no command is executed before the preceding command
completes itself.
Turning ON/OFF Power-level Error Correction
To turn on or off the power-level error correction, use the following
command:
:SOUR{1-36}:POW:PORT{1-4}:CORR
When the power calibration data measurement initiated with the
36}:POW:PORT{1-4}:CORR:COLL
is automatically turned on.
Reading/writing power calibration data array
The power calibration data array contains data used to perform error
correction for the power level at each measurement point (values obtained
by subtracting the value actually measured with the power meter from the
set power level value when measuring power calibration data at each
measurement point).
The number of power calibration data arrays that are assigned to individual
ports of individual channels can be up to 64 (16x4). To read/write one of
the power calibration data arrays, use the following command:
:SOUR{1-36}:POW:PORT{1-4}:CORR:DATA
command is complete, the power calibration
Programming
:SOUR{1-
1025