Agilent Technologies E5071C Manual page 61

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
7. Specify the IF bandwidth of the receiver as necessary. In this
measurement example, the IF bandwidth is set to 10 kHz because of
the need to lower the noise floor.
> IF Bandwidth >
Avg
STEP 2. Calibration
To turn the error correction ON, set the calibration type to the full 2-port
calibration and measure the calibration data.
For details about calibration, see Calibration.
1. Select the calibration kit suitable for the measurement cable. In this
measurement example, Calibration Kit 85032F is selected.
> Cal Kit > 85032F
Cal
2. Set the calibration type to the full 2-port calibration using the test
port 1 and 2.
> Calibrate > 2-Port Cal > Select Ports - 1-2
Cal
3. Connect the OPEN standard (included in the calibration kit) to the
other end of the measurement cable that is connected to the test
port 1 as shown in the following figure, and measure the open
calibration data at the test port 1. After measuring the open
calibration data, a checkmark
Open menu.
> Calibrate > 2-Port Cal > Reflection > Port1 Open
Cal
In the same way, measure the calibration data for the SHORT/LOAD
standards at the test port 1.
Connecting the OPEN/SHORT/LOAD standards
60
>
>
1
0
k/m
is displayed to the left of the Port 1