You can save the definition. See Saving Definition File of Calibration Kit.
Defining Calibration Kit Gender (m/f)
The gender of a calibration can be set using the subclass function in the
ENA. For example, subclass 1 can be used as "male" and subclass 2 can be
used as "female" or vice-versa. The display label, for example "Open (f)"
or "Open (m)", is just a display label and can be edited using
Kit > Define STDs and hence cannot be used to change the gender of the
calibration kit. To change the gender of the calibration kit, the E5071C
subclass function must be used.
For example, the gender of the default calibration kit (85032F) can be
checked using the following procedure:
1. Press
Cal
2. Click Cal Kit and select 85032F.
3. Select
Cal
The subclass gender is set from the E5071C side and not from the
Calibration kit side.
If both male/female are set using subclass in advance, the gender which is
used for the calibration can also be changed programmatically using
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SUBClass
The gender setting option using subclass is defined from the
E5071C connector type and not from the standard calibration kit
gender type.
Saving Definition File of Calibration Kit
You can save the definition file of the calibration kit for each standard into
a file on a storage device, then recall it later to reproduce the definition.
1. Press
Cal
2. Click Cal Kit, then select a calibration kit.
3. Click Modify Cal Kit > Export Cal Kit... to open the dialog box.
4. Specify a folder, enter a file name, and click Save.
The file is saved with the (.ckx) extension. Do not open and
modify it directly.
Never modify the contents (folders and files) of drives except
for drive D. Modifying the contents of drives other than drive D
may give serous damage to the functions and performance of
the analyzer.
Loading Definition File of Calibration Kit
1. Press
Cal
2. Click Cal Kit, then select a calibration kit.
key.
> Modify Cal Kit > Specify CLSs > Sub Class.
key.
key.
Measurement
> Modify
Cal
195