Full 3-Port and Full 4-Port Calibration using 2-Port ECal

Overview
Procedure
Other topics about Calibration with ECal
Overview
A VBA macro (ECal Assistant) is pre-installed in the E5071C to carry out a
full 3-port or a full 4-port calibration using the 2-port ECal.
ECal Assistant does not perform isolation calibration.
Procedure
1. Connect the USB port of the ECal module to the USB port on the
E5071C with a USB cable. The connection may be made while the
E5071C's power is on.
2. Allows the ECal module to warm up for 15 minutes until the module
indicates from WAIT to READY.
3. Press
Macro Setup
4. Click Load Project.
5. From the Open dialog box, select the VBA project file
D:\Agilent\ECalAssistant.vba and click the Open button.
6. Press
Macro Run
7. The following dialog box appears.
8. Click Next.
key.
key.
Measurement
159