123
too large
-
Expression
178
data not
allowed
-
Expression
170
error
27
External
test set
enabled
F
No.
Message
31
Failed to
configure
ECal
module
76
Failed to
control
The trace number in
data(tr)/mem(tr)/xAxis(tr) is out of range
# of trace available depends on the
maximum number of channel/traces
The port number in Advanced math
function is out of range
# of port available depends on the model
option
An error associated with execution has been
generated for which this instrument cannot
specify the error message. This code shows that
an error associated with execution defined in
11.5.1.1.5, IEEE488.2 has occurred. This error
occurs also when a calibration measurement is
aborted.
The absolute value of the exponent exceeds
32,000 (see 7.7.2.4.1, IEEE488.2).
An expression-data element has been received at
a position where this instrument does not accept
one.
When the expression data is put to syntactic
analysis, an error not corresponding to one of
Error Numbers -171 through -179 occurs.
When the external test set mode is turned ON,
the allowable calibration is limited. See
Calibration and Measurement Parameters.
Description
This error occurs when the control of the ECal
module fails at time of executing an ECal
command, SENS:CORR:COLL:ECAL:SOLTn or
SENS:CORR:OFFS:COLL:ECAL :xxxx. The failure
results from the failure to connect the ECal
module to the USB port, failure of the ECal
module, etc.
This error occurs when the external signal source
fails to respond during measurement, though it is
Product Information
759