The table below describes signals and time as shown in the above figure.
Signal,
Description
time
External
External trigger signal to be supplied.
Trig
Sampling
Time while the E5071C is actually performing
measurement.
Index
/Index signal of the handler I/O port. When the point
trigger function is ON and the low-latency external trigger
mode is ON, it goes to the High level only before staring
measurement of the first sweep point and returns to the
Low level after completing measurement of all
measurement points.
Point
Time until the E5071C is ready to accept a trigger for the
Trigger
next measurement point. The value depends on the
Period
measurement conditions and the settings of the E5071C.
td1
Time set as the external trigger delay time.
td2
Time that the E5071C requires internally for setting. This
depends on the settling time for setting the frequency, IF