1. Calibration Kit: Select the calibration kit or ECal module that is being
used.
2. Verification Kit: The corresponding Verification Kit to use appears in
the Verification Kit box. Refer to
and Verification Kit
3. Test Cable(s): Select the method of connecting ports and
Calibration/Verification Kit of E5071C. It is necessary to connect
Calibration Kit and Verification Kit to E5071C with the same cable
connection.
When calibration or verification is measured (2 port measurement),
E5071C ports are connected to ports of kit by the following options:
Single Cable: One side of the port is connected to the port of
E5071C and kit directly. The other port is connected through
Cable.
Cable Pair: Both ports are connected to both ports of E5071C
and kit through Cable.
Cable Flex Factor: Flex Factor determines how much of the
cable phase uncertainty should be included in determining the
limit lines.
the combination of Calibration Kit
table.