Warning Message

A warning message is displayed in the instrument message/Warning area
in the lower left part of the display against a gray background. Pushing a
front panel key or executing :DISP:CCL command clears the message.
This message simply appears on the display, being not known to a remote
environment such as a GPIB. This message is not displayed when another
error (against a red background) has already been displayed in the
instrument message/Warning area.
Messages
Auto port extension
not allowed
Calibration
extrapolated
Failed to set zero
type of power meter
Failed to zero and
calibrate power
meter
Fixture simulator not
allowed
Log sweep requires 2
octave minimum
Description
This message appears when you try to execute
the auto port extension when the sweep type is
power sweep or the frequency offset function is
ON.
If this message is displayed, the auto port
extension is not available.
The frequency range that can be calibrated
with the ECal module is exceeded. In this case,
the calibration data for the minimum frequency
or maximum frequency are used instead of the
data exceeding the allowable frequency range
for calibration.
This error occurs when zero type (External or
Internal) of power meter cannot be performed.
This error occurs when zeroing or calibration of
power meter cannot be performed.
This message appears when enabling the
frequency offset, while the fixture simulator
function is turned on.
If this message is displayed, the fixture
simulator function is automatically turned off.
If you change the sweep range that does not
satisfy the necessary condition of the
Product Information
777