Agilent Technologies E5071C Manual page 837

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
Statistical Analysis
The E5071C provides a mechanism that analyzes trace statistics. You can
analyze the following statistics:
Average
Standard deviation
Difference between the maximum and minimum values
To retrieve the results of statistical analysis, use the following command:
:CALC{1-36}:MST:DATA?
Also, you can use the following command to control whether to display the
results of statistical analysis on the LCD:
:CALC{1-36}:MST
You can retrieve the results of statistical analysis even after you
have turned off the display showing these results.
Alternatively, you can use the analysis commands to analyze the trace
statistics. When you use the analysis commands, you can analyze the trace
data in a specific range as well as the trace data throughout the entire
sweep range. For information on how to use the analysis commands, refer
to Analysis Commands.
Other topics about Analyzing Data