Agilent E5052A Signal Source Analyzer Programmer’s Guide Fourth Edition FIRMWARE REVISIONS This manual applies directly to instruments that have the firmware revision A.02.00. For additional information about firmware revisions, see Appendix A. Agilent Part No. E5052-90031 August 2005 Printed in Japan...
Page 2
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Agilent Technologies. Microsoft®,MS-DOS®,Windows®,Visual C++®,Visual Basic®,VBA® and Excel® are registered UNIX is a registered trademark in U.S.
Page 3
• User’s Guide (Part Number E5052-900x0, attached to Option ABA) This manual describes most of the basic information needed to use the E5052A. It provides a function overview, detailed operation procedure for each function (from preparation for measurement to analysis of measurement results), specifications, and supplemental information.
Making Effective Use of This Manual This chapter describes the contents of this guide.Using this chapter with the table of contents helps you to retrieve description of a subject you wish to understand as well as to obtain an overview of this guide.Also see the latter part of this chapter for brief description of usage of this guide, focusing on searching commands.
This guide describes programming method mainly aiming at learning how to write a program that remotely controls the E5052A using SCPI commands, focusing on sample usage with the HTBasic. Controlling the E5052A using the built-in VBA is not covered by this guide; it is described VBA Programmer's Guide VBA Programmer's .For using the E5052A VBA, see...
Page 25
Agilent E5052A than that indicated by the current printing date of this manual. The information in this manual applies directly to the E5052A model that has the serial number prefix listed on the title page of this manual. Appendix B, “Status Reporting System.”...
NOTE Some SCPI commands supported by the E5052A have optional syntax elements. In the command reference conventions, these elements are enclosed between square brackets ([]) or printed in lowercase letters. See “Syntax” on page 92 for more information.
Page 27
Making Effective Use of This Manual How to Use This Manual • : Spectrum monitor measurement • : Transient measurement • USER : User-defined window NOTE Some SCPI commands preclude the availability of certain measuring modes.
Page 28
Making Effective Use of This Manual How to Use This Manual...
Depending on the system controller and the interface, you can configure four types of remote control systems as shown in the table below. System controller Interface Overview System to control the E5052A and other devices connected via GPIB from the external GPIB controller. For more information, refer to “GPIB Remote Control System”...
Configuration of GPIB remote control system NOTE When the power of the E5052A is turned off, the SRQ signal of the built-in GPIB remains active. With the power turned off, disconnect the GPIB remote control system and the E5052A to prevent malfunction.
GPIB (for example, HTBasic or Agilent VEE). 3. Other devices (other instruments and/or specific-purpose peripherals) 4. GPIB cables for connecting the E5052A, the external controller, and other devices Size of system you can construct •...
The LAN (Local Area Network) remote control system provides two methods for controlling the E5052A: using the SICL-LAN server and using the telnet server. System configuration Use LAN cables to connect the E5052A and the external controller (computer). Figure 2-2 shows the configuration of the LAN remote control system. Figure 2-2...
Step 1. Turn on the SICL-LAN server of the E5052A. [System] - Misc Setup - Network Setup - SICL-LAN Server [ON] Step 2. Set the GPIB address of the E5052A for control with the SICL-LAN server. “XX” indicates the address number.
Page 35
Agilent VEE allows you to control the E5052A via the I/O interface. The following example shows how to control the E5052A when it is set as follows: the address of the SICL-LAN server is 17 and the IP address is 1.10.100.50.
Page 36
This host . The Select or Enter Gateway setting screen appears. Enter the IP address or host name of the E5052A (4 in figure). Figure 2-3 shows an example of control using the I/O interface set by the above procedure.
E5052A. A socket is an endpoint for a network connection; on the E5052A, ports 5024 and 5025 are provided for the sockets. Port 5024 is used for conversational control using telnet (user interface program for TELNET protocol) and port 5025 is used for control from a program.
Page 38
Step 2. At the MS-DOS prompt, type telnet 192.168.0.1 5024 and press the return key. Step 3. The telnet screen opens. Step 4. Type a command and press the return key; this command is sent to the E5052A and executed. If you enter a command that queries some data, the query response is displayed below the line you have entered the command.
Page 39
E5052A is 192.168.0.1). Enter 5025 in the field to specify the port for connection (1 in Figure 2-5) and enter the IP address or host name of the E5052A in the field to specify the host name (2 in Figure 2-5).
The USB remote control system controls instruments that use the name “alias.” There is no such address for GPIB connections. Use a LAN cable to connect the E5052A to an external controller (personal computer). Figure 2-6 shows an overview of the system configuration for the USB remote control system.
Page 41
Type B: USB (USBTMC) interface port Preparing E5052A You do not have to configure any softkey or command of the E5052A in order to control the E5052A from an external controller. Simply connect a USB cable to the USB interface port.
Page 42
Overview of Remote Control USB Remote Control System 1. Registering Alias When USB Cable Is Connected When new device is connected via USB cable, the following dialog box will appear automatically.Follow the steps below to register alias, when you use the Agilent I/O Libraries Suite 14.
Page 43
Changing Alias Control using C or Visual Basic You can control the E5052A by programming using Visual C++ or Visual Basic in the Windows environment as well as SICL/VISA.For further information on controlling the E5052A, see the manual of SICL or VISA.For Agilent I/O Libraries, use Agilent I/O Libraries M.01.01 or Agilent I/O Libraries Suite 14.
Page 44
Overview of Remote Control USB Remote Control System Control using Agilent VEE Agilent VEE allows you to control the E5052A via the direct I/O interface.The following example shows how to control the E5052A to which alias is given as SSA_USBIF. NOTE When using Agilent VEE for PC, use Agilent VEE Pro 7 for Windows or later version.
Page 45
Overview of Remote Control USB Remote Control System Figure 2-9 Sample Control Using Agilent VEE (USB)
Sending SCPI command messages Sending SCPI command messages Types and structure of commands The SCPI commands available for the E5052A are classified into two groups as follows. E5052A commands E5052A Commands specific to the E5052A. These cover all measurement functions of the E5052A as well as some general-purpose functions.
Overview of Remote Control Sending SCPI command messages Figure 2-10 Using colons and semicolons in command tree Grammar of messages This section describes the grammar of program messages sent via GPIB. Program messages are messages that the user sends to the instrument from the external controller to control the instrument.
Overview of Remote Control Sending SCPI command messages Remote mode The E5052A is not equipped for remote mode. Therefore, even if you send a GPIB command, it cannot enter into remote mode automatically. There is no local key to release remote mode.
Making a Measurement This chapter describes how to trigger the instrument to start a new measurement cycle and how to detect the end of a measurement cycle.
NOTE The E5052A has a trigger model for each measurement (see Figure 3-1); however, only one triggered measurement can take place at a time. This is because the hardware and the trigger port are shared by four instruments.
Page 51
Making a Measurement Trigger System Figure 3-1 Trigger system...
Making a Measurement Trigger System Trigger state and transition for each measurement The following subsections describe the states for each measurement and how the trigger system switches among the states. “Hold” State When one of the following commands has been executed, the trigger system switches to “Hold”...
Page 53
Making a Measurement Trigger System *1.This function is effective only for the transient measurement. This trigger source cannot be selected for other measurements. It is not displayed on the softkey either. To set the trigger source, use the following command (“xx” specifies the measurements (PN,SP,FP,TR) that are now waiting for trigger): •...
Making a Measurement Staring a Measurement Cycle (triggering the instrument) Staring a Measurement Cycle (triggering the instrument) Configuring the Instrument to Automatically Perform Continuous Measurement Step 1. Select a measurement to be triggered using the :TRIGger:MODE command on page 372. Step 2.
• Follow these steps: Step 1. Configure the E5052A so that it generates an SRQ when the operation status condition register's bit 4 (a bit that is set to 1 during measurement) is changed from 1 to 0. Step 2. Trigger the instrument to start a measurement cycle.
Page 56
Example 3-1 is a sample program that demonstrates how to use an SRQ to interrupt the end of measurement. This program triggers the spectrum measurement of E5052A and then terminates the program when the end signal is received. Example 3-1...
Making a Measurement Waiting for End of measurement PRINT "Measurement Complete" Using Wait Time Before creating your program, actually measure the time between the start and end of the measurement cycle. Then code your program so that the controller waits for the actually measured time by using the appropriate command (for example, the WAIT command for HTBasic).
Page 58
Making a Measurement Waiting for End of measurement...
Reading/Writing Measurement Data This chapter provides an overview of the Agilent E5052A's internal data transfer format and explains how to read and write measurement results (internal data array and marker position).
Data Types for Data Transfer Data Types for Data Transfer This section describes the data types that the E5052A uses for data transfer. Array type If you use any of the following commands, the data transfer will be executed using array-type data.
Reading/Writing Measurement Data Data Types for Data Transfer For transient measurement • :CALC:TR[1-1]:NARR:DATA:RDAT • :CALC:TR[1-1]:NARR:DATA:XDAT? • :CALC:TR[1-1]:TRAC[1-4]:DATA:FDAT • :CALC:TR[1-1]:TRAC[1-4]:DATA:FMEM • :CALC:TR[1-1]:TRAC[1-4]:DATA:UDAT • :CALC:TR[1-1]:TRAC[1-4]:DATA:UMEM • :CALC:TR[1-1]:WIDE:DATA:RDAT • :CALC:TR[1-1]:WIDE:DATA:XDAT? For sending and receiving data to/from VBA program • :PROG:VAR:ARR[1-10]:DATA Block type If you use any of the following commands, the data transfer will be executed using block-type data.
Reading/Writing Measurement Data Data Transfer Format Data Transfer Format Set the data transfer format to transfer data of either array or block type. To set the data transfer format, use the following command: :FORMat:DATA on page 272 • When you select the ASCII transfer format as the data transfer format, you can also specify the byte order.
Reading/Writing Measurement Data Data Transfer Format ASCII Transfer Format When you select the ASCII transfer format as the data transfer format, numbers are transferred as ASCII bytes, each of which corresponds to one of the formats shown below. Note that numbers are separated from one another with a comma (,) in accordance with the IEEE 488.2 specification.
Reading/Writing Measurement Data Data Transfer Format Binary Transfer Format You can select the binary transfer format from the IEEE 64-bit floating point format or the IEEE 32-bit floating point format depending on the controller you use. IEEE 64-bit floating point format When you select the IEEE 64-bit floating point binary transfer format as the data transfer format, numbers are transferred in the format shown in Figure 4-3.
Page 65
Reading/Writing Measurement Data Data Transfer Format Figure 4-5 IEEE 32-bit floating point binary transfer format This data transfer format uses a header that consists of a sharp character (#), the number 6 (which indicates the byte size of the <number of bytes transferred> part), and the <number of bytes transferred>...
Internal data processing Internal data processing Data flow Figure 4-7, Figure 4-9, Figure 4-10 and Figure 4-11 provide overviews of the E5052A’s internal data processing flows. For more information on each data processing flow, refer to User's Guide the “...
Page 67
Reading/Writing Measurement Data Internal data processing Figure 4-9 Data processing flow for frequency/power measurement with E5052A...
Page 68
Reading/Writing Measurement Data Internal data processing Figure 4-10 Data processing flow for transient measurement with E5052A...
Page 69
Internal data processing Figure 4-11 Data processing flow for user-defined function measurement with E5052A The enclosed part of the data processing flow represents the data array. By using SCPI commands, the data array can be read for processing and written for display on the...
Reading/Writing Measurement Data Internal data processing Internal data arrays Unformatted data arrays An unformatted array contains the raw data just as it was obtained through measurement and divided for each trace. The instrument retains the unformatted data arrays which are associated with each trace of individual measurements.
Page 71
Reading/Writing Measurement Data Internal data processing X-axis Data Arrays An x-axis data array contains the x-axis values for all measurement points. The instrument retains the x-axis arrays associated with individual measurements (for the transient measurement, narrow/wide only). X-axis data arrays are read-only. To retrieve one of the x-axis data arrays, use the following command: •...
Reading/Writing Measurement Data Retrieving Measurement Results Retrieving Measurement Results Markers allow you to retrieve measurement results at your specified points only. “Internal data arrays” on page 70 allow you to retrieve all measurement results throughout a particular trace. Retrieving measurement results at marker positions In using markers to retrieve measurement results at specified points, you can use up to six markers for each measurement graph, and you can move them to any point on the trace.
Page 73
Reading/Writing Measurement Data Retrieving Measurement Results formatted data arrays. The sample program in Example 4-1 uses the ASCII transfer format while the sample in Example 4-2 uses the binary transfer format. These sample programs show how to retrieve the formatted data arrays of measured phase noise.
Reading/Writing Measurement Data Entering Data in a Trace Entering Data in a Trace By using “Formatted data array” on page 70, you can change the data/memory trace on the LCD by writing the new data into the “Formatted memory arrays” on page 70. When you write data into the formatted data/memory array, you can choose either the ASCII or binary transfer format (see “Data Types for Data Transfer”...
Page 75
Reading/Writing Measurement Data Entering Data in a Trace ASSIGN @File TO * OUTPUT @Agte5052;":FORM:DATA REAL64" Head$="#6"&IVAL$(8*Nop,10) OUTPUT @Agte5052;":CALC:PN1:TRAC1:DATA:FDAT ";Head$; OUTPUT @Binary;Fdata(*),END...
Page 76
Reading/Writing Measurement Data Entering Data in a Trace...
Saving and Recalling (File Management) This chapter explains how to save and recall instrument status and measurement results onto/from the files. The chapter also discusses file management in general.
Auto recall The file saved with the name F:\autorec.sta or A:\autorec.sta will be automatically recalled when the E5052A is powered on. Saving measurement data Measurement data (in a formatted data array) can be saved in the file in CSV (Comma Separated Value) format.
Saving and Recalling (File Management) Saving/Recalling Saving images on LCD screen Images displayed on the LCD screen can be saved in a file in the bitmap (.bmp) or portable network graphics (.png) format. To save a screen image in a file, use the following command: :MMEMory:STORe:IMAGe on page 286 •...
• Also, file transfer from the E5052A to the external controller can be done by reading data from a file on the E5052A while using the commands as query and then writing them to the file on the controller. Retrieving data from storage To retrieve information from the E5052A’s built-in storage (usage, properties of file...
Example 5-1 shows a sample program for transferring files between the external controller and the E5052A. This program reads out data from a specified file on the E5052A and then writes them to a specified file on the external controller.
Working with Automatic Test Systems This chapter describes useful features that are available when the Agilent E5052A is integrated with an automatic test system.
Working with Automatic Test Systems Preventing Erroneous Key Operation on the Front Panel (Key Lock feature) Preventing Erroneous Key Operation on the Front Panel (Key Lock feature) When no operation is required from the front panel controls, the mouse, or the keyboard, disabling these input devices can prevent erroneous operation that might occur due to accidental touching.
SCPI commands should be processed quickly to improve throughput when the commands are frequently executed (for example, reading out the trace for each measurement). With the E5052A, the processing time for SCPI commands can be improved by decreasing the refresh rate of the LCD display.
*ESE on page 273 Follow these steps: Step 1. Set the E5052A so that it generates an SRQ when any of the error occurrence bits is set to 1 in the standard event status register. Step 2. When an SRQ is generated, the program interrupts the measurement cycle.
Page 87
Working with Automatic Test Systems Detecting Occurrence of an Error The error queue can be used in the following ways: 1. It is used as a branch for error handling. When an error queue is retrieved, it returns 0 as the error number and &dlq;No error&drq;...
Working with Automatic Test Systems Limit Test Limit Test This section describes how to define the limit lines and determine pass or fail with regard to the limit test function. For more on the concept of the limit test, refer to the Chapter describing “Data Analysis and Result Output”...
SCPI Command Reference This chapter describes the SCPI command reference for the Agilent E5052A. It describes the commands using their abbreviated format in alphabetical order. If you want to look up commands using their fully qualified format, refer to the index for the desired SCPI command.
Part with heading “Syntax” describes the syntax to send a command from the external controller to the E5052A. A syntax consists of a command part and a parameter part. The separator between the command part and the parameter part is a space.
SCPI Command Reference Parameters Parameters Part with heading “Parameters” describes necessary parameters when sending the command. When a parameter is a value type or a string type enclosed with <>, its description, allowable setup range, preset (factory-set) value, and so on are given; when a parameter is a selection type enclosed with {}, the description of each selection item is given.
Equivalent key Device Configuration Using E5052A and E5053A Microwave Downconverter There are 4 types of device configurations using the E5052A and the E5053A Microwave Downconverter including the stand-alone use of the E5052A: A-1, A-2, B, and C(Figure 7-1) Figure 7-1 Device Configurations Using E5052A and E5053A The table below lists the configuration ID used in this chapter with its description as well.
Page 95
SCPI Command Reference Equivalent key Configuration Description With downconverter turned on and the RF input set to ’E5052A Direct’ With downconverter turned on and the RF input set to ’Downconverter’...
SCPI Command Reference :ABORt E5052A commands This section describes the commands specific to the E5052A. :ABORt Syntax :ABORt Description Abort measurement (No Query) Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:ALLTrace:ACTive Syntax :CALCulate:FP[1-1]:ALLTrace:ACTive <numeric> :CALCulate:FP[1-1]:ALLTrace:ACTive? Description...
SCPI Command Reference :CALCulate:FP[1-1]:ALLTrace:LIMit:FAIL Description OFF or 0(Preset Set bandmarker coupling function to 'OFF' value) Equivalent key FP Menu -> Marker Function -> Couple FP Menu -> Marker Search -> Couple :CALCulate:FP[1-1]:ALLTrace:LIMit:FAIL Syntax :CALCulate:FP[1-1]:ALLTrace:LIMit:FAIL Description Reads out the limit test result (Query Only) Parameter Description ON or 1...
SCPI Command Reference :CALCulate:FP[1-1]:ALLTrace:MARKer:REFerence:NUMBer Description Turns on/off marker discrete function Parameter Description ON or 1 Enable marker discrete function OFF or 0(Preset Disable marker discrete function value) Equivalent key FP Menu -> Marker -> More Functions -> Discrete :CALCulate:FP[1-1]:ALLTrace:MARKer:REFerence:NUM Syntax :CALCulate:FP[1-1]:ALLTrace:MARKer:REFerence:NUMBer <numeric>...
SCPI Command Reference :CALCulate:FP[1-1]:DATA:RDATa Description OFF or 0(Preset Set delta marker mode mode to 'OFF' value) Equivalent key FP Menu -> Marker -> More Functions -> Ref Marker Mode :CALCulate:FP[1-1]:DATA:RDATa Syntax :CALCulate:FP[1-1]:DATA:RDATa <array> :CALCulate:FP[1-1]:DATA:RDATa? Description Sets/Reads raw data Parameter <Description> Range 1...3003 Preset value...
SCPI Command Reference :CALCulate:FP[1-1]:DATA:XDATa :CALCulate:FP[1-1]:DATA:XDATa Syntax :CALCulate:FP[1-1]:DATA:XDATa Description Reads X-axis data (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:ALLMarker:ACTive Syntax :CALCulate:FP[1-1]:TRACe[1-4]:ALLMarker:ACTive <numeric> :CALCulate:FP[1-1]:TRACe[1-4]:ALLMarker:ACTive? Description Selects active marker Parameter <Numeric> Range 1 to 6 Preset value Unit Resolution Equivalent key...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:X:SPAN :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:X:SPAN Syntax :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:X:SPAN <numeric> :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:X:SPAN? Description Sets/Reads the span value of bandmarker X Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key FP Menu -> Marker Function -> Band Marker X -> Span FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:X:STOP Parameter Description ON or 1 Set bandmarker X mode to 'ON' OFF or 0(Preset Set bandmarker X mode to 'OFF' value) Equivalent key FP Menu -> Marker Function -> Band Marker X -> Band Marker X FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:SPAN <Numeric> Resolution Equivalent key FP Menu -> Marker Function -> Band Marker Y -> Center FP Menu -> Marker Search -> Band Marker Y -> Center :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:SPAN Syntax :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:SPAN <numeric> :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:SPAN? Description Sets/Reads the span value of bandmarker Y Parameter <Numeric>...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:STATe FP Menu -> Marker Search -> Band Marker Y -> Start :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:STATe Syntax :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:STATe {ON|OFF|1|0} :CALCulate:FP[1-1]:TRACe[1-4]:BDMarker:Y:STATe? Description Turns on/off bandmarker Y Parameter Description ON or 1 Set bandmarker Y mode to 'ON' OFF or 0(Preset Set bandmarker Y mode to 'OFF' value) Equivalent key FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:DATA:FDATa Parameter <Numeric> Range 1 to 8 Preset value Unit Resolution Equivalent key FP Menu -> Trace View-> Copy to USER -> Copy to USER1 FP Menu -> Trace View-> Copy to USER -> Copy to USER2 FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:DATA:UDATa Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:DATA:UDATa Syntax :CALCulate:FP[1-1]:TRACe[1-4]:DATA:UDATa <array> :CALCulate:FP[1-1]:TRACe[1-4]:DATA:UDATa? Description Sets/Reads unformatted trace data Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:FORMat:FREQuency <Description> Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:FORMat:FREQuency Syntax :CALCulate:FP[1-1]:TRACe[1-4]:FORMat:FREQuency {HZ|HZV|DHZ|PCT|PPM} :CALCulate:FP[1-1]:TRACe[1-4]:FORMat:FREQuency? Description FP-frequency format Parameter Description HZ(Preset value) Set FP-frequency format to 'Hz' Set FP-frequency format to 'Hz/V' Set FP-frequency format to 'ΔHz' Set FP-frequency format to '%' Set FP-frequency format to 'ppm' This command is available when trace 1 or trace 4.
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y Syntax :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y {FRANge|BDMarker} :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y? Description Sets/Reads analysis/search range (Y-axis) Parameter Description FRANge Set analysis/search range (Y-axis) to 'Full Range' BDMarker(Preset Set analysis/search range (Y-axis) to 'Band Marker' value) Equivalent key FP Menu -> Marker Function -> Analysis Range (Y) :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:STATistics:DA Syntax :CALCulate:FP[1-1]:TRACe[1-4]:FUNCtion:STATistics:DATA...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:HOLD Description STATistics Set analysis type to 'Statistics' Equivalent key FP Menu -> Marker Function -> Analysis Type :CALCulate:FP[1-1]:TRACe[1-4]:HOLD Syntax :CALCulate:FP[1-1]:TRACe[1-4]:HOLD {OFF|MAXimum|MINimum} :CALCulate:FP[1-1]:TRACe[1-4]:HOLD? Description Selects/Reads trace data hold type Parameter Description OFF(Preset value) Set data hold to 'Off' MAXimum Set data hold to 'Maximum' MINimum...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:CLEar Description Sets/Reads the lower limit values of all measurement points Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent: CLEar Syntax :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:CLEar Description Clears the lower limit line (No Query) Equivalent key FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:DATA? Description Sets/Reads segment data of the lower limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] Syntax :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] Description Reads the limit test results of all measurement points in selected traces (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent:CLEar Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent: CLEar Syntax :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent:CLEar Description Clears the upper limit line (No Query) Equivalent key FP Menu -> Display-> Limit Test -> Delete Upper Limit Line :CALCulate:FP[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent: COUNt Syntax...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch:EXECute:LPEak Description Sets/Reads segment data of the upper limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch:E XECute:LPEak Syntax :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch:EXECute:LPEak Description Executes marker peak search left (No Query) Equivalent key FP Menu ->...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:STATe Description TARGet Set the marker tracking type to 'Target' Equivalent key FP Menu -> Marker Search -> Tracking :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:STATe Syntax :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:STATe {ON|OFF|1|0} :CALCulate:FP[1-1]:TRACe[1-4]:MARKer[1-6]:STATe? Description Turns on/off markers Parameter Description ON or 1 Set markers mode 'ON' OFF or 0(Preset Set markers mode 'OFF' value)
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:MATH:FUNCtion Equivalent key No equivalent key is available on the front panel. :CALCulate:FP[1-1]:TRACe[1-4]:MATH:FUNCtion Syntax :CALCulate:FP[1-1]:TRACe[1-4]:MATH:FUNCtion {NORMal|SUBTract|DIVide|ADD|MULTiply} :CALCulate:FP[1-1]:TRACe[1-4]:MATH:FUNCtion? Description Sets/Reads math operation type Parameter Description NORMal(Preset Set math operation type to 'Off' value) SUBTract Set math operation type to 'Data - Mem' DIVide Set math operation type to 'Data / Mem' Set math operation type to 'Data + Mem'...
SCPI Command Reference :CALCulate:FP[1-1]:TRACe[1-4]:SAPerture Parameter <Numeric> Range -200G to 200G Preset value Unit Resolution Equivalent key FP Menu -> Format -> Frequency Reference :CALCulate:FP[1-1]:TRACe[1-4]:SAPerture Syntax :CALCulate:FP[1-1]:TRACe[1-4]:SAPerture <numeric> :CALCulate:FP[1-1]:TRACe[1-4]:SAPerture? Description Sensitivity Aperture Parameter <Numeric> Range 100m to 20 Preset value Unit Resolution 100m Equivalent key...
SCPI Command Reference :CALCulate:PN[1-1]:ALLTrace:MARKer:DISCrete:STATe Parameter Description ON or 1 Set marker coupling function mode 'ON' OFF or 0(Preset Set marker coupling function mode 'OFF' value) Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:ALLTrace:MARKer:DISCrete:STATe Syntax :CALCulate:PN[1-1]:ALLTrace:MARKer:DISCrete:STATe {ON|OFF|1|0} :CALCulate:PN[1-1]:ALLTrace:MARKer:DISCrete:STATe? Description Sets/Reads marker movement (Continuous/Discrete)
SCPI Command Reference :CALCulate:PN[1-1]:DATA:RDATa Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:DATA:RDATa Syntax :CALCulate:PN[1-1]:DATA:RDATa <array> :CALCulate:PN[1-1]:DATA:RDATa? Description Sets/Reads the measurement raw data Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:ALLMarker:SEARch:DOMain:X Parameter <Numeric> Range 1 to 6 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:ALLMarker:SEARch:D OMain:X Syntax :CALCulate:PN[1-1]:TRACe[1-1]:ALLMarker:SEARch:DOMain:X {FRANge|BDMarker} :CALCulate:PN[1-1]:TRACe[1-1]:ALLMarker:SEARch:DOMain:X? Description Sets/Reads marker search range (X-axis) Parameter Description FRANge(Preset Set marker search range (X-axis) to 'Full Range' value) BDMarker...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STARt Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key PN Menu -> Marker Function -> Band Marker X -> Span PN Menu -> Marker Search -> Band Marker X -> Span :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STARt Syntax :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STARt <numeric>...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STOP Description OFF or 0(Preset Set bandmarker X mode 'OFF' value) Equivalent key PN Menu -> Marker Function -> Band Marker X -> Band Marker X PN Menu -> Marker Search -> Band Marker X -> Band Marker X :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STOP Syntax :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:X:STOP <numeric>...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:Y:SPAN PN Menu -> Marker Search -> Band Marker Y -> Center :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:Y:SPAN Syntax :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:Y:SPAN <numeric> :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:Y:SPAN? Description Sets/Reads the span value of bandmarker Y Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key PN Menu ->...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:BDMarker:Y:STOP Description Turns on/off bandmarker Y Parameter Description ON or 1 Set bandmarker Y mode 'ON' OFF or 0(Preset Set bandmarker Y mode 'OFF' value) Equivalent key PN Menu -> Marker Function -> Band Marker Y -> Band Marker Y PN Menu ->...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:DATA:FDATa <Numeric> Resolution Equivalent key PN Menu -> Trace View-> Copy to USER -> Copy to USER1 PN Menu -> Trace View-> Copy to USER -> Copy to USER2 PN Menu -> Trace View-> Copy to USER -> Copy to USER3 PN Menu ->...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:DATA:PDATa <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:DATA:PDATa Syntax :CALCulate:PN[1-1]:TRACe[1-1]:DATA:PDATa <array> :CALCulate:PN[1-1]:TRACe[1-1]:DATA:PDATa? Description Sets/Reads unformatted trace power data (dBc) Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:DATA:SDATa :CALCulate:PN[1-1]:TRACe[1-1]:DATA:SDATa Syntax :CALCulate:PN[1-1]:TRACe[1-1]:DATA:SDATa Description Reads the spurious judgement results (0/1) of trace data (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:DATA:SMEMory Syntax :CALCulate:PN[1-1]:TRACe[1-1]:DATA:SMEMory Description Reads the spurious judgement results (0/1) of memory data. (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:DOMain:X <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:DOMain:X Syntax :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:DOMain:X {FRANge|BDMarker} :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:DOMain:X? Description Sets/Reads analysis/search range (X-axis) Parameter Description FRANge Set analysis/search range (X-axis) to 'Full Range' BDMarker(Preset Set analysis/search range (X-axis) to 'Band Marker' value)
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:INTegral:DATA :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:INTegral:DAT Syntax :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:INTegral:DATA Description Reads the integrated phase noise, frequency range, RMS noise, RMS jitter, and residual FM of trace data. (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:INTegral:ME Mory Syntax :CALCulate:PN[1-1]:TRACe[1-1]:FUNCtion:INTegral:MEMory Description...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:HOLD Parameter Description OFF(Preset value) Set analysis type to 'Off' STATistics Set analysis type to 'Statistics' INTegral Set analysis type to 'Integral' Equivalent key PN Menu -> Marker Function -> Analysis Type :CALCulate:PN[1-1]:TRACe[1-1]:HOLD Syntax :CALCulate:PN[1-1]:TRACe[1-1]:HOLD {OFF|MAXimum|MINimum} :CALCulate:PN[1-1]:TRACe[1-1]:HOLD? Description Selects/Reads trace data hold type Parameter...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa <array> :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa? Description Sets/Reads the lower limit values of all measurement points Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent :CLEar Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:CLEar Description...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:DATA :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent :DATA Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:DATA <array> :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:DATA? Description Sets/Reads segment data of the lower limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:REPort[:DATA] Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:REPort[:DATA] Description Reads the limit test results of all measurement points in selected traces (Query Only)
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:LDATa :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:LDATa Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:LDATa <array> :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:LDATa? Description Sets/Reads the upper limit values of all measurement points Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent: CLEar Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:CLEar Description...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:DATA :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent: DATA Syntax :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:DATA <array> :CALCulate:PN[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:DATA? Description Sets/Reads segment data of the upper limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:E XECute:LPEak Syntax :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:EXECute:LPEak Description...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:TARGet:Y Description Sets/Reads the target transition definition Parameter Description POSitive Set the target transition definition to 'Positive' NEGative Set the target transition definition to 'Negative' BOTH(Preset Set the target transition definition to 'Both' value) Equivalent key PN Menu -> Marker Search -> Target -> Search Transition :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:T ARGet:Y Syntax...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:STATe Parameter Description OFF(Preset value) Set marker tracking type to 'Off' MAXimum Set marker tracking type to 'Maximum' MINimum Set marker tracking type to 'Minimum' PEAK Set marker tracking type to 'Peak' TARGet Set marker tracking type to 'Target' Equivalent key PN Menu ->...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:Y Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:Y Syntax :CALCulate:PN[1-1]:TRACe[1-1]:MARKer[1-6]:Y Description Reads the marker Y value (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:PN[1-1]:TRACe[1-1]:MATH:FUNCtion Syntax :CALCulate:PN[1-1]:TRACe[1-1]:MATH:FUNCtion {NORMal|SUBTract|DIVide|ADD|MULTiply} :CALCulate:PN[1-1]:TRACe[1-1]:MATH:FUNCtion? Description...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:SMOothing:STATe Parameter <Numeric> Range 50m to 25 Preset value Unit Resolution Equivalent key PN Menu -> Trace View -> Aperture :CALCulate:PN[1-1]:TRACe[1-1]:SMOothing:STATe Syntax :CALCulate:PN[1-1]:TRACe[1-1]:SMOothing:STATe {ON|OFF|1|0} :CALCulate:PN[1-1]:TRACe[1-1]:SMOothing:STATe? Description Turns on/off smoothing mode Parameter Description ON or 1 Set smoothing mode to 'ON' OFF or 0(Preset Set smoothing mode to 'OFF' value)
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:POWer :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:POWer Syntax :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:POWer {ON|OFF|1|0} :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:POWer? Description Turns on/off the spurious power value display Parameter Description ON or 1 Turn on the spurious power value display mode OFF or 0(Preset Turn off the spurious power value display mode value) Equivalent key PN Menu ->...
SCPI Command Reference :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:THReshold:TABLe:DATA :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:THReshold:T ABLe:DATA Syntax :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:THReshold:TABLe:DATA <array> :CALCulate:PN[1-1]:TRACe[1-1]:SPURious:THReshold:TABLe:DATA? Description Sets/Reads the threshold data Parameter <Description> Range 1...60 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:ALLTrace:LIMit:FAIL Syntax :CALCulate:SP[1-1]:ALLTrace:LIMit:FAIL Description Reads out the limit test result (Query Only) Parameter Description ON or 1...
SCPI Command Reference :CALCulate:SP[1-1]:ALLTrace:MARKer:DISCrete:STATe Parameter Description ON or 1 Set marker coupling function mode to 'ON' OFF or 0(Preset Set marker coupling function mode to 'OFF' value) Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:ALLTrace:MARKer:DISCrete:STATe Syntax :CALCulate:SP[1-1]:ALLTrace:MARKer:DISCrete:STATe {ON|OFF|1|0} :CALCulate:SP[1-1]:ALLTrace:MARKer:DISCrete:STATe? Description...
SCPI Command Reference :CALCulate:SP[1-1]:ALLTrace:MARKer:REFerence:STATe :CALCulate:SP[1-1]:ALLTrace:MARKer:REFerence:STAT Syntax :CALCulate:SP[1-1]:ALLTrace:MARKer:REFerence:STATe {ON|OFF|1|0} :CALCulate:SP[1-1]:ALLTrace:MARKer:REFerence:STATe? Description Turns on/off delta marker mode Parameter Description ON or 1 Set delta marker mode mode to 'ON' OFF or 0(Preset Set delta marker mode mode to 'OFF' value) Equivalent key SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:ACTive :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:ACTive Syntax :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:ACTive <numeric> :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:ACTive? Description Selects active marker Parameter <Numeric> Range 1 to 6 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:DO Main:X Syntax :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:DOMain:X {FRANge|BDMarker} :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:DOMain:X? Description Sets/Reads marker search range (X-axis) Parameter...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:PEAK Parameter Description FRANge(Preset Set marker search range (Y-axis) to 'Full Range' value) BDMarker Set marker search range (Y-axis) to 'Band Marker' Equivalent key SP Menu -> Marker Search -> Search Range (Y) :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:PE Syntax :CALCulate:SP[1-1]:TRACe[1-1]:ALLMarker:SEARch:PEAK Description Executes marker search all (No Query) Equivalent key SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STARt Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key SP Menu -> Marker Function -> Band Marker X -> Span SP Menu -> Marker Search -> Band Marker X -> Span :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STARt Syntax :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STARt <numeric>...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STOP Description OFF or 0(Preset Set bandmarker X mode to 'OFF' value) Equivalent key SP Menu -> Marker Function -> Band Marker X -> Band Marker X SP Menu -> Marker Search -> Band Marker X -> Band Marker X :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STOP Syntax :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:X:STOP <numeric>...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:Y:SPAN SP Menu -> Marker Search -> Band Marker Y -> Center :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:Y:SPAN Syntax :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:Y:SPAN <numeric> :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:Y:SPAN? Description Sets/Reads the span value of bandmarker Y Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:BDMarker:Y:STOP Description Turns on/off bandmarker Y Parameter Description ON or 1 Set bandmarker Y mode to 'ON' OFF or 0(Preset Set bandmarker Y mode to 'OFF' value) Equivalent key SP Menu -> Marker Function -> Band Marker Y -> Band Marker Y SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:DATA:FDATa <Numeric> Resolution Equivalent key SP Menu -> Trace View-> Copy to USER -> Copy to USER1 SP Menu -> Trace View-> Copy to USER -> Copy to USER2 SP Menu -> Trace View-> Copy to USER -> Copy to USER3 SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:DATA:UDATa <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:DATA:UDATa Syntax :CALCulate:SP[1-1]:TRACe[1-1]:DATA:UDATa <array> :CALCulate:SP[1-1]:TRACe[1-1]:DATA:UDATa? Description Sets/Reads unformatted trace data Parameter <Description> Range 1...1024 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:FORMat :CALCulate:SP[1-1]:TRACe[1-1]:FORMat Syntax :CALCulate:SP[1-1]:TRACe[1-1]:FORMat {DBM|DBV|WATT|VOLT|DBMHz|DBVHz|WHZ|VHZ} :CALCulate:SP[1-1]:TRACe[1-1]:FORMat? Description Selects SP format Parameter Description DBM(Preset Set SP format to 'dBm' value) Set SP format to 'dBV' WATT Set SP format to 'Watt' VOLT Set SP format to 'Volt' DBMHz Set SP format to 'dBm/Hz' DBVHz Set SP format to 'dBV/Hz'...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:FUNCtion:STATistics:DATA :CALCulate:SP[1-1]:TRACe[1-1]:FUNCtion:DOMain:Y? Description Sets/Reads analysis/search range (Y-axis) Parameter Description FRANge Set analysis/search range (Y-axis) to 'Full Range' BDMarker(Preset Set analysis/search range (Y-axis) to 'Band Marker' value) Equivalent key SP Menu -> Marker Function -> Analysis Range (Y) :CALCulate:SP[1-1]:TRACe[1-1]:FUNCtion:STATistics:DA Syntax :CALCulate:SP[1-1]:TRACe[1-1]:FUNCtion:STATistics:DATA...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:HOLD :CALCulate:SP[1-1]:TRACe[1-1]:HOLD Syntax :CALCulate:SP[1-1]:TRACe[1-1]:HOLD {OFF|MAXimum|MINimum} :CALCulate:SP[1-1]:TRACe[1-1]:HOLD? Description Selects/Reads trace data hold type Parameter Description OFF(Preset value) Set data hold to 'Off' MAXimum Set data hold to 'Maximum' MINimum Set data hold to 'Minimum' Equivalent key SP Menu -> Trace View -> Data Hold :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:FAIL Syntax :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:FAIL...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:CLEar <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent: CLEar Syntax :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent:CLEar Description Clears the lower limit line (No Query) Equivalent key SP Menu -> Display-> Limit Test -> Delete Lower Limit Line :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:SEGMent: COUNt Syntax...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:REPort[:DATA] Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:REPort[:DATA] Syntax :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:REPort[:DATA] Description Reads the limit test results of all measurement points in selected traces (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:CLEar Parameter <Description> Range 1...1024 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent: CLEar Syntax :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent:CLEar Description Clears the upper limit line (No Query) Equivalent key SP Menu -> Display-> Limit Test -> Delete Upper Limit Line :CALCulate:SP[1-1]:TRACe[1-1]:LIMit:UPPer:SEGMent: COUNt Syntax...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:EXECute:LPEak Description Sets/Reads segment data of the upper limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:E XECute:LPEak Syntax :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:SEARch:EXECute:LPEak Description Executes marker peak search left (No Query) Equivalent key SP Menu ->...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe Description TARGet Set the marker tracking type to 'TARGet' Equivalent key SP Menu -> Marker Search -> Tracking :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe Syntax :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe {ON|OFF|1|0} :CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe? Description Turns on/off markers Parameter Description ON or 1 Set markers to 'ON' OFF or 0(Preset Set markers to 'OFF' value)
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:MATH:FUNCtion Equivalent key No equivalent key is available on the front panel. :CALCulate:SP[1-1]:TRACe[1-1]:MATH:FUNCtion Syntax :CALCulate:SP[1-1]:TRACe[1-1]:MATH:FUNCtion {NORMal|SUBTract|DIVide|ADD|MULTiply} :CALCulate:SP[1-1]:TRACe[1-1]:MATH:FUNCtion? Description Selects math operation type Parameter Description NORMal(Preset Set math operation type to 'Off' value) SUBTract Set math operation type to 'Data - Mem' DIVide Set math operation type to 'Data / Mem' Set math operation type to 'Data + Mem'...
SCPI Command Reference :CALCulate:SP[1-1]:TRACe[1-1]:SMOothing:STATe <Numeric> Resolution Equivalent key SP Menu -> Trace View -> Aperture :CALCulate:SP[1-1]:TRACe[1-1]:SMOothing:STATe Syntax :CALCulate:SP[1-1]:TRACe[1-1]:SMOothing:STATe {ON|OFF|1|0} :CALCulate:SP[1-1]:TRACe[1-1]:SMOothing:STATe? Description Turns on/off smoothing mode Parameter Description ON or 1 Set smoothing mode to 'ON' OFF or 0(Preset Set smoothing mode to 'OFF' value) Equivalent key SP Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:ALLTrace:LIMit:FAIL :CALCulate:TR[1-1]:ALLTrace:BDMarker:X:COUPle:STATe? Description Turns on/off bandmarker coupling function Parameter Description ON or 1 Set bandmarker coupling function mode to 'ON' OFF or 0(Preset Set bandmarker coupling function mode to 'OFF' value) Equivalent key TR Menu -> Marker Function -> Couple TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:ALLTrace:MARKer:DISCrete:STATe Equivalent key TR Menu -> Marker -> Couple :CALCulate:TR[1-1]:ALLTrace:MARKer:DISCrete:STATe Syntax :CALCulate:TR[1-1]:ALLTrace:MARKer:DISCrete:STATe {ON|OFF|1|0} :CALCulate:TR[1-1]:ALLTrace:MARKer:DISCrete:STATe? Description Sets/Reads marker movement (Continuous/Discrete) Parameter Description ON or 1 Set marker movement (Continuous/Discrete) mode to 'ON' OFF or 0(Preset Set marker movement (Continuous/Discrete) mode to 'OFF' value) Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:NARRow:DATA:RDATa Parameter Description ON or 1 Set delta marker mode mode to 'ON' OFF or 0(Preset Set delta marker mode mode to 'OFF' value) Equivalent key TR Menu -> Marker -> More Functions -> Ref Marker Mode :CALCulate:TR[1-1]:NARRow:DATA:RDATa Syntax :CALCulate:TR[1-1]:NARRow:DATA:RDATa <array>...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:ALLMarker:SEARch:DOMain:X Parameter <Numeric> Range 1 to 6 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:ALLMarker:SEARch:D OMain:X Syntax :CALCulate:TR[1-1]:TRACe[1-4]:ALLMarker:SEARch:DOMain:X {FRANge|BDMarker} :CALCulate:TR[1-1]:TRACe[1-4]:ALLMarker:SEARch:DOMain:X? Description Sets/Reads marker search range (X-axis) Parameter Description FRANge(Preset Set marker search range (X-axis) to 'Full Range' value) BDMarker...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STARt Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key TR Menu -> Marker Function -> Band Marker X -> Span TR Menu -> Marker Search -> Band Marker X -> Span :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STARt Syntax :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STARt <numeric>...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STOP Description OFF or 0(Preset Set bandmarker X mode to 'OFF' value) Equivalent key TR Menu -> Marker Function -> Band Marker X -> Band Marker X TR Menu -> Marker Search -> Band Marker X -> Band Marker X :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STOP Syntax :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:X:STOP <numeric>...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:Y:SPAN TR Menu -> Marker Search -> Band Marker Y -> Center :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:Y:SPAN Syntax :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:Y:SPAN <numeric> :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:Y:SPAN? Description Sets/Reads the span value of bandmarker Y Parameter <Numeric> Range 0 to 2T Preset value Unit Resolution Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:BDMarker:Y:STOP Description Turns on/off bandmarker Y Parameter Description ON or 1 Set bandmarker Y mode to 'ON' OFF or 0(Preset Set bandmarker Y mode to 'OFF' value) Equivalent key TR Menu -> Marker Function -> Band Marker Y -> Band Marker Y TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:DATA:FDATa <Numeric> Resolution Equivalent key TR Menu -> Trace View-> Copy to USER -> Copy to USER1 TR Menu -> Trace View-> Copy to USER -> Copy to USER2 TR Menu -> Trace View-> Copy to USER -> Copy to USER3 TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:DATA:UDATa <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:DATA:UDATa Syntax :CALCulate:TR[1-1]:TRACe[1-4]:DATA:UDATa <array> :CALCulate:TR[1-1]:TRACe[1-4]:DATA:UDATa? Description Sets/Reads unformatted trace data Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:FREQuency :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:FREQuency Syntax :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:FREQuency {HZ|DHZ|PCT|PPM} :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:FREQuency? Description Sets/Reads the frequency format Parameter Description HZ(Preset value) Set the frequency format to 'Hz' Set the frequency format to 'ΔHz' Set the frequency format to '%' Set the frequency format to 'ppm' Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:PHASe:WRAP Parameter Description DEG(Preset Set phase format on transient measurement to 'Deg' value) Set phase format on transient measurement to 'Rad' GRAD Set phase format on transient measurement to 'Grad' Equivalent key TR Menu -> Format -> Phase Unit :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:PHASe:WRAP Syntax :CALCulate:TR[1-1]:TRACe[1-4]:FORMat:PHASe:WRAP {ON|OFF|1|0}...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y Syntax :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y {FRANge|BDMarker} :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:DOMain:Y? Description Sets/Reads analysis/search range (Y-axis) Parameter Description FRANge Set analysis/search range (Y-axis) to 'Full Range' BDMarker(Preset Set analysis/search range (Y-axis) to 'Band Marker' value) Equivalent key TR Menu -> Marker Function -> Analysis Range (Y) :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:STATistics:D Syntax :CALCulate:TR[1-1]:TRACe[1-4]:FUNCtion:STATistics:DATA...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:HOLD Description STATistics Set analysis type to 'Statistics' Equivalent key TR Menu -> Marker Function -> Analysis Type :CALCulate:TR[1-1]:TRACe[1-4]:HOLD Syntax :CALCulate:TR[1-1]:TRACe[1-4]:HOLD {OFF|MAXimum|MINimum} :CALCulate:TR[1-1]:TRACe[1-4]:HOLD? Description Sets/Reads trace data hold type Parameter Description OFF(Preset value) Set data hold to 'Off' MAXimum Set data hold to 'Maximum' MINimum...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:CLEar Description Sets/Reads the lower limit values of all measurement points Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent :CLEar Syntax :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:CLEar Description Clears the lower limit line (No Query) Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:SEGMent:DATA? Description Sets/Reads segment data of the lower limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] Syntax :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:REPort[:DATA] Description Reads the limit test results of all measurement points in selected traces (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent:CLEar Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent: CLEar Syntax :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent:CLEar Description Clears the upper limit line (No Query) Equivalent key TR Menu -> Display-> Limit Test -> Delete Upper Limit Line :CALCulate:TR[1-1]:TRACe[1-4]:LIMit:UPPer:SEGMent: COUNt Syntax...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch:EXECute:LPEak Description Sets/Reads segment data of the upper limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch: EXECute:LPEak Syntax :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:SEARch:EXECute:LPEak Description Executes marker peak search left (No Query) Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:STATe Description TARGet Set the marker tracking type to 'Target' Equivalent key TR Menu -> Marker Search -> Tracking :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:STATe Syntax :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:STATe {ON|OFF|1|0} :CALCulate:TR[1-1]:TRACe[1-4]:MARKer[1-6]:STATe? Description Turns on/off markers Parameter Description ON or 1 Set markers mode to 'ON' OFF or 0(Preset Set markers mode to 'OFF' value)
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:MATH:FUNCtion Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:TRACe[1-4]:MATH:FUNCtion Syntax :CALCulate:TR[1-1]:TRACe[1-4]:MATH:FUNCtion {NORMal|SUBTract|DIVide|ADD|MULTiply} :CALCulate:TR[1-1]:TRACe[1-4]:MATH:FUNCtion? Description Selects math operation type Parameter Description NORMal(Preset Set math operation type to 'Off' value) SUBTract Set math operation type to 'Data - Mem' DIVide Set math operation type to 'Data / Mem' Set math operation type to 'Data + Mem'...
SCPI Command Reference :CALCulate:TR[1-1]:TRACe[1-4]:SMOothing:APERture Parameter <Numeric> Range -200G to 200G Preset value Unit Resolution Equivalent key TR Menu -> Format -> Frequency Reference :CALCulate:TR[1-1]:TRACe[1-4]:SMOothing:APERture Syntax :CALCulate:TR[1-1]:TRACe[1-4]:SMOothing:APERture <numeric> :CALCulate:TR[1-1]:TRACe[1-4]:SMOothing:APERture? Description Smoothing aperture Parameter <Numeric> Range 50m to 25 Preset value Unit Resolution Equivalent key TR Menu ->...
SCPI Command Reference :CALCulate:TR[1-1]:WIDE:DATA:RDATa :CALCulate:TR[1-1]:WIDE:DATA:RDATa Syntax :CALCulate:TR[1-1]:WIDE:DATA:RDATa <array> :CALCulate:TR[1-1]:WIDE:DATA:RDATa? Description Sets/Reads the measurement raw data Parameter <Description> Range 1...1001 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:TR[1-1]:WIDE:DATA:XDATa Syntax :CALCulate:TR[1-1]:WIDE:DATA:XDATa Description Reads the X-axis data (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:USER[1-1]:ALLTrace:BDMarker:X:COUPle:STATe :CALCulate:USER[1-1]:ALLTrace:BDMarker:X:COUPle:S TATe Syntax :CALCulate:USER[1-1]:ALLTrace:BDMarker:X:COUPle:STATe {ON|OFF|1|0} :CALCulate:USER[1-1]:ALLTrace:BDMarker:X:COUPle:STATe? Description Turns on/off bandmarker coupling function Parameter Description ON or 1 Set bandmarker coupling function mode to 'ON' OFF or 0(Preset Set bandmarker coupling function mode to 'OFF' value) Equivalent key USER Menu ->...
SCPI Command Reference :CALCulate:USER[1-1]:ALLTrace:MARKer:DISCrete:STATe Parameter Description ON or 1 Set marker coupling function mode to 'ON' OFF or 0(Preset Set marker coupling function mode to 'OFF' value) Equivalent key USER Menu -> Marker -> Couple :CALCulate:USER[1-1]:ALLTrace:MARKer:DISCrete:ST Syntax :CALCulate:USER[1-1]:ALLTrace:MARKer:DISCrete:STATe {ON|OFF|1|0} :CALCulate:USER[1-1]:ALLTrace:MARKer:DISCrete:STATe? Description Sets/Reads marker discrete mode Parameter...
SCPI Command Reference :CALCulate:USER[1-1]:ALLTrace:MARKer:REFerence:STATe Equivalent key USER Menu -> Marker -> More Functions -> Ref Marker :CALCulate:USER[1-1]:ALLTrace:MARKer:REFerence:S TATe Syntax :CALCulate:USER[1-1]:ALLTrace:MARKer:REFerence:STATe {ON|OFF|1|0} :CALCulate:USER[1-1]:ALLTrace:MARKer:REFerence:STATe? Description Turns on/off delta marker mode Parameter Description ON or 1 Set delta marker mode mode to 'ON' OFF or 0(Preset Set delta marker mode mode to 'OFF' value)
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:ALLMarker:SEARch:DOMain:Y Description Sets/Reads marker search X range Parameter Description FRANge(Preset Set marker search X range to 'Full Range' value) BDMarker Set marker search X range to 'Band Marker' Equivalent key USER Menu -> Marker Search -> Search Range (X) :CALCulate:USER[1-1]:TRACe[1-8]:ALLMarker:SEARch: DOMain:Y Syntax...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:SPAN Parameter <Numeric> Range -1T to 1T Preset value Unit Resolution Equivalent key USER Menu -> Marker Function -> Band Marker X -> Center USER Menu -> Marker Search -> Band Marker X -> Center :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:SPAN Syntax :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:SPAN <numeric>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:STATe <Numeric> Unit Resolution Equivalent key USER Menu -> Marker Function -> Band Marker X -> Start USER Menu -> Marker Search -> Band Marker X -> Start :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:STATe Syntax :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:STATe {ON|OFF|1|0} :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:X:STATe? Description Turns on/off bandmarker X Parameter Description ON or 1...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:CENTer :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:CENTer Syntax :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:CENTer <numeric> :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:CENTer? Description Sets/Reads the center value of bandmarker Y Parameter <Numeric> Range -1T to 1T Preset value Unit Resolution Equivalent key USER Menu -> Marker Function -> Band Marker Y -> Center USER Menu ->...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:STATe Parameter <Numeric> Range -1T to 1T Preset value Unit Resolution Equivalent key USER Menu -> Marker Function -> Band Marker Y -> Start USER Menu -> Marker Search -> Band Marker Y -> Start :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:STATe Syntax :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:STATe {ON|OFF|1|0} :CALCulate:USER[1-1]:TRACe[1-8]:BDMarker:Y:STATe? Description...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:DATA:COPY <Numeric> Resolution Equivalent key USER Menu -> Marker Function -> Band Marker Y -> Stop USER Menu -> Marker Search -> Band Marker Y -> Stop :CALCulate:USER[1-1]:TRACe[1-8]:DATA:COPY Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:COPY <numeric> Description Copies trace data to the user trace (No Query) Parameter <Numeric>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:DATA:FMEMory <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:DATA:FMEMory Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:FMEMory <array> :CALCulate:USER[1-1]:TRACe[1-8]:DATA:FMEMory? Description Sets/Reads formatted memory data Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:DATA:STARt Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:DATA:STARt Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:STARt Description start frequency (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:DATA:STOP Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:STOP...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:DATA:UMEMory :CALCulate:USER[1-1]:TRACe[1-8]:DATA:UMEMory Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:UMEMory <array> :CALCulate:USER[1-1]:TRACe[1-8]:DATA:UMEMory? Description Sets/Reads unformatted memory data Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:DATA:XDATa Syntax :CALCulate:USER[1-1]:TRACe[1-8]:DATA:XDATa <array> :CALCulate:USER[1-1]:TRACe[1-8]:DATA:XDATa? Description Sets/Reads the X data Parameter <Description>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:FUNCtion:TYPE :CALCulate:USER[1-1]:TRACe[1-8]:FUNCtion:TYPE Syntax :CALCulate:USER[1-1]:TRACe[1-8]:FUNCtion:TYPE {OFF|STATistics} :CALCulate:USER[1-1]:TRACe[1-8]:FUNCtion:TYPE? Description Analysis type Parameter Description OFF(Preset value) Set marker search TARGet to 'Off' STATistics Set marker search TARGet to 'Statistics' Equivalent key USER Menu -> Marker Function -> Analysis Type :CALCulate:USER[1-1]:TRACe[1-8]:HOLD Syntax :CALCulate:USER[1-1]:TRACe[1-8]:HOLD {OFF|MAXimum|MINimum} :CALCulate:USER[1-1]:TRACe[1-8]:HOLD?
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:LDATa When the limit test is set to OFF, OFF or 0 is always read out. Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:LDATa Syntax :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:LDATa <array> :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:LDATa? Description Sets/Reads the lower limit values of all measurement points Parameter <Description>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGMent:DATA <Numeric> Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGM ent:DATA Syntax :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGMent:DATA <array> :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGMent:DATA? Description Sets/Reads segment data of the lower limit line Parameter <Description> Range 1...400 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:LDATa Parameter Description ON or 1 Turn on the limit test function mode OFF or 0(Preset Turn off the limit test function mode value) Equivalent key USER Menu -> Display -> Limit Test -> Limit Test :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:LDATa Syntax :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:LDATa <array>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:SEGMent:DATA Parameter <Numeric> Range 1 to 100 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:SEGMe nt:DATA Syntax :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:SEGMent:DATA <array> :CALCulate:USER[1-1]:TRACe[1-8]:LIMit:UPPer:SEGMent:DATA? Description Sets/Reads segment data of the upper limit line Parameter <Description>...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TARGet:TRAnsitio Description BOTH Set marker-search-peak polality type to 'Both' Equivalent key USER Menu -> Marker Search -> Peak -> Peak Polarity :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARc h:TARGet:TRAnsition Syntax :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TARGet:TRAnsition {POSitive|NEGative|BOTH} :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TARGet:TRAnsition? Description marker-target transition type Parameter Description POSitive Set marker-target transition type to 'Positive' NEGative Set marker-target transition type to 'Negative' BOTH(Preset...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TRACking:TYPE :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARc h:TRACking:TYPE Syntax :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TRACking:TYPE {OFF|MAXimum|MINimum|PEAK|TARGet} :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:SEARch:TRACking:TYPE? Description Sets/Reads the marker tracking type Parameter Description OFF(Preset value) Set search tracking type to 'Off' MAXimum Set search tracking type to 'Maximum' MINimum Set search tracking type to 'Minimum' PEAK Set search tracking type to 'Peak' TARGet...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:Y Parameter <Numeric> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:Y Syntax :CALCulate:USER[1-1]:TRACe[1-8]:MARKer[1-6]:Y Description Reads the marker y position (Query Only) Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:MATH:FUNCtion Syntax :CALCulate:USER[1-1]:TRACe[1-8]:MATH:FUNCtion...
SCPI Command Reference :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:APERture Equivalent key No equivalent key is available on the front panel. :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:APERtur Syntax :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:APERture <numeric> :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:APERture? Description Sets/Reads the smoothing aperture value Parameter <Numeric> Range 50m to 25 Preset value Unit Resolution Equivalent key USER Menu -> Trace View -> Aperture :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:STATe Syntax :CALCulate:USER[1-1]:TRACe[1-8]:SMOothing:STATe {ON|OFF|1|0}...
SCPI Command Reference :CONTrol:HANDler:B[:DATA] Parameter <Numeric> Range 0 to 255 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :CONTrol:HANDler:B[:DATA] Syntax :CONTrol:HANDler:B[:DATA] <numeric> Description Outputs data using port B (No Query) Parameter <Numeric> Range 0 to 255 Preset value Unit...
SCPI Command Reference :CONTrol:HANDler:C:MODE Equivalent key No equivalent key is available on the front panel. :CONTrol:HANDler:C:MODE Syntax :CONTrol:HANDler:C:MODE {INPut|OUTPut} :CONTrol:HANDler:C:MODE? Description Selects input/output mode on port C Parameter Description INPut(Preset Set input/output mode on port C to 'Input' value) OUTPut Set input/output mode on port C to 'Output' Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :CONTrol:HANDler:E[:DATA] Parameter Description INPut(Preset Set input/output mode on port D to 'Input' value) OUTPut Set input/output mode on port D to 'Output' Equivalent key No equivalent key is available on the front panel. :CONTrol:HANDler:E[:DATA] Syntax :CONTrol:HANDler:E[:DATA] <numeric> :CONTrol:HANDler:E[:DATA]? Description Inputs/outputs data using port E(port C + port D;...
SCPI Command Reference :CONTrol:HANDler:OUTPut[1-2][:DATA] :CONTrol:HANDler:OUTPut[1-2][:DATA] Syntax :CONTrol:HANDler:OUTPut[1-2][:DATA] <numeric> :CONTrol:HANDler:OUTPut[1-2][:DATA]? Description Sets/Reads OUTPUT1 and/or OUTPUT2 Parameter <Numeric> Range 0 to 1 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :DISPlay:CLOCk Syntax :DISPlay:CLOCk {ON|OFF|1|0} :DISPlay:CLOCk? Description Turns on/off internal clock display Parameter...
SCPI Command Reference :DISPlay:ECHO:CLEar Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :DISPlay:ECHO:CLEar Syntax :DISPlay:ECHO:CLEar Description Clears echo window (No Query) Equivalent key PN Menu -> Macro Setup -> Echo Window Menu -> Clear Echo SP Menu ->...
SCPI Command Reference :DISPlay:ECHO:STATe :DISPlay:ECHO:FSIZe? Description Sets/Reads the font size in echo window Parameter <Numeric> Range 9|10|11|12|14|16|18|20|22|24|26|28|36|48|72|96|112 Preset value Unit Resolution Equivalent key PN Menu -> Macro Setup -> Echo Window Menu -> Echo Font Size SP Menu -> Macro Setup -> Echo Window Menu -> Echo Font Size FP Menu ->...
SCPI Command Reference :DISPlay:FP[1-1]:ALLTrace:PERSistence:CLEar :DISPlay:ENABle? Description Enable/disable trace update Parameter Description ON or 1(Preset Set trace update mode to 'ON' value) OFF or 0 Set trace update mode to 'OFF' Equivalent key PN Menu -> Display -> Update SP Menu -> Display -> Update FP Menu ->...
SCPI Command Reference :DISPlay:FP[1-1]:ANNotation:MEASurement:STATe Description RIGHt Set the marker information position to 'Right' Equivalent key FP Menu -> Display -> Marker Information :DISPlay:FP[1-1]:ANNotation:MEASurement:STATe Syntax :DISPlay:FP[1-1]:ANNotation:MEASurement:STATe {ON|OFF|1|0} :DISPlay:FP[1-1]:ANNotation:MEASurement:STATe? Description Turns on/off measurement conditions Parameter Description ON or 1(Preset Set measurement conditions mode to 'ON' value) OFF or 0 Set measurement conditions mode to 'OFF'...
SCPI Command Reference :DISPlay:FP[1-1]:LABel:DATA Parameter Description Set Y graticule label to 'OFF' SHORt(Preset Set Y graticule label to 4 digits value) MIDDle Set Y graticule label to 8 digits LONG Set Y graticule label to 12 digits Equivalent key FP Menu -> Display -> Y # of Digits :DISPlay:FP[1-1]:LABel:DATA Syntax :DISPlay:FP[1-1]:LABel:DATA <String>...
SCPI Command Reference :DISPlay:FP[1-1]:LIMit:FSIGn Equivalent key FP Menu -> Display -> Title Label :DISPlay:FP[1-1]:LIMit:FSIGn Syntax :DISPlay:FP[1-1]:LIMit:FSIGn {ON|OFF|1|0} :DISPlay:FP[1-1]:LIMit:FSIGn? Description Turns on/off the limit test judgement display Parameter Description ON or 1(Preset Turn on the limit test judgement display mode value) OFF or 0 Turn off the limit test judgement display mode Equivalent key...
SCPI Command Reference :DISPlay:FP[1-1]:STATe Parameter Description D11_23(Preset Set the trace layout to 'x3' value) D12_34 Set the trace layout to 'x4' Equivalent key FP Menu -> Display -> Allocate :DISPlay:FP[1-1]:STATe Syntax :DISPlay:FP[1-1]:STATe {ON|OFF|1|0} :DISPlay:FP[1-1]:STATe? Description measurement display on/off. At least one meas window must be turned on. Parameter Description ON or 1(Preset...
SCPI Command Reference :DISPlay:FP[1-1]:TRACe[1-4]:PERSistence:CLEar Parameter Description Set data and/or memory trace to 'Off' DATA(Preset Set data and/or memory trace to 'Data' value) MEMory Set data and/or memory trace to 'Memory' BOTH Set data and/or memory race to 'Both' (data and memory) Equivalent key FP Menu ->...
SCPI Command Reference :DISPlay:FP[1-1]:Y[:SCALe]:DIVisions Parameter <Numeric> Range 0 to 30 Preset value Unit Resolution Equivalent key FP Menu -> Scale -> Reference Position :DISPlay:FP[1-1]:Y[:SCALe]:DIVisions Syntax :DISPlay:FP[1-1]:Y[:SCALe]:DIVisions <numeric> :DISPlay:FP[1-1]:Y[:SCALe]:DIVisions? Description # of Y division Parameter <Numeric> Range 4 to 30 Preset value Unit Resolution Equivalent key...
SCPI Command Reference :DISPlay:MESSage:CLEar :DISPlay:MESSage:CLEar Syntax :DISPlay:MESSage:CLEar Description Clears caution/message (No Query) Equivalent key No equivalent key is available on the front panel. :DISPlay:PN[1-1]:ALLTrace:PERSistence:CLEar Syntax :DISPlay:PN[1-1]:ALLTrace:PERSistence:CLEar Description Clears all persistent traces (No Query) Equivalent key No equivalent key is available on the front panel. :DISPlay:PN[1-1]:ANNotation:MARKer:POSition Syntax :DISPlay:PN[1-1]:ANNotation:MARKer:POSition {LEFT|RIGHt}...
SCPI Command Reference :DISPlay:PN[1-1]:GRATicule:AXIS:Y:RELative Equivalent key PN Menu -> Display -> Meas Condition :DISPlay:PN[1-1]:GRATicule:AXIS:Y:RELative Syntax :DISPlay:PN[1-1]:GRATicule:AXIS:Y:RELative {ON|OFF|1|0} :DISPlay:PN[1-1]:GRATicule:AXIS:Y:RELative? Description Sets/Reads ralative Y-scale Parameter Description ON or 1 Set relative Y-sclae mode to 'ON' OFF or 0(Preset Set relative Y-scale mode to 'OFF' value) Equivalent key PN Menu ->...
SCPI Command Reference :DISPlay:PN[1-1]:LABel:STATe Parameter <String> Range Preset value "" Unit Resolution Equivalent key PN Menu -> Display -> Edit Title Label :DISPlay:PN[1-1]:LABel:STATe Syntax :DISPlay:PN[1-1]:LABel:STATe {ON|OFF|1|0} :DISPlay:PN[1-1]:LABel:STATe? Description Shows/Hides window title label Parameter Description ON or 1 Show window title label OFF or 0(Preset Hide window title label value)
SCPI Command Reference :DISPlay:PN[1-1]:MAXimize :DISPlay:PN[1-1]:MAXimize Syntax :DISPlay:PN[1-1]:MAXimize {ON|OFF|1|0} :DISPlay:PN[1-1]:MAXimize? Description Maximizes active trace Parameter Description ON or 1 Set maximize active trace mode to 'ON' OFF or 0(Preset Set maximize active trace mode to 'OFF' value) Equivalent key No equivalent key is available on the front panel. :DISPlay:PN[1-1]:STATe Syntax :DISPlay:PN[1-1]:STATe {ON|OFF|1|0}...
SCPI Command Reference :DISPlay:PN[1-1]:TRACe[1-1]:LABel:DATA Parameter Description ON or 1 Enable the marker list OFF or 0(Preset Disable the marker list value) Equivalent key PN Menu -> Marker -> Marker List :DISPlay:PN[1-1]:TRACe[1-1]:LABel:DATA Syntax :DISPlay:PN[1-1]:TRACe[1-1]:LABel:DATA <String> :DISPlay:PN[1-1]:TRACe[1-1]:LABel:DATA? Description Edits trace title lable on phase noise measurement Parameter <String>...
SCPI Command Reference :DISPlay:PN[1-1]:TRACe[1-1]:MODE :DISPlay:PN[1-1]:TRACe[1-1]:MODE Syntax :DISPlay:PN[1-1]:TRACe[1-1]:MODE {OFF|DATA|MEMory|BOTH} :DISPlay:PN[1-1]:TRACe[1-1]:MODE? Description Sets/Reads data and/or memory trace display Parameter Description Set data and/or memory trace to 'Off' DATA(Preset Set data and/or memory trace to 'Data' value) MEMory Set data and/or memory trace to 'Memory' BOTH Set data and/or memory race to 'Both' (data and memory) Equivalent key...
SCPI Command Reference :DISPlay:PN[1-1]:TRACe[1-1]:Y[:SCALe]:RPOSition :DISPlay:PN[1-1]:TRACe[1-1]:Y[:SCALe]:RPOSition Syntax :DISPlay:PN[1-1]:TRACe[1-1]:Y[:SCALe]:RPOSition <numeric> :DISPlay:PN[1-1]:TRACe[1-1]:Y[:SCALe]:RPOSition? Description scale reference position Parameter <Numeric> Range 0 to 30 Preset value Unit Resolution Equivalent key PN Menu -> Scale -> Reference Position :DISPlay:PN[1-1]:Y[:SCALe]:DIVisions Syntax :DISPlay:PN[1-1]:Y[:SCALe]:DIVisions <numeric> :DISPlay:PN[1-1]:Y[:SCALe]:DIVisions? Description # of Y division Parameter <Numeric>...
SCPI Command Reference :DISPlay:SP[1-1]:ALLTrace:PERSistence:CLEar Parameter Description ON or 1(Preset Show softkey value) OFF or 0 Hide softkey Equivalent key No equivalent key is available on the front panel. :DISPlay:SP[1-1]:ALLTrace:PERSistence:CLEar Syntax :DISPlay:SP[1-1]:ALLTrace:PERSistence:CLEar Description Clears all persistent traces (No Query) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :DISPlay:SP[1-1]:MAXimize Parameter Description ON or 1(Preset Turn on the limit test judgement display mode value) OFF or 0 Turn off the limit test judgement display mode Equivalent key SP Menu -> Display -> Limit Test -> Fail Sign :DISPlay:SP[1-1]:MAXimize Syntax :DISPlay:SP[1-1]:MAXimize {ON|OFF|1|0}...
SCPI Command Reference :DISPlay:SP[1-1]:TABLe[:STATe] USER Menu -> Measurement View -> Show Window -> Spectrum Monitor :DISPlay:SP[1-1]:TABLe[:STATe] Syntax :DISPlay:SP[1-1]:TABLe[:STATe] {ON|OFF|1|0} :DISPlay:SP[1-1]:TABLe[:STATe]? Description Turns on/off the marker list Parameter Description ON or 1 Enable the marker list OFF or 0(Preset Disable the marker list value) Equivalent key SP Menu ->...
SCPI Command Reference :DISPlay:SP[1-1]:TRACe[1-1]:MODE Parameter Description ON or 1(Preset Turn on the limit line mode value) OFF or 0 Turn off the limit line mode Equivalent key SP Menu -> Display -> Limit Test -> Limit Line :DISPlay:SP[1-1]:TRACe[1-1]:MODE Syntax :DISPlay:SP[1-1]:TRACe[1-1]:MODE {OFF|DATA|MEMory|BOTH} :DISPlay:SP[1-1]:TRACe[1-1]:MODE? Description Sets/Reads data and/or memory trace display...
SCPI Command Reference :DISPlay:SP[1-1]:TRACe[1-1]:Y[:SCALe]:AUTO Parameter Description ON or 1 Set persistence mode to 'ON' OFF or 0(Preset Set persistence mode to 'OFF' value) Equivalent key SP Menu -> Trace View -> Persistence -> Persistence Mode :DISPlay:SP[1-1]:TRACe[1-1]:Y[:SCALe]:AUTO Syntax :DISPlay:SP[1-1]:TRACe[1-1]:Y[:SCALe]:AUTO Description Executes autoscale (No Query) Equivalent key SP Menu ->...
SCPI Command Reference :DISPlay:TR[1-1]:ALLTrace:PERSistence:CLEar <Numeric> Unit Resolution Equivalent key SP Menu -> Scale -> Divisions :DISPlay:TR[1-1]:ALLTrace:PERSistence:CLEar Syntax :DISPlay:TR[1-1]:ALLTrace:PERSistence:CLEar Description Clears all stored traces (No Query) Equivalent key No equivalent key is available on the front panel. :DISPlay:TR[1-1]:ALLTrace:Y:SCALe:AUTO Syntax :DISPlay:TR[1-1]:ALLTrace:Y:SCALe:AUTO Description auto scale all (No Query) Equivalent key TR Menu ->...
SCPI Command Reference :DISPlay:TR[1-1]:GRATicule:AXIS:Y:RELative Parameter Description ON or 1(Preset Set measurement conditions mode to 'ON' value) OFF or 0 Set measurement conditions mode to 'OFF' Equivalent key TR Menu -> Display -> Meas Condition :DISPlay:TR[1-1]:GRATicule:AXIS:Y:RELative Syntax :DISPlay:TR[1-1]:GRATicule:AXIS:Y:RELative {ON|OFF|1|0} :DISPlay:TR[1-1]:GRATicule:AXIS:Y:RELative? Description Sets/Reads relative Y-scale Parameter Description...
SCPI Command Reference :DISPlay:TR[1-1]:MAXimize Parameter Description ON or 1(Preset Turn on the limit test judgement display mode value) OFF or 0 Turn off the limit test judgement display mode Equivalent key TR Menu -> Display -> Limit Test -> Fail Sign :DISPlay:TR[1-1]:MAXimize Syntax :DISPlay:TR[1-1]:MAXimize {ON|OFF|1|0}...
SCPI Command Reference :DISPlay:TR[1-1]:TABLe[:STATe] USER Menu -> Measurement View -> Show Window -> Transient :DISPlay:TR[1-1]:TABLe[:STATe] Syntax :DISPlay:TR[1-1]:TABLe[:STATe] {ON|OFF|1|0} :DISPlay:TR[1-1]:TABLe[:STATe]? Description Turns on/off the marker list Parameter Description ON or 1 Enable the marker list OFF or 0(Preset Disable the marker list value) Equivalent key TR Menu ->...
SCPI Command Reference :DISPlay:TR[1-1]:TRACe[1-4]:MODE Parameter Description ON or 1(Preset Turn on the limit line mode value) OFF or 0 Turn off the limit line mode Equivalent key TR Menu -> Display -> Limit Test -> Limit Line :DISPlay:TR[1-1]:TRACe[1-4]:MODE Syntax :DISPlay:TR[1-1]:TRACe[1-4]:MODE {OFF|DATA|MEMory|BOTH} :DISPlay:TR[1-1]:TRACe[1-4]:MODE? Description Sets/Reads data and/or memory trace display...
SCPI Command Reference :DISPlay:TR[1-1]:TRACe[1-4]:Y[:SCALe]:AUTO Parameter Description ON or 1 Set persistence mode to 'ON' OFF or 0(Preset Set persistence mode to 'OFF' value) Equivalent key TR Menu -> Trace View -> Persistence -> Persistence Mode :DISPlay:TR[1-1]:TRACe[1-4]:Y[:SCALe]:AUTO Syntax :DISPlay:TR[1-1]:TRACe[1-4]:Y[:SCALe]:AUTO Description Executes autoscale (No Query) Equivalent key TR Menu ->...
SCPI Command Reference :DISPlay:UPDate:IMMediate <Numeric> Unit Resolution Equivalent key TR Menu -> Scale -> Divisions :DISPlay:UPDate:IMMediate Syntax :DISPlay:UPDate:IMMediate Description Update display force (No Query) Equivalent key No equivalent key is available on the front panel. :DISPlay:USER[1-1]:ALLTrace:PERSistence:CLEar Syntax :DISPlay:USER[1-1]:ALLTrace:PERSistence:CLEar Description Clears all stored traces (No Query) Equivalent key USER Menu ->...
SCPI Command Reference :DISPlay:USER[1-1]:ANNotation:MEASurement:STATe :DISPlay:USER[1-1]:ANNotation:MEASurement:STATe Syntax :DISPlay:USER[1-1]:ANNotation:MEASurement:STATe {ON|OFF|1|0} :DISPlay:USER[1-1]:ANNotation:MEASurement:STATe? Description Turns on/off measurement conditions Parameter Description ON or 1(Preset Set measurement conditions mode to 'ON' value) OFF or 0 Set measurement conditions mode to 'OFF' Equivalent key USER Menu -> Display -> Meas Condition :DISPlay:USER[1-1]:GRATicule:AXIS:Y:RELative Syntax :DISPlay:USER[1-1]:GRATicule:AXIS:Y:RELative {ON|OFF|1|0}...
SCPI Command Reference :DISPlay:USER[1-1]:LABel:DATA Description SHORt(Preset Set Y graticule label to 4 digits value) MIDDle Set Y graticule label to 8 digits LONG Set Y graticule label to 12 digits Equivalent key USER Menu -> Display -> Y # of Digits :DISPlay:USER[1-1]:LABel:DATA Syntax :DISPlay:USER[1-1]:LABel:DATA <String>...
SCPI Command Reference :DISPlay:USER[1-1]:LIMit:FSIGn :DISPlay:USER[1-1]:LIMit:FSIGn Syntax :DISPlay:USER[1-1]:LIMit:FSIGn {ON|OFF|1|0} :DISPlay:USER[1-1]:LIMit:FSIGn? Description Turns on/off the limit test judgement display Parameter Description ON or 1(Preset Turn on the limit test judgement display mode value) OFF or 0 Turn off the limit test judgement display mode Equivalent key USER Menu ->...
SCPI Command Reference :DISPlay:USER[1-1]:TABLe[:STATe] Description OFF or 0(Preset hide user defined window value) Equivalent key PN Menu -> Measurement View -> Show Window -> User SP Menu -> Measurement View -> Show Window -> User FP Menu -> Measurement View -> Show Window -> User TR Menu ->...
SCPI Command Reference :DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE Equivalent key USER Menu -> Trace View -> Trace Label :DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE Syntax :DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE {ON|OFF|1|0} :DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE? Description Turns on/off the limit line Parameter Description ON or 1(Preset Turn on the limit line mode value) OFF or 0 Turn off the limit line mode Equivalent key USER Menu ->...
SCPI Command Reference :DISPlay:USER[1-1]:TRACe[1-8]:STATe Parameter Description ON or 1 Clears persistent data OFF or 0(Preset Does not clear persistent data value) Equivalent key USER Menu -> Trace View -> Persistence -> Persistence Mode :DISPlay:USER[1-1]:TRACe[1-8]:STATe Syntax :DISPlay:USER[1-1]:TRACe[1-8]:STATe {ON|OFF|1|0} :DISPlay:USER[1-1]:TRACe[1-8]:STATe? Description Sets/Reads persistence mode Parameter Description ON or 1(Preset...
SCPI Command Reference :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:AUTO :DISPlay:USER[1-1]:TRACe[1-8]:X:UNIT? Description X axis unit Parameter <String> Range Preset value "U" Unit Resolution Equivalent key USER Menu -> Scale -> X Unit :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:AUTO Syntax :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:AUTO Description Executes autoscale (No Query) Equivalent key USER Menu -> Scale -> Auto Scale :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:PDIVision Syntax :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:PDIVision <numeric>...
SCPI Command Reference :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:RPOSition Description scale reference level Parameter <Numeric> Range -200G to 200G Preset value Unit Resolution Equivalent key USER Menu -> Scale -> Reference Value USER Menu -> Scale -> Marker -> Reference :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:RPOSition Syntax :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:RPOSition <numeric> :DISPlay:USER[1-1]:TRACe[1-8]:Y[:SCALe]:RPOSition? Description scale reference position Parameter...
SCPI Command Reference :DISPlay:USER[1-1]:Y[:SCALe]:DIVisions <String> Preset value "U" Unit Resolution Equivalent key USER Menu -> Scale -> Y Unit :DISPlay:USER[1-1]:Y[:SCALe]:DIVisions Syntax :DISPlay:USER[1-1]:Y[:SCALe]:DIVisions <numeric> :DISPlay:USER[1-1]:Y[:SCALe]:DIVisions? Description # of Y division Parameter <Numeric> Range 4 to 30 Preset value Unit Resolution Equivalent key USER Menu ->...
SCPI Command Reference :FORMat:DATA Parameter Description NORMal(Preset Set byte order so that a byte containing MSB (Most Significant value) Bit) is transferred first SWAPped Set byte order so that a byte containing LSB (Least Significant Bit) is transferred first Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :HCOPy:IMMediate Description Sets/Reads inverse color print mode Parameter Description NORMal(Preset Set inverse color print mode to 'Off' value) INVert Set inverse color print mode to 'On' Equivalent key PN Menu -> System -> Invert Image SP Menu -> System -> Invert Image FP Menu ->...
SCPI Command Reference *ESR <Numeric> Resolution Equivalent key No equivalent key is available on the front panel. *ESR Syntax *ESR Description Reads standard event status register value (Query Only) Equivalent key No equivalent key is available on the front panel. *IDN Syntax *IDN...
SCPI Command Reference *RST *RST Syntax *RST Description Preset (No Query) Equivalent key No equivalent key is available on the front panel. *SRE Syntax *SRE <numeric> *SRE? Description Sets service request enable register Parameter <Numeric> Range 0 to 255 Preset value Unit Resolution Equivalent key...
SCPI Command Reference :INITiate:FP[1-1]:IMMediate Description Sets/Reads the trigger continuous mode Parameter Description ON or 1 Set trigger continuous mode to 'ON' OFF or 0(Preset Set trigger continuous mode to 'OFF' value) Equivalent key No equivalent key is available on the front panel. :INITiate:FP[1-1]:IMMediate Syntax :INITiate:FP[1-1]:IMMediate...
SCPI Command Reference :INITiate:SP[1-1]:IMMediate :INITiate:SP[1-1]:CONTinuous? Description Sets/Reads trigger continuous mode Parameter Description ON or 1 Set trigger continuous mode to 'ON' OFF or 0(Preset Set trigger continuous mode to 'OFF' value) Equivalent key No equivalent key is available on the front panel. :INITiate:SP[1-1]:IMMediate Syntax :INITiate:SP[1-1]:IMMediate...
SCPI Command Reference :MMEMory:CATalog :MMEMory:CATalog Syntax :MMEMory:CATalog Description List file.folder name of the current directory (Query Only) Equivalent key No equivalent key is available on the front panel. :MMEMory:COPY Syntax :MMEMory:COPY <String 1>,<String 2> Description Copies file (No Query) Parameter <String 1>...
SCPI Command Reference :MMEMory:DELete <String 1> Unit Resolution block Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:DELete Syntax :MMEMory:DELete <String> Description Deletes file/directory (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :MMEMory:FP[1-1]:TRACe[1-4]:LOAD:LIMit:UPPer <String> Unit Resolution Equivalent key FP Menu -> Display -> Limit Test -> Import Lower Limit Line... :MMEMory:FP[1-1]:TRACe[1-4]:LOAD:LIMit:UPPer Syntax :MMEMory:FP[1-1]:TRACe[1-4]:LOAD:LIMit:UPPer <String> Description Reads the upper limit line (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key FP Menu ->...
SCPI Command Reference :MMEMory:LOAD:CORRection:POWer Description Saves memory trace data (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:LOAD:CORRection:POWer Syntax :MMEMory:LOAD:CORRection:POWer <String> Description Loads correction data for a specified power (No Query) Parameter <String>...
SCPI Command Reference :MMEMory:LOAD:PROGram :MMEMory:LOAD:PROGram Syntax :MMEMory:LOAD:PROGram <String> Description Loads VBA project/module (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:LOAD:STATe Syntax :MMEMory:LOAD:STATe <String> Description Recalls settings (No Query) Parameter <String>...
SCPI Command Reference :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer <String> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer Syntax :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer <String> Description Reads the lower limit line (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key PN Menu ->...
SCPI Command Reference :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:SPURious:THReshold :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:SPURious:THRes hold Syntax :MMEMory:PN[1-1]:TRACe[1-1]:LOAD:SPURious:THReshold <String> Description Reads the threshold data (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key PN Menu -> Trace View -> Spurious -> Import Threshold Table... :MMEMory:PN[1-1]:TRACe[1-1]:STORe[:DATA] Syntax :MMEMory:PN[1-1]:TRACe[1-1]:STORe[:DATA] <String> Description Saves trace data (No Query) Parameter...
SCPI Command Reference :MMEMory:SP[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer <String> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:SP[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer Syntax :MMEMory:SP[1-1]:TRACe[1-1]:LOAD:LIMit:LOWer <String> Description Reads the lower limit line (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key SP Menu ->...
SCPI Command Reference :MMEMory:SP[1-1]:TRACe[1-1]:STORe[:DATA] :MMEMory:SP[1-1]:TRACe[1-1]:STORe[:DATA] Syntax :MMEMory:SP[1-1]:TRACe[1-1]:STORe[:DATA] <String> Description Saves trace data (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:SP[1-1]:TRACe[1-1]:STORe:MEMory Syntax :MMEMory:SP[1-1]:TRACe[1-1]:STORe:MEMory <String> Description Saves memory trace data (No Query) Parameter <String>...
SCPI Command Reference :MMEMory:STORe:PROGram <String> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:STORe:PROGram Syntax :MMEMory:STORe:PROGram <String> Description Saves VBA project (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:STORe:STATe Syntax :MMEMory:STORe:STATe <String>...
SCPI Command Reference :MMEMory:STORe:STYPe :MMEMory:STORe:STYPe Syntax :MMEMory:STORe:STYPe {STAT|DST} :MMEMory:STORe:STYPe? Description Selects save state type Parameter Description STAT(Preset Set save state type to 'State Only' value) Set save state type to 'State & Data' Equivalent key PN Menu -> Save/Recall -> Save State -> Save Type SP Menu ->...
SCPI Command Reference :MMEMory:TR[1-1]:TRACe[1-4]:STORe[:DATA] Parameter <String> Range Preset value Unit Resolution Equivalent key TR Menu -> Display -> Limit Test -> Import Upper Limit Line... :MMEMory:TR[1-1]:TRACe[1-4]:STORe[:DATA] Syntax :MMEMory:TR[1-1]:TRACe[1-4]:STORe[:DATA] <String> Description Saves trace data (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key...
SCPI Command Reference :MMEMory:USER[1-1]:TRACe[1-8]:LOAD:LIMit:LOWer :MMEMory:USER[1-1]:TRACe[1-8]:LOAD:LIMit:LOWer Syntax :MMEMory:USER[1-1]:TRACe[1-8]:LOAD:LIMit:LOWer <String> Description Reads the lower limit line (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key USER Menu -> Display -> Limit Test -> Import Lower Limit Line... :MMEMory:USER[1-1]:TRACe[1-8]:LOAD:LIMit:UPPer Syntax :MMEMory:USER[1-1]:TRACe[1-8]:LOAD:LIMit:UPPer <String> Description Reads the upper limit line (No Query) Parameter...
SCPI Command Reference :MMEMory:USER[1-1]:TRACe[1-8]:STORe:MEMory <String> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :MMEMory:USER[1-1]:TRACe[1-8]:STORe:MEMory Syntax :MMEMory:USER[1-1]:TRACe[1-8]:STORe:MEMory <String> Description Saves selected memory trace data (No Query) Parameter <String> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :PROGram:SKEY:ITEM[1-8]:ENABle Equivalent key No equivalent key is available on the front panel. :PROGram:SKEY:ITEM[1-8]:ENABle Syntax :PROGram:SKEY:ITEM[1-8]:ENABle {ON|OFF|1|0} :PROGram:SKEY:ITEM[1-8]:ENABle? Description Turns on/off user defined softkey function Parameter Description ON or 1 Enable user defined softkey function OFF or 0(Preset Disable user defined softkey function value) Equivalent key...
SCPI Command Reference :PROGram:VARiable:ARRay[1-10]:DATA <String> Resolution Equivalent key No equivalent key is available on the front panel. :PROGram:VARiable:ARRay[1-10]:DATA Syntax :PROGram:VARiable:ARRay[1-10]:DATA <array> :PROGram:VARiable:ARRay[1-10]:DATA? Description User defined array data Parameter <Description> Range 1...1601 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :PROGram:VARiable:ARRay[1-10]:POINts Syntax :PROGram:VARiable:ARRay[1-10]:POINts <numeric>...
SCPI Command Reference :PROGram:VARiable:DOUBle[1-10] :PROGram:VARiable:DOUBle[1-10] Syntax :PROGram:VARiable:DOUBle[1-10] <numeric> :PROGram:VARiable:DOUBle[1-10]? Description User defined 64bit floating variable Parameter <Numeric> Range Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :PROGram:VARiable:INTeger[1-10] Syntax :PROGram:VARiable:INTeger[1-10] <numeric> :PROGram:VARiable:INTeger[1-10]? Description User defined integer variable Parameter <Numeric>...
:SENSe:ATTenuation:LEVel <numeric> :SENSe:ATTenuation:LEVel? Description Sets/Reads the input attenuator level Parameter When the E5052A is used stand-alone, or with the downconverter turned off, or with the downconverter on and the RF input is set to 'E5052A Direct' <Description> Range 0 to 35...
SCPI Command Reference :SENSe:CORRection:POWer:DATA :SENSe:CORRection:POWer:DATA Syntax :SENSe:CORRection:POWer:DATA <array> :SENSe:CORRection:POWer:DATA? Description Sets/Reads the frequency where the correction is performed and the correction values Parameter <Description> Range 1...2002 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :SENSe:CORRection:POWer:STATe Syntax :SENSe:CORRection:POWer:STATe {ON|OFF|1|0}...
DCONverter Sets the input port to ’Downconverter’ (Preset value) DIRect Sets the input port to ’E5052A Direct’ Equivalent key PN Menu -> Input Port -> Downconverter -> RF Input SP Menu -> Input Port -> Downconverter -> RF Input FP Menu -> Input Port -> Downconverter -> RF Input TR Menu ->...
SCPI Command Reference :SENSe:DCONverter:MANual:MEXTernal[1-2]:BIAS:CURRent :SENSe:DCONverter:MANual:MEXTernal[1-2]:BIAS:CUR Rent Syntax :SENSe:DCONverter:MANual:MEXTernal[1-2]:BIAS:CURRent <numeric> :SENSe:DCONverter:MANual:MEXTernal[1-2]:BIAS:CURRent? Description Sets/Reads the bias current to be supplied to the external mixer Parameter <Description> Range -10m to 10m Preset value Unit Resolution Equivalent key PN Menu -> System -> Instrument Setup -> Downconverter Manual Setup -> Current SP Menu ->...
SCPI Command Reference :SENSe:DCONverter:MEXTernal USER Menu -> System -> Instrument Setup -> Downconverter Manual Setup -> Mixer1 Bias :SENSe:DCONverter:MEXTernal Syntax :SENSe:DCONverter:MEXTernal {ON|OFF} :SENSe:DCONverter:MEXTernal? Description Sets the use of the external mixer on or off and reads its settings Parameter <Description> Use of the external mixer 'ON' OFF (Preset Use of the external mixer 'Not Used'...
SCPI Command Reference :SENSe:FP[1-1]:DCONverter:SSEarch:EXECute :SENSe:FP[1-1]:DCONverter:FREQuency? Description Sets/Reads information on the furequency supplied to the downconverter. Parameter When the frequency band is 3G to 10GHz <Description> Range 3G to 10G Preset value Unit Resolution 100m When the frequency band is 9G to 26.5GHz <Description>...
SCPI Command Reference :SENSe:FP[1-1]:FREQuency:RESolution Description HIGH(Preset Set frequency band to 'HIGH' value) Equivalent key FP Menu -> Setup -> Frequency Band :SENSe:FP[1-1]:FREQuency:RESolution Syntax :SENSe:FP[1-1]:FREQuency:RESolution {NARRow|MIDDle|WIDE} :SENSe:FP[1-1]:FREQuency:RESolution? Description Sets/Reads frequency resolution Parameter Description NARRow Set frequency resolution to 'NARRow' MIDDle Set frequency resolution to 'MIDDle' WIDE(Preset Set frequency resolution to 'WIDE' value)
SCPI Command Reference :SENSe:FP[1-1]:SWEep:DWELl :SENSe:FP[1-1]:SWEep:DWELl Syntax :SENSe:FP[1-1]:SWEep:DWELl <numeric> :SENSe:FP[1-1]:SWEep:DWELl? Description Sets/Reads the point delay value Parameter <Numeric> Range 0 to 1 Preset value Unit Resolution 100u Equivalent key FP Menu -> Setup -> Point Delay :SENSe:FP[1-1]:SWEep:TIME:DATA Syntax :SENSe:FP[1-1]:SWEep:TIME:DATA Description Reads the measurement time (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :SENSe:PN[1-1]:AVERage:STATe <Numeric> Preset value Unit Resolution Equivalent key PN Menu -> Average -> Avg Factor :SENSe:PN[1-1]:AVERage:STATe Syntax :SENSe:PN[1-1]:AVERage:STATe {ON|OFF|1|0} :SENSe:PN[1-1]:AVERage:STATe? Description Turns on/off averaging mode Parameter Description ON or 1 Enable averaging function OFF or 0(Preset Disable averaging function value) Equivalent key PN Menu ->...
SCPI Command Reference :SENSe:PN[1-1]:DCONverter:FREQuency :SENSe:PN[1-1]:DCONverter:FREQuency Syntax :SENSe:PN[1-1]:DCONverter:FREQuency <numeric> :SENSe:PN[1-1]:DCONverter:FREQuency? Description Sets/Reads the input frequecny to be supplied to the downconverter Parameter When the frequency band is 3G to 10GHz <Description> Range 3G to 10G Preset value Unit Resolution 100m When the frequency band is 9G to 26.5GHz <Description>...
Set frequency band to '99M - 1.5GHz' BAND4(Preset Set frequency band to '300M - 7GHz' value) When the downconverter is turned on and with the RF input is set to ’E5052A Direct’ <Description> BAND1 Set frequency band to '10M - 41MHz'...
SCPI Command Reference :SENSe:PN[1-1]:SWEep:POINts Description FAST Set the quality level to 'Fast' Equivalent key PN Menu -> Setup -> Measurement Quality :SENSe:PN[1-1]:SWEep:POINts Syntax :SENSe:PN[1-1]:SWEep:POINts Description Reads the number of measurement points (Query Only) Equivalent key No equivalent key is available on the front panel. :SENSe:ROSCillator:SOURce Syntax :SENSe:ROSCillator:SOURce...
Description Sets/Reads the carrier frequency band Parameter When the E5052A is used stand-alone, or with the downconverter turned off, or with the downconverter on and the RF input is set to 'E5052A Direct' Description Set the carrier frequency band to '10M - 1.5GHz'...
Sets/Reads the center value of frequency span Parameter When the E5052A is used stand-alone, or with the downconverter turned off (Allowable setting range varies depending on whether the frequency offset is on or off , the LO frequency within the frequency offset, and the conversion mode used in the...
Resolution 100m When the downconverter is turned on and the RF input is set to 'E5052A Direct'(Allowable setting range varies depending on whether the frequency offset is on or off , the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
Resolution 100m When the downconverter is turned on and the RF input is set to 'E5052A Direct' RF (Allowable setting range varies depending on whether the frequency offset is on or off , the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
Resolution 100m When the downconverter is turne on and the RF input is set to 'E5052A Direct' (Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
SCPI Command Reference :SENSe:SP[1-1]:POWer:RLEVel When the downconverter is turne on and the RF input is set to 'Downconverter' <Description> Range 3.00000005G to 26.5075G Preset value 3.0075G Unit Resolution 100m Equivalent key SP Menu -> Start/Center -> Stop SP Menu -> Stop/Span -> Stop SP Menu ->...
TR Menu -> Average -> Averaging :SENSe:TR[1-1]:NARRow:FREQuency:PREFerence Syntax :SENSe:TR[1-1]:NARRow:FREQuency:PREFerence <numeric> :SENSe:TR[1-1]:NARRow:FREQuency:PREFerence? Description Sets/Reads phase reference freuency Parameter When the E5052A is used stand-alone, or with the downconverter turned off (Allowable setting range varies depending on whether the frequency offset is on or...
Unit Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct'(Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
Unit Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct'(Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
SCPI Command Reference :SENSe:TR[1-1]:NARRow:SWEep:POINts When the downconverter is turne on and the RF input is set to 'Downconverter' <Description> Range 3G to 26.5G Preset value Unit Resolution 100m Equivalent key TR Menu -> Setup -> Target Freq TR Menu -> Marker To -> Marker -> Target Freq :SENSe:TR[1-1]:NARRow:SWEep:POINts Syntax :SENSe:TR[1-1]:NARRow:SWEep:POINts...
SCPI Command Reference :SENSe:TR[1-1]:NARRow:TIME:SPAN Description Selects reference position for time span of transient measurement Parameter Description LEFT Set reference position to 'LEFT' CENTer(Preset Set reference position to 'CENTer' value) RIGHt Set reference position to 'RIGHt' Equivalent key TR Menu -> Time Offset -> Narrow Ref Position TR Menu ->...
<Description> Range 160M to 117G Preset value 3.15G Unit Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct' with the frequency offset is off <Description> Range 150M|300M|600M|900M|1.2G|1.5G|1.8G|2.4G|3G Preset value...
Unit Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct' with the frequency offset is on (Allowable setting range varies depending on whether the LO frequency within the frequency offset, and the conversion mode used in the frequency offset) <Description>...
SCPI Command Reference :SENSe:TR[1-1]:WIDE:TIME:REFerence Parameter <Numeric> Range -80m to 1.1 Preset value Unit Resolution Equivalent key TR Menu -> Time Offset -> Wide Time Offset TR Menu -> Span -> Wide Time Offset :SENSe:TR[1-1]:WIDE:TIME:REFerence Syntax :SENSe:TR[1-1]:WIDE:TIME:REFerence {LEFT|CENTer|RIGHt} :SENSe:TR[1-1]:WIDE:TIME:REFerence? Description Selects reference position for time span Parameter Description LEFT...
SCPI Command Reference :SENSe:UDConverter:HARMonic <Numeric> Unit Resolution Equivalent key TR Menu -> Time Offset -> Wide Span TR Menu -> Span -> Wide Span :SENSe:UDConverter:HARMonic Syntax :SENSe:UDConverter:HARMonic <numeric> :SENSe:UDConverter:HARMonic? Description Sets/Reads the factor of the frequency offset Parameter <Description> Range 1 to 20 Preset value Unit...
SCPI Command Reference :SENSe:UDConverter:MODE setting range varies depending on the set value of Harmonic) <Description> Range 10M to 110G Preset value Unit Resolution 100m When the conversion mode used in the frequency offset is RF=N*LO+IF (Allowable setting range varies depending on the set value of Harmonic) <Description>...
SCPI Command Reference :SOURce:FP[1-1]:SWEep:POINts Parameter Description CONTrol(Preset Set sweep parameter to 'CONTrol' value) POWer Set sweep parameter to 'POWer' Equivalent key FP Menu -> Setup -> Sweep Parameter :SOURce:FP[1-1]:SWEep:POINts Syntax :SOURce:FP[1-1]:SWEep:POINts <numeric> :SOURce:FP[1-1]:SWEep:POINts? Description Sets/Reads the number of measurement points Parameter <Numeric>...
SCPI Command Reference :SOURce:FP[1-1]:VOLTage:CONTrol:SPAN FP Menu -> Stop/Span -> DC Control Center :SOURce:FP[1-1]:VOLTage:CONTrol:SPAN Syntax :SOURce:FP[1-1]:VOLTage:CONTrol:SPAN <numeric> :SOURce:FP[1-1]:VOLTage:CONTrol:SPAN? Description Vcontrol span Parameter <Numeric> Range 0 to 50 Preset value 100u Unit Resolution 100u Equivalent key FP Menu -> Start/Center -> DC Control Span FP Menu ->...
SCPI Command Reference :SOURce:FP[1-1]:VOLTage:POWer:CENTer Description Vcontrol stop Parameter <Numeric> Range -15 to 35 Preset value 100u Unit Resolution 100u Equivalent key FP Menu -> Start/Center -> DC Control Stop FP Menu -> Stop/Span -> DC Control Stop :SOURce:FP[1-1]:VOLTage:POWer:CENTer Syntax :SOURce:FP[1-1]:VOLTage:POWer:CENTer <numeric> :SOURce:FP[1-1]:VOLTage:POWer:CENTer? Description Vpower center...
SCPI Command Reference :SOURce:FP[1-1]:VOLTage:POWer:STARt Parameter <Numeric> Range 0 to 16 Preset value Unit Resolution Equivalent key FP Menu -> Start/Center -> DC Power Span FP Menu -> Stop/Span -> DC Power Span :SOURce:FP[1-1]:VOLTage:POWer:STARt Syntax :SOURce:FP[1-1]:VOLTage:POWer:STARt <numeric> :SOURce:FP[1-1]:VOLTage:POWer:STARt? Description Vpower start Parameter <Numeric>...
Set the frequency band in the auto frequency control function to value) '300M-7GHz' When the downconverter is turned on and the RF input is set to 'E5052A Direct' <Description> Set the frequency band in the auto frequency control function to '10M-1.5GHz'...
SCPI Command Reference :SOURce:VOLTage:CONTrol:AFC:IMMediate USER Menu -> DC Control Voltage -> Auto Freq Control -> Freqency Band :SOURce:VOLTage:CONTrol:AFC:IMMediate Syntax :SOURce:VOLTage:CONTrol:AFC:IMMediate Description Executes the auto frequency control once. (No Query) Equivalent key PN Menu -> DC Control Voltage -> Auto Freq Control -> AFC Status SP Menu ->...
SCPI Command Reference :SOURce:VOLTage:CONTrol:AFC:LIMit:HIGH Description Sets/Reads the maximum number of iterations for the DC control voltage-setting loops Parameter <Numeric> Range 1 to 99 Preset value Unit Resolution Equivalent key PN Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration SP Menu ->...
SCPI Command Reference :SOURce:VOLTage:CONTrol:AFC:SENSitivity :SOURce:VOLTage:CONTrol:AFC:LIMit:LOW? Description Sets/Reads the minimum DC control voltage limit Parameter <Numeric> Range -15 to 35 Preset value Unit Resolution 100u Equivalent key PN Menu -> DC Control Voltage -> Auto Freq Control -> Min Ctrl Voltage Limit SP Menu ->...
Resolution 100m When the downconverter is turned on and the RF input is set to 'E5052A Direct' (Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, and the conversion mode used in...
SCPI Command Reference :SOURce:VOLTage:CONTrol:AFC:TOLerance the frequency offset) <Description> Range 10M to 113G Preset value Unit Resolution 100m When the downconverter is turned on and the RF input is set to 'Downconverter' <Description> Range 3G to 26.5G Preset value Unit Resolution 100m Equivalent key PN Menu ->...
SCPI Command Reference :SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQuire TR Menu -> DC Control Voltage -> Auto Freq Control -> Tolerance USER Menu -> DC Control Voltage -> Auto Freq Control -> Tolerance :SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQ uire Syntax :SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQuire Description Executes DC CTRL DRIFT CAL (No Query) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :SOURce:VOLTage:CONTrol:LEVel:AMPLitude SP Menu -> DC Control Voltage -> DC Control Delay FP Menu -> DC Control Voltage -> DC Control Delay TR Menu -> DC Control Voltage -> DC Control Delay USER Menu -> DC Control Voltage -> DC Control Delay :SOURce:VOLTage:CONTrol:LEVel:AMPLitude Syntax :SOURce:VOLTage:CONTrol:LEVel:AMPLitude <numeric>...
SCPI Command Reference :SOURce:VOLTage:CONTrol:LIMit:HIGH SP Menu -> DC Control Voltage -> DC Control Output FP Menu -> DC Control Voltage -> DC Control Output TR Menu -> DC Control Voltage -> DC Control Output USER Menu -> DC Control Voltage -> DC Control Output :SOURce:VOLTage:CONTrol:LIMit:HIGH Syntax :SOURce:VOLTage:CONTrol:LIMit:HIGH <numeric>...
SCPI Command Reference :SOURce:VOLTage:POWer:DELay Equivalent key PN Menu -> DC Control Voltage -> Min Ctrl Voltage Limit SP Menu -> DC Control Voltage -> Min Ctrl Voltage Limit FP Menu -> DC Control Voltage -> Min Ctrl Voltage Limit TR Menu -> DC Control Voltage -> Min Ctrl Voltage Limit USER Menu ->...
SCPI Command Reference :SOURce:VOLTage:POWer:LEVel:STATe <Numeric> Resolution Equivalent key PN Menu -> DC Power Voltage -> DC Power Voltage SP Menu -> DC Power Voltage -> DC Power Voltage FP Menu -> DC Power Voltage -> DC Power Voltage TR Menu -> DC Power Voltage -> DC Power Voltage USER Menu ->...
SCPI Command Reference :SOURce:VOLTage:POWer:LIMit:LOW <Numeric> Unit Resolution Equivalent key PN Menu -> DC Power Voltage -> Max Pwr Voltage Limit SP Menu -> DC Power Voltage -> Max Pwr Voltage Limit FP Menu -> DC Power Voltage -> Max Pwr Voltage Limit TR Menu ->...
SCPI Command Reference :STATus:OPERation:BIT12:CONDition Parameter <Numeric> Range 0 to 14 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:BIT12:CONDition Syntax :STATus:OPERation:BIT12:CONDition Description Reads operation-program status register (Query Only) Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:BIT12:ENABle Syntax :STATus:OPERation:BIT12:ENABle <numeric>...
SCPI Command Reference :STATus:OPERation:BIT12:NTRansition :STATus:OPERation:BIT12:NTRansition Syntax :STATus:OPERation:BIT12:NTRansition <numeric> :STATus:OPERation:BIT12:NTRansition? Description Sets/Reads operation-program status negative transition filter value Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:BIT12:PTRansition Syntax :STATus:OPERation:BIT12:PTRansition <numeric>...
SCPI Command Reference :STATus:OPERation:CONDition Parameter <Numeric> Range 0 to 14 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:CONDition Syntax :STATus:OPERation:CONDition Description Reads operation status conditional register value (Query Only) Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:ENABle Syntax :STATus:OPERation:ENABle <numeric>...
SCPI Command Reference :STATus:OPERation:NTRansition :STATus:OPERation:NTRansition Syntax :STATus:OPERation:NTRansition <numeric> :STATus:OPERation:NTRansition? Description Sets/Reads operation status negative transition filter value Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:OPERation:PTRansition Syntax :STATus:OPERation:PTRansition <numeric>...
SCPI Command Reference :STATus:QUEStionable:CONDition :STATus:QUEStionable:CONDition Syntax :STATus:QUEStionable:CONDition Description Reads questionable status conditional register value (Query Only) Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:CURRent:ENABle Syntax :STATus:QUEStionable:CURRent:ENABle <numeric> :STATus:QUEStionable:CURRent:ENABle? Description Sets/Reads questionable-current status enable register Parameter <Numeric> Range 0 to 65535 Preset value...
SCPI Command Reference :STATus:QUEStionable:DCONverter[:EVENt] <Description> Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:DCONverter[:EVENt] Syntax :STATus:QUEStionable:DCONverter[:EVENt] Description Reads questionable downconverter status event register value (Query Only) Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:ENABle Syntax :STATus:QUEStionable:ENABle <numeric>...
SCPI Command Reference :STATus:QUEStionable:LIMit:ENABle Description Reads the questionable limit conditional register value (Query Only) Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:ENABle Syntax :STATus:QUEStionable:LIMit:ENABle <numeric> :STATus:QUEStionable:LIMit:ENABle? Description Sets/Reads the questionable limit status enable value Parameter <Numeric> Range 0 to 65535 Preset value...
SCPI Command Reference :STATus:QUEStionable:LIMit:FP[1-1][:EVENt] Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:FP[1-1][:EVENt] Syntax :STATus:QUEStionable:LIMit:FP[1-1][:EVENt] Description Reads the questionable limit FP status event register value (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :STATus:QUEStionable:LIMit:NTRansition Parameter <Numeric> Range 0 to 65535 Preset value 32767 Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:NTRansition Syntax :STATus:QUEStionable:LIMit:NTRansition <numeric> :STATus:QUEStionable:LIMit:NTRansition? Description Sets/Reads the questionable limit status negative transition filter value Parameter <Numeric>...
SCPI Command Reference :STATus:QUEStionable:LIMit:PN[1-1][:EVENt] Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:PN[1-1][:EVENt] Syntax :STATus:QUEStionable:LIMit:PN[1-1][:EVENt] Description Reads the questionable limit PN status event register value (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :STATus:QUEStionable:LIMit:PTRansition Parameter <Numeric> Range 0 to 65535 Preset value 32767 Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:PTRansition Syntax :STATus:QUEStionable:LIMit:PTRansition <numeric> :STATus:QUEStionable:LIMit:PTRansition? Description Sets/Reads the questionable limit status positive transition filter value Parameter <Numeric>...
SCPI Command Reference :STATus:QUEStionable:LIMit:SP[1-1][:EVENt] Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:SP[1-1][:EVENt] Syntax :STATus:QUEStionable:LIMit:SP[1-1][:EVENt] Description Reads the questionable limit SP status event register value (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :STATus:QUEStionable:LIMit:TR[1-1]:CONDition Parameter <Numeric> Range 0 to 65535 Preset value 32767 Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:TR[1-1]:CONDition Syntax :STATus:QUEStionable:LIMit:TR[1-1]:CONDition Description Reads the questionable limit TR conditional register value (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :STATus:QUEStionable:LIMit:TR[1-1]:NTRansition :STATus:QUEStionable:LIMit:TR[1-1]:NTRansition Syntax :STATus:QUEStionable:LIMit:TR[1-1]:NTRansition <numeric> :STATus:QUEStionable:LIMit:TR[1-1]:NTRansition? Description Sets/Reads the questionable limit TR status negative transition filter value Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:TR[1-1]:PTRansition Syntax :STATus:QUEStionable:LIMit:TR[1-1]:PTRansition <numeric>...
SCPI Command Reference :STATus:QUEStionable:LIMit:USER[1-1]:ENABle :STATus:QUEStionable:LIMit:USER[1-1]:ENABle Syntax :STATus:QUEStionable:LIMit:USER[1-1]:ENABle <numeric> :STATus:QUEStionable:LIMit:USER[1-1]:ENABle? Description Sets/Reads the questionable limit USER1 status enable value Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:LIMit:USER[1-1][:EVENt] Syntax :STATus:QUEStionable:LIMit:USER[1-1][:EVENt] Description...
SCPI Command Reference :STATus:QUEStionable:LIMit:USER[1-1]:PTRansition :STATus:QUEStionable:LIMit:USER[1-1]:PTRansition Syntax :STATus:QUEStionable:LIMit:USER[1-1]:PTRansition <numeric> :STATus:QUEStionable:LIMit:USER[1-1]:PTRansition? Description Sets/Reads the questionable limit USER1 status positive transition filter value Parameter <Numeric> Range 0 to 65535 Preset value 32767 Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:MISC:ENABle Syntax :STATus:QUEStionable:MISC:ENABle <numeric>...
SCPI Command Reference :STATus:QUEStionable:NTRansition :STATus:QUEStionable:NTRansition Syntax :STATus:QUEStionable:NTRansition <numeric> :STATus:QUEStionable:NTRansition? Description Sets/Reads questionable status negative transition filter value Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:PHASe:ENABle Syntax :STATus:QUEStionable:PHASe:ENABle <numeric>...
SCPI Command Reference :STATus:QUEStionable:POWer:ENABle :STATus:QUEStionable:POWer:ENABle Syntax :STATus:QUEStionable:POWer:ENABle <numeric> :STATus:QUEStionable:POWer:ENABle? Description Sets/Reads questionable-power status enable register Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:POWer[:EVENt] Syntax :STATus:QUEStionable:POWer[:EVENt] Description Reads questionable-power status event register value (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :STATus:QUEStionable:REFerence:ENABle :STATus:QUEStionable:REFerence:ENABle Syntax :STATus:QUEStionable:REFerence:ENABle <numeric> :STATus:QUEStionable:REFerence:ENABle? Description Sets/Reads questionable-reference signal status enable register Parameter <Numeric> Range 0 to 65535 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :STATus:QUEStionable:REFerence[:EVENt] Syntax :STATus:QUEStionable:REFerence[:EVENt] Description Reads questionable-reference signal status event register value (Query Only) Equivalent key...
SCPI Command Reference :SYSTem:BEEPer:COMPlete:IMMediate USER Menu -> System -> Backlight :SYSTem:BEEPer:COMPlete:IMMediate Syntax :SYSTem:BEEPer:COMPlete:IMMediate Description Makes beep sound for operation completion (No Query) Equivalent key PN Menu -> System -> Misc Setup -> Beeper -> Test Beep Complete SP Menu -> System -> Misc Setup -> Beeper -> Test Beep Complete FP Menu ->...
SCPI Command Reference :SYSTem:BEEPer:WARNing:STATe USER Menu -> System -> Misc Setup -> Beeper -> Test Beep Warning :SYSTem:BEEPer:WARNing:STATe Syntax :SYSTem:BEEPer:WARNing:STATe {ON|OFF|1|0} :SYSTem:BEEPer:WARNing:STATe? Description Turns on/off the beep for warning Parameter Description ON or 1 Set the beep for warning mode to 'ON' OFF or 0(Preset Set the beep for warning mode to 'OFF' value)
SCPI Command Reference :SYSTem:ERRor[:NEXT] <Numeric 2> Unit Resolution <Numeric 3> Range 1 to 31 Preset value Unit Resolution Equivalent key No equivalent key is available on the front panel. :SYSTem:ERRor[:NEXT] Syntax :SYSTem:ERRor[:NEXT] Description Gets error code & description (Query Only) Equivalent key No equivalent key is available on the front panel.
SCPI Command Reference :SYSTem:POFF Parameter Description ON or 1 Set touch screen and mouse lock state mode to 'ON' OFF or 0(Preset Set touch screen and mouse lock state mode to 'OFF' value) Equivalent key No equivalent key is available on the front panel. :SYSTem:POFF Syntax :SYSTem:POFF...
SCPI Command Reference :SYSTem:TIME USER Menu -> Display -> Security Level :SYSTem:TIME Syntax :SYSTem:TIME <numeric 1>,<numeric 2>,<numeric 3> :SYSTem:TIME? Description Sets/Gets system time Parameter <Numeric 1> Range 0 to 23 Preset value Unit Resolution <Numeric 2> Range 0 to 59 Preset value Unit Resolution...
Parameter When the E5052A is used stand-alone, or with the downconverter turned off (Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, the conversion mode used in the frequency offset, the target frequency and frequency range of the narrow band mode) <Numeric>...
<Numeric> Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct' (Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, the conversion mode used in the frequency offset, the target frequency and frequency range of the narrowband mode) <Description>...
Unit Resolution When the downconverter is turned on and the RF input is set to 'E5052A Direct' (Allowable setting range varies depending on whether the frequency offset is on or off, the LO frequency within the frequency offset, the conversion mode used in the...
Page 376
SCPI Command Reference :TRIGger:TR[1-1]:WIDE:VIDeo:FREQuency:CENTer frequency offset, and the transient frequency range of the wideband mode) <Description> Range 50M to 113G Preset value Unit Resolution 100m When the downconverter is turned on and the RF input is set to 'Downconverter' (Allowable setting range varies depending on the transient frequency range of the wideband mode) <Description>...
SCPI Command Reference List by function Command list List by function Bellow table shows the SCPI command list by function. Function Setting/Execution item Command 24Bit I/O control Outputs data using port A :CONT:HAND:A[:DATA] Outputs data using port B :CONT:HAND:B[:DATA] Inputs/Outputs data using port C :CONT:HAND:C[:DATA] Selects input/output mode on port C :CONT:HAND:C:MODE...
Page 378
SCPI Command Reference List by function Function Setting/Execution item Command DC sources Executes DC CTRL DRIFT CAL :SOUR:VOLT:CONT:CORR:COLL:ACQ DC CTRL DRIFT CAL state :SOUR:VOLT:CONT:CORR[:STAT] DC Control delay setting(sec) :SOUR:VOLT:CONT:DEL DC sources Sets/Reads DC Control voltage :SOUR:VOLT:CONT:LEV:AMPL (Continued) Turns on/off DC Control voltage output :SOUR:VOLT:CONT:LEV:STAT Sets/Reads maximum voltage limit for DC Control :SOUR:VOLT:CONT:LIM:HIGH...
Page 379
SCPI Command Reference List by function Function Setting/Execution item Command File operation Catalog directory :MMEM:CAT Coies file :MMEM:COPY File transfer through SCPI :MMEM:DATA Deletes file/directory :MMEM:DEL Loads program :MMEM:LOAD:PROG Recalls settings :MMEM:LOAD:STAT Creates a directory :MMEM:MDIR Saves screen image :MMEM:STOR:IMAG Saves VBA project :MMEM:STOR:PROG Saves settings...
Page 380
SCPI Command Reference List by function Function Setting/Execution item Command Frequency, RF Cleasr persistence mode :DISP:FP[1-1]:TRAC[1-4]:PERS:CLE power and DC Sets/Reads persistence mode :DISP:FP[1-1]:TRAC[1-4]:PERS:STAT current measurement - Display Executes autoscale :DISP:FP[1-1]:TRAC[1-4]:Y[:SCAL]:AUTO (Continued) Sets/Reads sclae per division :DISP:FP[1-1]:TRAC[1-4]:Y[:SCAL]:PDIV Sets/Reads scale reference level :DISP:FP[1-1]:TRAC[1-4]:Y[:SCAL]:RLEV Sets/Reads scale reference position :DISP:FP[1-1]:TRAC[1-4]:Y[:SCAL]:RPOS...
Page 381
SCPI Command Reference List by function Function Setting/Execution item Command Frequency, RF Turns on/off the limit line :DISP:FP[1-1]:TRAC[1-4]:LIM:LINE power and DC Turns on/off the limit test judgement display :DISP:FP[1-1]:LIM:FSIG current measurement - Limit Test Reads the upper limit line :MMEM:FP[1-1]:TRAC[1-4]:LOAD:LIM:UPP (Continued) Reads the lower limit line :MMEM:FP[1-1]:TRAC[1-4]:LOAD:LIM:LOW...
Page 382
SCPI Command Reference List by function Function Setting/Execution item Command Frequency, RF Executes marker search minimum :CALC:FP[1-1]:TRAC[1-4]:MARK[1-6]:SEAR:EXEC: power and DC current measurement Executes marker peak search :CALC:FP[1-1]:TRAC[1-4]:MARK[1-6]:SEAR:EXEC:P - Marker/analysis (Continued) Executes marker peak search right :CALC:FP[1-1]:TRAC[1-4]:MARK[1-6]:SEAR:EXEC:R Executes marker target search right :CALC:FP[1-1]:TRAC[1-4]:MARK[1-6]:SEAR:EXEC:R Executes marker target search :CALC:FP[1-1]:TRAC[1-4]:MARK[1-6]:SEAR:EXEC:T...
Page 383
SCPI Command Reference List by function Function Setting/Execution item Command Frequency, RF Sets/Reads the center value of DC Control voltage :SOUR:FP[1-1]:VOLT:CONT:CENT power and DC Sets/Reads the span value of DC Control voltage :SOUR:FP[1-1]:VOLT:CONT:SPAN current measurement - Measurement Sets/Reads the start value of DC Control volage :SOUR:FP[1-1]:VOLT:CONT:STAR conditions (Continued)
Page 384
SCPI Command Reference List by function Function Setting/Execution item Command Others (Continued) Powers off the instrument :SYST:POFF Presets instrument :SYST:PRES Phase noise Sets/Reads data hold type :CALC:PN[1-1]:TRAC[1-1]:HOLD measurement - Selects math operation type :CALC:PN[1-1]:TRAC[1-1]:MATH:FUNC Display Copies data to memory :CALC:PN[1-1]:TRAC[1-1]:MATH:MEM Sets/Reads the smoothing aperture value :CALC:PN[1-1]:TRAC[1-1]:SMO:APER Turns on/off smoothing...
Page 385
SCPI Command Reference List by function Function Setting/Execution item Command Phase noise Reads out the limit test result :CALC:PN[1-1]:ALLT:LIM:FAIL measurement - Limit Turns on/off the limit test function :CALC:PN[1-1]:TRAC[1-1]:LIM[:STAT] Test Sets/Reads the number of segments in the upper limit :CALC:PN[1-1]:TRAC[1-1]:LIM:UPP:SEGM:COUN line Sets/Reads the number of segments in the lower limit :CALC:PN[1-1]:TRAC[1-1]:LIM:LOW:SEGM:COUN...
Page 386
SCPI Command Reference List by function Function Setting/Execution item Command Phase noise Sets/Reads the start value of bandmarker Y :CALC:PN[1-1]:TRAC[1-1]:BDM:Y:STAR measurement - Turns on/off bandmarker Y :CALC:PN[1-1]:TRAC[1-1]:BDM:Y:STAT Marker/analysis (Continued) Sets/Reads the stop value of bandmarker Y :CALC:PN[1-1]:TRAC[1-1]:BDM:Y:STOP Sets/Reads analysis/search range (X-axis) :CALC:PN[1-1]:TRAC[1-1]:FUNC:DOM:X Sets/Reads analysis/search range (Y-axis) :CALC:PN[1-1]:TRAC[1-1]:FUNC:DOM:Y...
Page 387
SCPI Command Reference List by function Function Setting/Execution item Command Phase noise Turns on/off the marker list :DISP:PN[1-1]:TABL[:STAT] measurement - Reads the integrated phase noise, frequency range, :CALC:PN[1-1]:TRAC[1-1]:FUNC:INT:DATA Marker/analysis RMS noise, RMS jitter, and residual FM of trace (Continued) data. Reads the integrated phase noise, frequency range, :CALC:PN[1-1]:TRAC[1-1]:FUNC:INT:MEM RMS noise, RMS jitter, and residual FM of memory...
Page 388
SCPI Command Reference List by function Function Setting/Execution item Command Phase noise Turns on/off the spurious power value display :CALC:PN[1-1]:TRAC[1-1]:SPUR:POW measurement - Sets/Reads the raw power data (dBc) :CALC:PN[1-1]:DATA:PDAT Spurious display Sets/Reads unformatted trace power data (dBc) :CALC:PN[1-1]:TRAC[1-1]:DATA:PDAT Sets/Reads unformatted memory power data (dBc) :CALC:PN[1-1]:TRAC[1-1]:DATA:PMEM Reads the spurious judgement results (0/1) of trace :CALC:PN[1-1]:TRAC[1-1]:DATA:SDAT...
Page 389
SCPI Command Reference List by function Function Setting/Execution item Command Spectrum monitor - Turns on/off relative Y-scale :DISP:SP[1-1]:GRAT:AXIS:Y:REL Display (Continued) Shows/Hides Y graticule label :DISP:SP[1-1]:GRAT:AXIS:Y:STAT Edits window title label :DISP:SP[1-1]:LAB:DATA Shows/Hides Window Title Label :DISP:SP[1-1]:LAB:STAT Maximizes active trace :DISP:SP[1-1]:MAX Turns on/off spectrum monitor mode :DISP:SP[1-1]:STAT Edits trance title label :DISP:SP[1-1]:TRAC[1-1]:LAB:DATA...
Page 390
SCPI Command Reference List by function Function Setting/Execution item Command Spectrum monitor - Reads the limit test results of all measurement points :CALC:SP[1-1]:TRAC[1-1]:LIM:REP[:DATA] Limit Test in selected traces (Continued) Turns on/off the limit line :DISP:SP[1-1]:TRAC[1-1]:LIM:LINE Turns on/off the limit test judgement display :DISP:SP[1-1]:LIM:FSIG Reads the upper limit line :MMEM:SP[1-1]:TRAC[1-1]:LOAD:LIM:UPP...
Page 391
SCPI Command Reference List by function Function Setting/Execution item Command Spectrum monitor - Executes marker search minimum :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:EXEC: Marker/Analysis (Continued) Executes marker peak search :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:EXEC:P Executes marker peak search right :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:EXEC:R Executes marker target search right :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:EXEC:R Executes marker target search :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:EXEC:T Sets/Reads the peak excursion value :CALC:SP[1-1]:TRAC[1-1]:MARK[1-6]:SEAR:PEAK:E...
Page 392
SCPI Command Reference List by function Function Setting/Execution item Command Spectrum monitor - Sets/Reads the reference level of frequency span :SENS:SP[1-1]:POW:RLEV Measurement conditions (Continued) Spectrum monitor - Sets/Reads the raw data :CALC:SP[1-1]:DATA:RDAT Reads/writes the Reads X-axis data :CALC:SP[1-1]:DATA:XDAT data Sets/Reads formatted trace data :CALC:SP[1-1]:TRAC[1-1]:DATA:FDAT Sets/Reads formatted memory data :CALC:SP[1-1]:TRAC[1-1]:DATA:FMEM...
Page 393
SCPI Command Reference List by function Function Setting/Execution item Command Status report system Reads questionable-current status event register :STAT:QUES:CURR[:EVEN] (Continued) value Sets/Reads questionable downconverter status enable :STAT:QUES:DCON:ENAB register value Reads questionable downconverter status event :STAT:QUES:DCON[:EVEN] register value Sets/Reads questionable status enable register :STAT:QUES:ENAB Reads questionable status event register value :STAT:QUES[:EVEN]...
Page 394
SCPI Command Reference List by function Function Setting/Execution item Command Status report system Reads the questionable limit conditional register :STAT:QUES:LIM:PN[1-1]:COND (Continued) value (PN) Sets/Reads the questionable limit status enable value :STAT:QUES:LIM:PN[1-1]:ENAB (PN) Reads the questionable limit status event register :STAT:QUES:LIM:PN[1-1][:EVEN] value (PN) Sets/Reads the questionable limit status positive :STAT:QUES:LIM:PN[1-1]:PTR...
Page 395
SCPI Command Reference List by function Function Setting/Execution item Command Transient Selects active trace :CALC:TR[1-1]:ALLT:ACT measurement - Selects phase format on transient measurement :CALC:TR[1-1]:TRAC[1-4]:FORM:PHAS:UNIT Display Turns on/off wrap-phase :CALC:TR[1-1]:TRAC[1-4]:FORM:PHAS:WRAP Sets/Reads data hold :CALC:TR[1-1]:TRAC[1-4]:HOLD Selects math operation type :CALC:TR[1-1]:TRAC[1-4]:MATH:FUNC Copies data to memory :CALC:TR[1-1]:TRAC[1-4]:MATH:MEM Sets/Reads the smoothing aperture value :CALC:TR[1-1]:TRAC[1-4]:SMO:APER...
Page 396
SCPI Command Reference List by function Function Setting/Execution item Command Transient Reads out the limit test result :CALC:TR[1-1]:ALLT:LIM:FAIL measurement - Limit Turns on/off the limit test function :CALC:TR[1-1]:TRAC[1-4]:LIM[:STAT] Test Sets/Reads the number of segments in the upper limit :CALC:TR[1-1]:TRAC[1-4]:LIM:UPP:SEGM:COUN line Sets/Reads the number of segments in the lower limit :CALC:TR[1-1]:TRAC[1-4]:LIM:LOW:SEGM:COUN line...
Page 397
SCPI Command Reference List by function Function Setting/Execution item Command Transient Sets/Reads the span value of bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BDM:Y:SPAN measurement - Sets/Reads the start value of bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BDM:Y:STAR Marker/analysis (Continued) Turns on/off bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BDM:Y:STAT Sets/Reads the stop value of bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BDM:Y:STOP Sets/Reads analysis/search range (X-axis) :CALC:TR[1-1]:TRAC[1-4]:FUNC:DOM:X...
Page 398
SCPI Command Reference List by function Function Setting/Execution item Command Transient Sets/Reads the marker information position :DISP:TR[1-1]:ANN:MARK:POS measurement - Turns on/off the marker list :DISP:TR[1-1]:TABL[:STAT] Marker/analysis (Continued) Transient Triggers continuous :INIT:TR[1-1]:CONT measurement - Triggers once :INIT:TR[1-1]:IMM Measurement Transient Clears averaging :SENS:TR[1-1]:AVER:CLE measurement - Sets/Reads averaging count...
Page 399
SCPI Command Reference List by function Function Setting/Execution item Command Transient Sets/Reads X axis data :CALC:TR[1-1]:WIDE:DATA:XDAT measurement - Reads the number of measurement points :SENS:TR[1-1]:NARR:SWE:POIN Reads/writes the data (Continued) Reads the nnumber of measurement points :SENS:TR[1-1]:WIDE:SWE:POIN User defined Selects active trace :CALC:USER[1-1]:ALLT:ACT window - Display Selects math operation type...
Page 400
SCPI Command Reference List by function Function Setting/Execution item Command User defined Reads out the limit test result :CALC:USER[1-1]:ALLT:LIM:FAIL window - Limit Test Turns on/off the limit test function :CALC:USER[1-1]:TRAC[1-8]:LIM[:STAT] Sets/Reads the number of segments in the upper limit :CALC:USER[1-1]:TRAC[1-8]:LIM:UPP:SEGM:COUN line Sets/Reads the number of segments in the lower limit :CALC:USER[1-1]:TRAC[1-8]:LIM:LOW:SEGM:COU...
Page 401
SCPI Command Reference List by function Function Setting/Execution item Command User defined bandmarker Y center :CALC:USER[1-1]:TRAC[1-8]:BDM:Y:CENT window - bandmarker Y span :CALC:USER[1-1]:TRAC[1-8]:BDM:Y:SPAN Marker/analysis (Continued) bandmarker Y start :CALC:USER[1-1]:TRAC[1-8]:BDM:Y:STAR Turns on/off bandmarker Y :CALC:USER[1-1]:TRAC[1-8]:BDM:Y:STAT bandmarker Y stop :CALC:USER[1-1]:TRAC[1-8]:BDM:Y:STOP Sets/Reads analysis range (X-axis) :CALC:USER[1-1]:TRAC[1-8]:FUNC:DOM:X Sets/Reads analysis range(Y-axis) :CALC:USER[1-1]:TRAC[1-8]:FUNC:DOM:Y...
Page 402
SCPI Command Reference List by function Function Setting/Execution item Command User defined marker y position :CALC:USER[1-1]:TRAC[1-8]:MARK[1-6]:Y window - Sets/Reads the marker information position :DISP:USER[1-1]:ANN:MARK:POS Marker/analysis (Continued) Turns on/off the marker list :DISP:USER[1-1]:TABL[:STAT] User defined Sets/Reads formatted trace data :CALC:USER[1-1]:TRAC[1-8]:DATA:FDAT window - Sets/Reads formatted momory data :CALC:USER[1-1]:TRAC[1-8]:DATA:FMEM Reads/writes the...
SCPI Command Reference Commands with Variable Parameters and/or Setting Ranges Depending on Device Configurataion Commands with Variable Parameters and/or Setting Ranges Depending on Device Configurataion The table below lists the commands that have variable parameters and/or setting ranges depending on the device configuration. Table 7-1 SCPI Commands with Variable Parameters and/or Setting Ranges Depending on Device Configuration Command...
Page 404
SCPI Command Reference Commands with Variable Parameters and/or Setting Ranges Depending on Device Configurataion Table 7-1 SCPI Commands with Variable Parameters and/or Setting Ranges Depending on Device Configuration Command Description :SENSe:TR[1-1]:WIDE:FREQuency:MAXimum Sets/Reads transient frequency range in the wideband mode :SENSe:UDConverter:HARMonic Sets/Reads the frequency offset factor :SENSe:UDConverter:LO Sets/Reads the LO frequency of the frequency offset...
SCPI Command Reference Softkey Functions Softkey Functions Bellow table shows the SCPI command list by measurement window and softkey. FP Menu Table 7-2 FP Menu Key Operation Function SCPI Command Attenuator Sets/Reads Input Attenuator level on 5dB Step :SENS:ATT:LEV Input Attenuator Average Turns on/off averaging function :SENS:FP[1-1]:AVER:STAT...
Page 406
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Sets/Reads the minimum DC Control voltage limit :SOUR:VOLT:CONT:LIM:LOW Min Ctrl Voltage Limit DC Power Voltage Sets/Reads DC Power delay (sec) :SOUR:VOLT:POW:DEL DC Power Delay Turns on/off DC Power voltage :SOUR:VOLT:POW:LEV:STAT DC Power Output DC Power Voltage...
Page 407
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Format Selects frequency format (Hz or Hz/V) :CALC:FP[1-1]:TRAC[1-4]:FOR Frequency Format M:FREQ Sets/Reads the frequency reference. :CALC:FP[1-1]:TRAC[1-4]:REF: Frequency FREQ Reference Sensitivity Aperture Sets/Reads the sensitivity aperture :CALC:FP[1-1]:TRAC[1-4]:SAP Input Port Downconverter Downconverter...
Page 408
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Saves VBA Project :MMEM:STOR:PROG Save Project Marker Clear Marker Menu Clears all the markers All OFF Turns on/off marker 1 :CALC:FP[1-1]:TRAC[1-4]:MAR Marker 1 K[1-6]:STAT Turns on/off marker 2 :CALC:FP[1-1]:TRAC[1-4]:MAR Marker 2 K[1-6]:STAT...
Page 409
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Marker Function Sets/Reads analysis/search range (X-axis) :CALC:FP[1-1]:TRAC[1-4]:FUN Analysis Range (X) C:DOM:X Sets/Reads analysis/search range (Y-axis) :CALC:FP[1-1]:TRAC[1-4]:FUN Analysis Range (Y) C:DOM:Y Analysis Type Sets/Reads analysis type :CALC:FP[1-1]:TRAC[1-4]:FUN C:TYPE Band Marker X Turns on/off bandmarker X...
Page 410
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Sets/Reads the stop value of bandmarker X :CALC:FP[1-1]:TRAC[1-4]:BDM Stop :X:STOP Band Marker Y Turns on/off bandmarker Y :CALC:FP[1-1]:TRAC[1-4]:BDM Band Marker Y :Y:STAT Sets/Reads the center value of bandmarker Y :CALC:FP[1-1]:TRAC[1-4]:BDM Center :Y:CENT...
Page 411
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Executes marker target search :CALC:FP[1-1]:TRAC[1-4]:MAR Search Target K[1-6]:SEAR:EXEC:TARG Sets/Reads the target transition definition :CALC:FP[1-1]:TRAC[1-4]:MAR Target Transition K[1-6]:SEAR:TARG:TRAN Sets/Reads the marker target value :CALC:FP[1-1]:TRAC[1-4]:MAR Target Value K[1-6]:SEAR:TARG:Y Sets/Reads the marker tracking type :CALC:FP[1-1]:TRAC[1-4]:MAR Tracking...
Page 412
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Save/Recall Opens windows explorer Explorer... Recalls state file by file name :MMEM:LOAD:STAT Recall by filename Recall State Recalls settings :MMEM:LOAD:STAT Autorec Opens file dialog File Dialog... State01 Recalls state file from register 1 :MMEM:LOAD:STAT...
Page 413
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Sets/Reads the reference level value :DISP:FP[1-1]:TRAC[1-4]:Y[:SC Reference Value AL]:RLEV Sets/Reads scale per division :DISP:FP[1-1]:TRAC[1-4]:Y[:SC Scale/Div AL]:PDIV Setup Carrier Search Searches carrier signal and reflects the result to the input frequency of the :SENS:FP[1-1]DCON:SSE:EXEC downconverter Sets/Reads frequency resolution...
Page 414
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Sets/Reads the stop value of DC Control voltage :SOUR:FP[1-1]:VOLT:CONT:ST DC Control Stop DC Power Center Sets/Reads the center value of DC Power voltage :SOUR:FP[1-1]:VOLT:POW:CEN Sets/Reads the span value of DC Power voltage :SOUR:FP[1-1]:VOLT:POW:SPA DC Power Span Sets/Reads the start value of DC Power voltage...
Page 415
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Frequency Offset (User Downconv.) Sets/Reads the conversion mode of the frequency offset :SENS:UDC:MODE Conversion Mode Sets/Reads the frequency offset :SENS:UDC:STAT Frequency Offset Sets/Reads the frequency offset factor :SENS:UDC:HARM Harmonic # LO Frequency...
Page 416
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Disables from mouse/touch screen operations :SYST:KLOC:MOUS Touch Screen & Mouse Lock Network Setup Sets MAC address MAC Address Enables/disables network connections Network Configuration Sets network ID of the instrument Network Identification Sets SICL-LAN address...
Page 417
SCPI Command Reference Softkey Functions Table 7-2 FP Menu Key Operation Function SCPI Command Copies trace data to the user trace :CALC:FP[1-1]:TRAC[1-4]:DAT Copy to User A:COPY Copies data to memory :CALC:FP[1-1]:TRAC[1-4]:MAT Data -> Mem H:MEM Data hold :CALC:FP[1-1]:TRAC[1-4]:HOL Data Hold Sets/Reads math operation type :CALC:FP[1-1]:TRAC[1-4]:MAT Data Math...
Page 418
SCPI Command Reference Softkey Functions PN Menu Table 7-3 PN Menu Key Operation Function SCPI Command Attenuator Sets/Reads Input Attenuator level on 5dB Step :SENS:ATT:LEV Input Attenuator Average Turns on/off averaging function :SENS:PN[1-1]:AVER:STAT Averaging Restarts averaging :SENS:PN[1-1]:AVER:CLE Averaging Restart Sets/Reads average count :SENS:PN[1-1]:AVER:COUN Avg Factor Sets/Reads the number of correlation...
Page 419
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command DC Power Voltage Sets/Reads DC Power delay (sec) :SOUR:VOLT:POW:DEL DC Power Delay Turns on/off DC Power voltage :SOUR:VOLT:POW:LEV:STAT DC Power Output DC Power Voltage Sets/Reads DC Power voltage :SOUR:VOLT:POW:LEV:AMPL Sets/Reads the maximum DC Power voltage limit :SOUR:VOLT:POW:LIM:HIGH...
Page 420
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Sets/Reads the signal supplied to the RF input port :SENS:DCON:INP RF Input Sets the use of the external mixer on or off and reads its settings :SENS:DCON:MEXT External Mixer Macro Setup...
Page 422
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Sets/Reads the span value of bandmarker X :CALC:PN[1-1]:TRAC[1-1]:BD Span M:X:SPAN Start Sets/Reads the start value of bandmarker X :CALC:PN[1-1]:TRAC[1-1]:BD M:X:STAR Sets/Reads the stop value of bandmarker X :CALC:PN[1-1]:TRAC[1-1]:BD Stop M:X:STOP...
Page 423
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Peak Sets/Reads the peak excursion value :CALC:PN[1-1]:TRAC[1-1]:MA Peak Excursion RK[1-6]:SEAR:PEAK:EXC Sets/Reads the marker peak-search polarity :CALC:PN[1-1]:TRAC[1-1]:MA Peak Polarity RK[1-6]:SEAR:PEAK:POL Search Left Executes marker peak search left :CALC:PN[1-1]:TRAC[1-1]:MA RK[1-6]:SEAR:EXEC:LPE Executes marker peak search...
Page 424
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Show Window Turns on/off frequency, power and DC current measurement mode :DISP:FP[1-1]:STAT Freq & Power Turns on/off phase noise measurement mode :DISP:PN[1-1]:STAT Phase Noise Turns on/off spectrum monitor mode :DISP:SP[1-1]:STAT Spectrum Monitor...
Page 425
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Saves state file to register 2 :MMEM:STOR:STAT State02 Saves state file to register 3 :MMEM:STOR:STAT State03 Saves state file to register 4 :MMEM:STOR:STAT State04 State05 Saves state file to register 5 :MMEM:STOR:STAT Saves state file to register 6 :MMEM:STOR:STAT...
Page 426
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Sets 40MHz to the stop frequency :SENS:PN[1-1]:FREQ:STOP 40MHz Sets 5MHz to the stop frequency :SENS:PN[1-1]:FREQ:STOP 5MHz System Aborts printing :HCOP:ABOR Abort Printing Turns on/off backlight :SYST:BACK:STAT Backlight Saves screen image :MMEM:STOR:IMAG...
Page 427
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Selects print mode :HCOP:IMAG Invert Image Misc Setup Beeper Beep Turns on/off the beep for operation completion :SYST:BEEP:COMP:STAT Complete Turns on/off the beep for warning :SYST:BEEP:WARN:STAT Beep Warning Makes beep sound for operation completion :SYST:BEEP:COMP:IMM...
Page 428
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Enables/disables Telnet server Telnet Server Outputs print :HCOP:IMM Print Executes printer setup Printer Setup ... Reads product information Product Information Service Menu Displays softkeys associated with Administrator Menu. This function is Administrator not available to general users.
Page 429
SCPI Command Reference Softkey Functions Table 7-3 PN Menu Key Operation Function SCPI Command Smoothing on/off :CALC:PN[1-1]:TRAC[1-1]:SM Smoothing O:STAT Spurious Clears the threshold data :CALC:PN[1-1]:TRAC[1-1]:SPU Clear Threshold R:THR:TABL:CLE Table Import Threshold Reads the threshold data :MMEM:PN[1-1]:TRAC[1-1]:LO AD:SPUR:THR Table ... Disables the spurious power value display :CALC:PN[1-1]:TRAC[1-1]:SPU Normalized R:POW...
Page 430
SCPI Command Reference Softkey Functions SP Menu Table 7-4 SP Menu Key Operation Function SCPI Command Attenuator Sets/Reads Input Attenuator level on 5dB Step :SENS:ATT:LEV Input Attenuator Average/BW Turns on/off averaging function :SENS:SP[1-1]:AVER:STAT Averaging Restarts averaging :SENS:SP[1-1]:AVER:CLE Averaging Restart Sets/Reads averaging type :SENS:SP[1-1]:AVER:TYPE Averaging Type Sets/Reads the averaging count...
Page 431
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Sets/Reads the minimum DC Control voltage limit :SOUR:VOLT:CONT:LIM:LOW Min Ctrl Voltage Limit DC Power Voltage Sets/Reads DC Power delay (sec) :SOUR:VOLT:POW:DEL DC Power Delay DC Power Output Turns on/off DC Power voltage :SOUR:VOLT:POW:LEV:STAT Sets/Reads DC Power voltage...
Page 432
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Format Sets/Reads the detector mode :SENS:SP[1-1]:DET:FUNC Detector Mode SP format :CALC:SP[1-1]:TRAC[1-1]:FOR Format Input Port Downconverter Sets the use of the downconverter on or off, or reads its setting :SENS:DCON[:STAT] Downconverter Sets/Reads the signal supplied to the RF input port...
Page 433
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Marker Clear Marker Menu Clears all the markers All OFF Marker 1 Turns on/off marker 1 :CALC:SP[1-1]:TRAC[1-1]:MAR K[1-6]:STAT Turns on/off marker 2 :CALC:SP[1-1]:TRAC[1-1]:MAR Marker 2 K[1-6]:STAT Turns on/off marker 3 :CALC:SP[1-1]:TRAC[1-1]:MAR Marker 3...
Page 434
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Sets/Reads analysis type :CALC:SP[1-1]:TRAC[1-1]:FUN Analysis Type C:TYPE Band Marker X Turns on/off bandmarker X :CALC:SP[1-1]:TRAC[1-1]:BDM Band Marker X :X:STAT Sets/Reads the center value of bandmarker X :CALC:SP[1-1]:TRAC[1-1]:BDM Center :X:CENT...
Page 435
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Sets/Reads the span value of bandmarker Y :CALC:SP[1-1]:TRAC[1-1]:BDM Span :Y:SPAN Sets/Reads the start value of bandmarker Y :CALC:SP[1-1]:TRAC[1-1]:BDM Start :Y:STAR Sets/Reads the stop value of bandmarker Y :CALC:SP[1-1]:TRAC[1-1]:BDM Stop :Y:STOP...
Page 436
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Marker To Sets/Reads the center value of frequency span :SENS:SP[1-1]:FREQ:CENT Marker -> Center Sets/Reads the start value of frequency span :SENS:SP[1-1]:FREQ:STAR Marker -> Start Sets/Reads the stop value of frequency span :SENS:SP[1-1]:FREQ:STOP Marker ->...
Page 437
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Saves memory trace data :MMEM:SP[1-1]:TRAC[1-1]:ST Save Memory Trace OR:MEM Save State Saves settings :MMEM:STOR:STAT Autorec File Dialog... Opens file dialog Selects instrument state type (Entire or instrument state only) :MMEM:STOR:STYP Save Type Saves state file to register 1...
Page 438
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Sets the magnification of carrier frequency when center frequency is set. Harmonic # Sets/Reads the center value of frequency span :SENS:SP[1-1]:FREQ:CENT Center Sets/Reads the span value of frequency span :SENS:SP[1-1]:FREQ:SPAN Span Sets/Reads the start value of frequency span...
Page 439
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Downconverter Manual Setup Sets/Reads the bias current to be supplied to the external mixer :SENS:DCON:MAN:MEXT[1-2]: Current BIAS:CURR Sets/Reads the IF gain of the external mixer :SENS:DCON:MAN:IFG[1-2] IF Gain 1 IF Gain 2 Sets/Reads the LO frequency of the external mixer...
Page 440
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Opens control panel Control Panel ... GPIB Setup Turns on/off system controller mode System Controller Configuration Sets the address for controlling the analyzer from a controller via GPIB. Talker/Listene r Address Key Lock...
Page 441
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Test Menu Performs internal test Power On Test Performs display test Display Test Front Panel Performs front panel key (hard key) test Performs touch screen calibration Adjust Touch Screen Displays the connection status of...
Page 442
SCPI Command Reference Softkey Functions Table 7-4 SP Menu Key Operation Function SCPI Command Triggers source :TRIG:SP[1-1]:SOUR Source Selects measurement mode :TRIG:MODE Trigger to Spectrum Monitor...
Page 443
SCPI Command Reference Softkey Functions TR Menu Table 7-5 TR Menu Key Operation Function SCPI Command Attenuator Sets/Reads Input Attenuator level on 5dB Step :SENS:ATT:LEV Input Attenuator Average Turns on/off averaging function :SENS:TR[1-1]:AVER:STAT Averaging Averaging Restart Restarts averaging :SENS:TR[1-1]:AVER:CLE Sets/Reads averaging count :SENS:TR[1-1]:AVER:COUN Avg Factor DC Control Voltage...
Page 444
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command DC Power Voltage Sets/Reads DC Power delay (sec) :SOUR:VOLT:POW:DEL DC Power Delay Turns on/off DC Power voltage :SOUR:VOLT:POW:LEV:STAT DC Power Output Sets/Reads DC Power voltage :SOUR:VOLT:POW:LEV:AMPL DC Power Voltage Sets/Reads the maximum DC Power voltage limit :SOUR:VOLT:POW:LIM:HIGH...
Page 445
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Sets/Reads the reference frequency :CALC:TR[1-1]:TRAC[1-4]:REF: Frequency FREQ Reference Selects phase format on transient measurement :CALC:TR[1-1]:TRAC[1-4]:FOR Phase Unit M:PHAS:UNIT Turns on/off wrap-phase :CALC:TR[1-1]:TRAC[1-4]:FOR Wrap Phase M:PHAS:WRAP Input Port Downconverter Sets the use of the downconverter on or off, or reads its settingǵ...
Page 446
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Saves VBA project :MMEM:STOR:PROG Save Project Marker Clear Marker Menu Clears all the markers All OFF Turns on/off marker 1 :CALC:TR[1-1]:TRAC[1-4]:MA Marker 1 RK[1-6]:STAT Turns on/off marker 2 :CALC:TR[1-1]:TRAC[1-4]:MA Marker 2 RK[1-6]:STAT...
Page 447
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Marker Function Sets/Reads analysis/search range (X-axis) :CALC:TR[1-1]:TRAC[1-4]:FUN Analysis Range (X) C:DOM:X Sets/Reads analysis/search range (Y-axis) :CALC:TR[1-1]:TRAC[1-4]:FUN Analysis Range (Y) C:DOM:Y Analysis Type Sets/Reads analysis type :CALC:TR[1-1]:TRAC[1-4]:FUN C:TYPE Band Marker X Turns on/off bandmarker X...
Page 448
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Sets/Reads the stop value of bandmarker X :CALC:TR[1-1]:TRAC[1-4]:BD Stop M:X:STOP Band Marker Y Turns on/off bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BD Band Marker Y M:Y:STAT Sets/Reads the center value of bandmarker Y :CALC:TR[1-1]:TRAC[1-4]:BD Center M:Y:CENT...
Page 449
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Executes marker target search :CALC:TR[1-1]:TRAC[1-4]:MA Search Target RK[1-6]:SEAR:EXEC:TARG Sets/Reads the target transition definition :CALC:TR[1-1]:TRAC[1-4]:MA Target Transition RK[1-6]:SEAR:TARG:TRAN Sets/Reads the marker target value :CALC:TR[1-1]:TRAC[1-4]:MA Target Value RK[1-6]:SEAR:TARG:Y Sets/Reads the marker tracking type :CALC:TR[1-1]:TRAC[1-4]:MA Tracking...
Page 450
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Recalls state file from register 2 :MMEM:LOAD:STAT State02 Recalls state file from register 3 :MMEM:LOAD:STAT State03 Recalls state file from register 4 :MMEM:LOAD:STAT State04 Recalls state file from register 5 :MMEM:LOAD:STAT State05 Recalls state file from register 6...
Page 451
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Setup Sets/Reads frequency transient range (Narrowband) :SENS:TR[1-1]:NARR:FREQ:RA Freq Range Sets/Reads maximum input level :SENS:TR[1-1]:POW:INP:LEV: Max Input Level Phase Reference Sets/Reads phase reference freuency :SENS:TR[1-1]:NARR:FREQ:PR Recalc Phase Reference Sets/Reads the offset value of the phase reference frequency :CALC:TR[1-1]:TRAC[1-4]:FOR...
Page 452
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command System Aborts printing :HCOP:ABOR Abort Printing Turns on/off backlight :SYST:BACK:STAT Backlight Saves screen image :MMEM:STOR:IMAG Dump Screen Image Instrument Setup Correction File Dialog ... Loads correction data for a specified power :MMEM:LOAD:CORR:POW Loads correction data for a specified power :MMEM:LOAD:CORR:POW...
Page 453
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Misc Setup Beeper Turns on/off the beep for operation completion :SYST:BEEP:COMP:STAT Beep Complete Beep Warning Turns on/off the beep for warning :SYST:BEEP:WARN:STAT Makes beep sound for operation completion :SYST:BEEP:COMP:IMM Test Beep Complete...
Page 454
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Outputs print :HCOP:IMM Print Executes printer setup Printer Setup ... Reads product information Product Information Service Menu Displays softkeys associated with Administrator Menu. This function is Administrator Menu not available to general users.
Page 455
SCPI Command Reference Softkey Functions Table 7-5 TR Menu Key Operation Function SCPI Command Trace View Sets/Reads smoothing aperture value :CALC:TR[1-1]:TRAC[1-4]:SM Aperture O:APER Copies trace data to the user trace :CALC:TR[1-1]:TRAC[1-4]:DAT Copy to User A:COPY Data -> Mem Copies data to memory :CALC:TR[1-1]:TRAC[1-4]:MAT H:MEM Data hold...
Page 456
SCPI Command Reference Softkey Functions USER Menu Table 7-6 User Menu Key Operation Function SCPI Command Attenuator Sets/Reads Input Attenuator level on 5dB Step :SENS:ATT:LEV Input Attenuator DC Control Voltage Auto Freq Control Turns on/off the auto frequency control function. :SOUR:VOLT:CONT:AFC[:STAT AFC Status Executes the auto frequency control once.
Page 457
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Sets/Reads the minimum DC Power voltage limit :SOUR:VOLT:POW:LIM:LOW Min Pwr Voltage Limit Display Edits the measurement window title label :DISP:USER[1-1]:LAB:DATA Edit Title Label Limit Test Clears the lower limit line :CALC:USER[1-1]:TRAC[1-8]:L Delete Lower IM:LOW:SEGM:CLE...
Page 458
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Echo Window Menu Clears Echo window :DISP:ECHO:CLE Clear Echo Sets/Reads the font size on Echo window :DISP:ECHO:FSIZ Echo Font Size Turns on/off the Echo window :DISP:ECHO:STAT Echo Window Loads and executes the macro selected on file names.
Page 460
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Sets/Reads the stop value of bandmarker X :CALC:USER[1-1]:TRAC[1-8]:B Stop DM:X:STOP Band Marker Y Turns on/off bandmarker Y :CALC:USER[1-1]:TRAC[1-8]:B Band Marker Y DM:Y:STAT Sets/Reads the center value of bandmarker Y :CALC:USER[1-1]:TRAC[1-8]:B Center DM:Y:CENT...
Page 461
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Peak Sets/Reads the peak excursion value :CALC:USER[1-1]:TRAC[1-8]: Peak Excursion MARK[1-6]:SEAR:PEAK:EXC Sets/Reads the marker peak-search polality :CALC:USER[1-1]:TRAC[1-8]: Peak Polarity MARK[1-6]:SEAR:PEAK:POL Search Left Executes marker peak search left :CALC:USER[1-1]:TRAC[1-8]: MARK[1-6]:SEAR:EXEC:LPE Executes marker peak search...
Page 462
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Turns on/off spectrum monitor mode :DISP:SP[1-1]:STAT Spectrum Monitor Transient Turns on/off transient measurement mode :DISP:TR[1-1]:STAT Turns on/off user defined window :DISP:USER[1-1]:STAT User Selects spectrum monitor mode :DISP:WIND:ACT Spectrum Monitor Transient...
Page 463
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Saves state file to register 5 :MMEM:STOR:STAT State05 Saves state file to register 6 :MMEM:STOR:STAT State06 Scale Auto Scale Executes autoscale :DISP:USER[1-1]:TRAC[1-8]:Y[: SCAL]:AUTO Executes autoscale for all traces on user defined window :DISP:USER[1-1]:ALLT:Y:SCAL Auto Scale All :AUTO...
Page 464
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Sets/Reads the IF gain of the external mixer :SENS:DCON:MAN:IFG[1-2] IF Gain 1 IF Gain 2 Sets/Reads the LO frequency of the external mixer :SENS:DCON:MAN:LO[1-2]:FR Frequency Frequency Sets/Reads the LO level of the external mixer :SENS:DCON:MAN:LO[1-2]:LE LO1 Level...
Page 465
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command GPIB Setup Turns on/off system controller mode System Controller Configuration Sets the address for controlling the analyzer from a controller via GPIB Talker/Listene r Address Key Lock Front Panel &...
Page 466
SCPI Command Reference Softkey Functions Table 7-6 User Menu Key Operation Function SCPI Command Performs display test Display Test Performs front panel key (hard key) test Front Panel Performs touch screen calibration Adjust Touch Screen Displays the connection status of E5053A Test E5053A Trace View...
Page 467
Agilent E5052A than that indicated by the current printing date of this manual. The information in this manual applies directly to the E5052A model that has the serial number prefix listed on the title page of this manual.
Page 468
Manual Changes Manual Changes Manual Changes To adapt this manual to your Agilent E5052A, refer to Table A-1 and Table A-2. Table A-1 Manual Changes by Serial Number Serial Prefix or Number Make Manual Changes MY442 or later Change 4...
Page 469
Manual Changes Manual Changes Change 4 The equipment with prefix MY441 or earlier does not support the USB (USBTMC) interface port nor the removable hard disk function. Change 3 The following functions are integrated newly into the firmware version A.01.50 onward. They are not supported by the firmware version A.02.00 or earlier.
Page 470
Manual Changes Manual Changes Change 1 The functions listed below are limited when option 011 is installed. Table A-3 Limited functions when option 011 is installed Functions Limitations [PN] Start frequency Minimum value is 10 Hz [PN] Correlation Not Available (The value is fixed as 1) [PN] IF Gain Not Available (The value is fixed as 10 dB) [FP] Trigger Mode...
Page 471
Status Reporting System This appendix describes the status reporting system of the Agilent E5052A.
Page 472
Status Reporting System General Status Register Model General Status Register Model The Agilent E5052A has a status reporting system to report the condition of the instrument. Figure B-1 General status register model The status reporting system has a hierarchical structure as shown in Figure B-1. When the instrument satisfies a particular condition, the corresponding bit of the event register is set to 1.
Page 473
General Status Register Model Event Register This reflects the corresponding condition of the E5052A (e.g., occurrence of an event) as a bit status. These bits continuously monitor changes in the E5052A’s state and change the bit status when the condition for each bit is met (e.g., changing bit status to “1” if a specific event occurs).
Page 474
“1” when bit 3 of the condition register makes a negative transition, that is, when it changes from 1 to 0. Figure B-2 Transition filter and condition register In the E5052A, the following registers provide a condition register and transition filter: • Operation status register •...
Page 475
Figure B-5. The status byte register is a summary of registers in the lower level. This section describes the E5052A’s status registers in each hierarchy. Each bit of the status register is described in Table B-1 through Table B-10.
Page 476
Status Reporting System Status Register Structure Figure B-3 Status register structure (1 of 6) Appendix B...
Page 477
Status Reporting System Status Register Structure Figure B-4 Status register structure (2 of 6) Appendix B...
Page 478
Status Reporting System Status Register Structure Figure B-5 Status register structure (3 of 6) Appendix B...
Page 479
Status Reporting System Status Register Structure Figure B-6 Status register structure (4 of 6) Appendix B...
Page 480
Status Reporting System Status Register Structure Figure B-7 Status register structure (5 of 6) Appendix B...
Page 481
Status Reporting System Status Register Structure Figure B-8 Status register structure (6 of 6) Appendix B...
Page 482
Status Reporting System Status Register Structure Table B-1 Status Bit Definitions of Status Byte Register Name Description Position 0, 1 Not used Always 0. Set to “1” if error/event queue contains data; reset to “0” Error/Event Queue when all data has been retrieved. Questionable Status Register Set to “1”...
Page 483
Not used Always 0 Power ON Set to “1” when the E5052A is powered ON, or when the firmware is restarted. Issuing the *CLS command will clear all of the bits from the standard event status register. Appendix B...
Page 484
0 - 14 Defined by the user Set to “1” as defined by the user. Not used Always 0. The user-defined bits (0 to 14) can be specified by the user as required and are not used by the E5052A. Appendix B...
Page 485
Always 0. RF frequency out of range Set to “1” when DUT’s oscillation frequency is out of E5052A’s measurement range. In this case, DUT’s oscillation frequency should be verified. Questionable Phase Status Register Set to “1” when one of the enable bits in the Summary questionable phase status register is set to “1.”...
Page 486
Phase Lock Loop Unlocked Set to “1” when Phase Lock Loop is not locked. PLL Frequency Out of Range Set to “1” when PLL frequency is out of E5052A’s measurement range. PLL Input Overflow Set to “1” when PLL input overflows.
Page 487
Status Reporting System Status Register Structure Table B-9 Status Bit Definitions of Questionable Reference Signal Status Event Register Name Description Position Not used Always 0. External Reference Unlock Set to “1” when external reference is not locked. 2-15 Not used Always 0.
Page 488
Status Reporting System Status Register Structure register. Table B-12 Status Bit Definitions of Questionable Limit PN1 Status Event Register Name Description Position Not used Always 0. Trace 1 Limit Test Fail Set to “0” when a measurement cycle begins; set to “1” when the measurement cycle finishes and returns “fail”...
Page 489
Status Reporting System Status Register Structure event register. Table B-15 Status Bit Definitions of Questionable Limit TR1 Status Event Register Name Description Position Not used Always 0. Trace 1 Limit Test Fail Set to “0” when a measurement cycle begins; set to “1”...
Page 490
Status Reporting System Status Register Structure Table B-16 Status Bit Definitions of Questionable Limit USER1 Status Event Register Name Description Position 9-15 Not used Always 0. Issuing the *CLS command will clear all of the bits in the questionable limit USER1 status event register.
Page 491
Status Reporting System Using the Status Reporting System Using the Status Reporting System You can manage the status report system by using the following commands in any combination: • *CLS on page 273 • *SRE on page 275 • *STB on page 275 *ESE on page 273 •...
Page 492
Status Reporting System Using the Status Reporting System :STATus:QUEStionable:LIMit:TR[1-1]:ENABle on page 359 • :STATus:QUEStionable:LIMit:TR[1-1][:EVENt] on page 359 • :STATus:QUEStionable:LIMit:TR[1-1]:NTRansition on page 360 • • :STATus:QUEStionable:LIMit:TR[1-1]:PTRansition on page 360 • :STATus:QUEStionable:LIMit:USER[1-1]:CONDition on page 360 :STATus:QUEStionable:LIMit:USER[1-1]:ENABle on page 361 • :STATus:QUEStionable:LIMit:USER[1-1][:EVENt] on page 361 •...
Page 493
Communication with External Instruments Using 24-bit I/O Port This chapter provides necessary information for communicating with external instruments (for example, a handler in a production line) by using the 24-bit I/O port equipped with the Agilent E5052A.
Page 494
Communication with External Instruments Using 24-bit I/O Port 24-bit I/O Port Overview 24-bit I/O Port Overview The E5052A 24-bit I/O port provides four independent parallel ports for data I/O associated with several control signal lines and the power line. All signals operate in TTL logic.
Page 495
Communication with External Instruments Using 24-bit I/O Port I/O Signal Pin Layout and Description I/O Signal Pin Layout and Description Figure A-2 illustrates the layout of the I/O signal pins on the 24-bit interface connector, and Table C-1 on page 496 briefly describes these signals. Figure C-2 24-bit interface connector pin layout NOTE...
Page 496
Communication with External Instruments Using 24-bit I/O Port I/O Signal Pin Layout and Description Table C-1 Description of the 24-bit Interface I/O Signals Signal name Signal Description number direction —— Ground /INPUT1 Input When this port receives a negative pulse, /OUTPUT1 and /OUTPUT2 are changed to the Low level.
Page 497
Communication with External Instruments Using 24-bit I/O Port I/O Signal Pin Layout and Description Table C-1 Description of the 24-bit Interface I/O Signals Signal name Signal Description number direction PORT C STATUS Output Port C status signal. This signal is changed to the High level when port C is configured to output port.
Page 498
Communication with External Instruments Using 24-bit I/O Port Inputting/Outputting Data Inputting/Outputting Data The E5052A 24-bit I/O port provides the ports for data I/O shown in Table C-2. Table C-2 I/O Ports Port Name Usage Data Structure Port A Output Port B...
Page 499
Communication with External Instruments Using 24-bit I/O Port Inputting/Outputting Data Reading data input to port When port C, D, or E is configured to input port, the binary data represented with High(0)/Low(1) of each bit of the port is read as decimal data. To retrieve the data, use the following commands as query: Port Name Command...
Page 500
Communication with External Instruments Using 24-bit I/O Port Preset States at Power-on Preset States at Power-on The 24-bit I/O port is set at power-on as follows (not affected at reset). Port A High (All Bits) Port B High (All Bits) Port C Input Port D...
Page 501
Communication with External Instruments Using 24-bit I/O Port Timing Chart and Pulse Width Timing Chart and Pulse Width When the formatted data calculation is completed during the sweep, a negative pulse is provided. The pulse width of the sweep completion signal is shown in Figure C-3. Table C-3 Value of T1 in Figure C-3 (typical) Typical Value...
Page 502
Communication with External Instruments Using 24-bit I/O Port Timing Chart and Pulse Width Figure C-5 shows a timing chart of the pulse input to INPUT1, /OUTPUT1 signal output and /OUTPUT2 signal output. Table C-5 Values of T1 through T2 in Figure C-5 (typical) Minimum value 1 μs Pulse width of /INPUT1...
Page 503
Communication with External Instruments Using 24-bit I/O Port Electrical Characteristics Electrical Characteristics Input signal All input signals are TTL compatible. Table C-7 shows the electrical characteristics of the input signals, and Figure C-7 shows the circuit diagram of the input signals. Table C-7 Electrical Characteristics of Input Signals (typical) Maximum rate input voltage...
Page 504
Communication with External Instruments Using 24-bit I/O Port Electrical Characteristics Output signal All output signals are TTL compatible. Table C-8 shows the electrical characteristics of output signals, and Figure C-8 shows the circuit diagram of the output signals. Table C-8 Electrical Characteristics Of Output Signals (typical) Maximum rate output current -10 mA to 10 mA...
Page 505
Error Messages The Agilent E5052A provides error messages to indicate its operating status. This appendix describes the error messages of the E5052A, listed in alphabetical order.
Page 506
A21 flash ROM write error This error occurs when writing the system calibration data to ROM on the A21 board fails. Contact Agilent Technologies’ Customer Contact given at the end of this guide or the company from which you bought the device.
Page 507
The current through the DC POWER connector is too large. Downconverter Fan Stop This error occurs when the fan of the downconverter stops. Contact Agilent Technologies’ Customer Contact given at the end of this guide or the company from which you bought the device.
Page 508
Error Messages Error number: -200 suspected. Contact Agilent Technologies’ Customer Contact given at the end of this guide or the company from which you bought the device. -200 Execution error An error associated with execution has been generated for which this instrument cannot specify the error message.
Page 509
DUT that has an unstable output level is measured. Aside from the above reasons, there is the possibility of a device failure. Contact an Agilent Technologies sales office or the company from which you bought the device.
Page 510
Error Messages Error number: 20 Marker search failed This error occurs when marker search fails. -109 Missing parameter The number of parameters is less than that required for the command, or the parameter has not been entered. No downconverter unit connected This error occurs when the downconverter is not turned on or when the USB (USBTMC) cable is not connected even though the downconverter is enabled.
Page 511
Stop the frequency modulation. • A large harmonics component is included in the DUT’s output signal. Insert a low-pass filter between the DUT’s output terminal and the E5052A RF IN connector to eliminate the harmonics component. PLL frequency range over This error occurs when the internal PLL circuit becomes unlocked while the measurement is in progress.
Page 512
One of the errors designated as “system errors” in this instrument has occurred. Target freq out of range This error indicates the 2nd, 3rd, or nth order harmonics is not within the E5052A measurement range. At this time, the previous measurement conditions still remain (any of sweep parameters are not changed).
Page 513
Error Messages Error number: 571 Warning Message A warning message is displayed in the instrument message/Warning area in the lower-left part of the display against a gray background. Touching Entry Off key at the front panel executing the :DISPlay:MESSage:CLEar command on page 237 clears the message. Moreover, about a specific warning message, when a sweep is started again, a warning message may disappear.
Page 514
Error Messages Error number: 508 20 dB. Set RF ATT 25dB This message occurs when the input attenuator is set improperly. Set the input attenuator to 25 dB. Set RF ATT 30dB This message occurs when the input attenuator is set improperly. Set the input attenuator to 30 dB.
Page 515
Looking up SCPI commands, 26 Deleting a file (directory), 80 Detecting Occurrence of an Error, 86 Device, 94 Manual Changes, 468 Device Configuration Using E5052A and E5053A Manual trigger, 52 Microwave Downconverter, 94 Measurement State, 53 Device selector, 32 measuring mode, 26...