Possible cases of frequency ranges
Example 1, When frequency ranges specified with
subclasses overlap
Example 2, When frequency ranges specified with
subclasses do not cover a part of a measurement
frequency range
Example 3, When a frequency range specified with
a subclass lies outside the measurement frequency
range
Example 4, When a part of a frequency range
specified with a subclass lies outside the
measurement frequency range
Example 1, When frequency ranges specified with subclasses overlap
For the measurement frequency range (1 GHz - 8 GHz), if you execute
calibration with a standard of subclass 1 (1 GHz-5 GHz) and then with a
standard of subclass 2 (4 GHz-8 GHz), the standard last executed is
applied for the overlapping portion (4 GHz-5 GHz).
Example 2, When frequency ranges specified with subclasses do not cover a part of a measurement frequency range
For a measurement frequency range (1 GHz to 8 GHz), if you execute
calibration with a standard of subclass 1 (1 GHz to 5 GHz) and then with a
standard of subclass 2 (6 GHz to 8 GHz), Done is NOT available for the
undefined portion (5 GHz-6 GHz).
Measurement
Execution of
calibration
Possible
Impossible
Possible
Possible
199