The definitions for sweep delay time and sweep time, which you can
specify in the segment sweep, are shown in figure below.

Turning ON/OFF Segments Individually

From the firmware revision A.09.60, each segment can be turned ON or
OFF individually. The minimum number of points turned ON must be 2.
If the total number of points turned ON are less than 2, the segments with
the highest frequency ranges is automatically turned ON to meet the
minimum number of points.

Frequency/Order Base Display

You can choose between frequency-based and order-based display as the
method of displaying traces when executing a segment sweep.