Measuring Saw Bandpass Filter Using Segment Sweep; Determine The Segment Sweep Conditions - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C

Measuring SAW Bandpass Filter using Segment Sweep

Overview
Procedure
Other Measurement Examples
Overview
This section illustrates how to use the segment sweep function to evaluate
a SAW bandpass filter with a center frequency of 947.5 MHz.
Procedure
Here, the DUT is evaluated by following the steps.
Step
1.
Determine the Segment Sweep
Conditions
2.
Create a Segment Sweep Table
3.
Select the Segment Sweep as the
Sweep Type
4.
Execute the Calibration
5.
Connect the DUT
6.
Execute the Measurement
7.
Define the Setup for Display

1. Determine the Segment Sweep Conditions

The following figure shows the results of evaluating the transmission
characteristics of the SAW bandpass filter in the range of 440 MHz to 3
GHz by using the linear sweep.
Transmission characteristics of SAW bandpass filter (440 MHz to 3
GHz, linear sweep)
394
Description
The segment sweep conditions are
determined according to the
characteristics of the DUT.
The segment sweep conditions are
entered in the E5071C.
The segment sweep is selected as
the sweep type.
A 2-port ECal is performed between
the test ports connecting the DUT.
The DUT is connected.
A trigger is applied to perform the
measurement.
The choice is made between
frequency base and order base as the
method of displaying segments.