Agilent Technologies E5071C Manual page 892

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

2
Query Error
3
Instrument
Dependent
Error
4
Execution
Error
5
Command
Error
6
Not used
1. Set to "1" when the E5071C receives a
data output request but there is no data to
output.
2. Set to "1" when the data of the
E5071C's output queue has been cleared
because of a new message received before
the completion of data output.
Set to "1" when an error has occurred and
the error is not a command, query, or
execution error.
1. Set to "1" when any parameter in an
SCPI command exceeds its input range or
is inconsistent with the E5071C's
capabilities.
2. Set to "1" when an SCPI command
cannot be properly executed due to some
condition of the E5071C.
1. Set to "1" when an IEEE 488.2 syntax
error occurs (a command sent to the
E5071C does not follow the IEEE 488.2
syntax). Possible violations include the
command parameter violating the E5071C
listening formats or being unacceptable.
2. Set to "1" when a semantic error
occurs. Possible causes include a
command containing misspellings being
sent to the E5071C or an IEEE 488.2
command not supported by the E5071C
being sent.
3. Set to "1" when GET (Group Execution
Trigger) is input while a program message
is being received.
Always 0