Agilent Technologies E5071C Manual page 443

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
Number of points: 201
2. Set sweep conditions for channels 2, 3, and 4 according to the same
setting of channel 1.
6. Setting Balance Conversion Topology
For channel 2, set the balance conversion topology in order to perform
measurement that includes the balanced port.
Setting Description
Set channel 2 to the active
channel.
Set DUT port 1 to unbalance and
DUT port 2 to balance.
Set the destination to which DUT
port 1 (unbalance) is connected to
test port 1 of the analyzer.
Set the destination to which DUT
port 2 (balance) is connected to
test ports 3 and 4 of the analyzer.
Unbalance-balance conversion for
trace 1: ON
Fixture simulator: ON
7. Selecting Measurement Parameter
1. Select the measurement parameter for trace 1 of channel 1.
Setting Description
Set channel 1 to the active
channel.
Measurement parameter for
trace 1: S12
2. The subscript of the measurement parameter means the test
port of the E5071C. Check the test port assignment and select
the measurement parameter.
442
> Point >
Sweep Setup
>
1
x1
Key Operation
(press it until
Channel Next
channel 2 is activated)
> Fixture Simulator >
Analysis
Topology > Device > SE-Bal
Port 1 (se) > 1
Port 2 (bal) > 3-4
BalUn (set it to ON)
Fixture Simulator (set it to ON)
Key Operation
(press it until
Channel Next
channel 1 is activated)
> S12
Meas
>
>
2
0