Agilent E5061B Network Analyzers Service Guide Second Edition Manufacturing No. E5061-90105 December 2010...
Page 3
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Agilent Technologies. Microsoft®,MS-DOS®,Windows®,Visual C++®,Visual Basic®,VBA® and Excel® are registered UNIX is a registered trademark in U.S.
Page 4
The E5061B complies with INSTALLATION CATEGORY II as well as POLLUTION DEGREE 2 in IEC61010-1. The E5061B is an INDOOR USE product. NOTE The LEDs in the E5061B are Class 1 in accordance with IEC60825-1, CLASS 1 LED PRODUCT NOTE This equipment is MEASUREMENT CATEGORY I (CAT I).
Page 5
safety features are maintained in operational condition. • Dangerous Procedure Warnings Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. WARNING Dangerous voltage levels, capable of causing death, are present in this instrument. Use extreme caution when handling, testing, and adjusting this instrument.
Page 6
Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility, or to the calibration facilities of other International Standards Organization members.
Page 7
Typeface Conventions Sample (bold) Boldface type is used when a term is defined or emphasised. Sample (Italic) Italic type is used for emphasis. Indicates a hardkey (key on the front panel or external keyboard) labeled “Sample.” “key” may be omitted. menu/button/box Indicates a menu/button/box on the screen labeled Sample...
General Information The Service Guide is a guide to servicing the E5061B RF Network Analyzer. The guide contains information requisite to troubleshooting, and repairs.
Do not allow any computer virus to infect the system. This machine has no virus check function nor anti-virus software installed. Agilent Technologies will not be held liable for any failure or damage arising from negligence regarding these prohibitions and warnings.
Chapter 1 , “General Information,” The Service Guide is a guide to servicing the E5061B RF Network Analyzer. The guide contains information requisite to troubleshooting, and repairs. Chapter 2 , “Troubleshooting,”...
The supplement for this manual is identified by this manual's printing data and is available from Agilent Technologies. If the serial prefix or number of an instrument is lower than that on the title page of this manual, see Appendix A, Manual Changes. For information concerning, a serial number prefix that is not listed on the title page or in the Manual change supplement, contact the nearest Agilent Technologies office.
General Information Required Equipment Required Equipment Table 1-1 lists the recommended equipment for performing maintenance on the E5061B. Table 1-1 Recommended Test Equipment Equipment Recommended Model Qty. Frequency Counter Agilent 53181A woth option 010 & Opt.030 Power Meter Agilent E4419B...
Page 17
General Information Required Equipment Table 1-1 Recommended Test Equipment Equipment Recommended Model Qty. Adapter Agilent p/n 1250-0780 Agilent p/n 1250-0077 Agilent p/n 1250-1473 Agilent p/n 1250-1474 Agilent p/n 1250-2879 Torque Wrench Agilent p/n 8710-1766 USB/GPIB Interface Agilent 82357A Personal Computer with GPIB board *1.P: Performance Tests, A: Adjustment.
Troubleshooting Introduction Introduction WARNING These servicing instructions are for use by qualified personnel only. To avoid possible electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from its power supply beforehand.
If you wish to return to the Measurement View, double-click "Network Analyzer" icon. NOTE If you need to shut down the E5061B and again turn on, perform in accordance with the following procedure. a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with mouse.
29. Step 2. Check the display • If no display appears on the LCD after the E5061B is turned on, go to “No Display Troubleshooting” on page 22. • If the E5061B stops in booting process despite something being displayed on the LCD, go to “Booting Process Troubleshooting”...
Page 22
Troubleshooting To Troubleshoot the Instrument Figure 2-2 Primary trouble isolation flowchart Chapter 2...
No Display Troubleshooting No Display Troubleshooting If the E5061B displays nothing despite it is powered from proper ac power line, isolate the failure in accordance with the procedure shown in Figure 2-3. Connect the keyboard to the E5061B rear panel connector, turn the power on and start trouble isolation.
Page 24
Step 1. Check fan operation and DC monitor LED If the rear panel fan (blower) doesn't run, a failure in power supply is assumed. Remove the E5061B outer cover and check if the following LEDs light: • +3.3 V and +5 V dc monitor LEDs on A50/A60 CPU Module as shown in Figure 2-4 •...
Page 25
No Display Troubleshooting Step 2. Check system fan inside. If the system fan on the chassis inside the E5061B don't run, problem seems in the A51 Measurement I/F board or the flat cable between the A51 Measurement I/F board and CPU module.
Troubleshooting Booting Process Troubleshooting Booting Process Troubleshooting Figure 2-7 represents the booting process flow in the E5061B. If the E5061B stops in the booting process, troubleshoot using the following step-by-step procedure. Figure 2-7 Booting process flowchart Chapter 2...
Page 27
Troubleshooting Booting Process Troubleshooting Step 1. Splash Screen The splash screen is displayed with Agilent logo as shown in Figure 2-8. If the splash screen is displayed, you can assume that the A50/A60 CPU Module is functioning correctly. NOTE While the splash screen is displayed, if you want to run the BIOS setup utility, push F2 key as soon as in the screen.
Page 28
The operating system automatically checks the device drivers, which are necessary to use the E5061B functions and are installed in the system before the E5061B is shipped from Agilent factory. If the operating system doesn't detect them, a message box is displayed. In this case, install the device driver.
Page 29
Troubleshooting Booting Process Troubleshooting Figure 2-11 Firmware revision and option information If the display whited out, entirely blued or appeared with a dialog box, a mass storage problem is suspected. Try to perform the mass storage recovery procedure. NOTE If a message of "Will Shut Down in Five Seconds" is displayed in place of "Initializing.." and the shutdown occurs, the DSP board fails in starting up.
Page 30
Troubleshooting Booting Process Troubleshooting Figure 2-12 Measurement view Step 6. Power-on self test The power-on self test is executed once automatically before the measurement starts. While the power-on self test is in progress, "Power on test" is displayed at the left in the instrument status bar.
Power On Self Test Power-on self-test always takes place once the E5061B is turned on. When a failure is detected, a message "Power on self test failed" is displayed. The content of the power-on self-test is the same as a part of the diagnostic test program and includes the dc power supply voltage, source PLL synthesizer and level controller (ALC) tests.
Do not operate front panel keys, keyboard and mouse during the diagnostic test. Changing the instrument settings while the diagnostic test is in progress will cause incorrect test results. Step 1. Connect a mouse and an external keyboard to the E5061B’s rear panel connector. Step 2. Press key.
Step 11. To exit the diagnostic test, click "Exit" button. Program Overall Instrument Information The option and serial number for the E5061B, and test date can be entered automatically. Test Conditions The test conditions, “Temperature”, “Humidity” and person doing the test (“Tested by”) can be entered.
Troubleshooting Troubleshooting Using Diagnostic Test Figure 2-15 Test dialog box To execute the test, click “Execute” button. To return the E5061B Diagnostic test program dialog box (Main Menu), click “Back to Main Menu”. The test program will prompt you to connect cables or Short termination to the E5061B.
Page 35
Troubleshooting Troubleshooting Using Diagnostic Test test, replace the faulty board assembly as shown in Table 2-3. Table 2-3 Diagnostic tests failure troubleshooting information (Option 3L5) Test Test group Failured Probable faulty board assembly Test A1/A11 A9 LF Source Receiver Source Module Module Bias...
Page 36
Troubleshooting Troubleshooting Using Diagnostic Test Table 2-4 Diagnostic tests failure troubleshooting information (option 115/117/135/137/215/217/235/237) Test Test group Failured Test Probable faulty board assembly A11 Source A3/A4 Receiver Module Module DC-BUS Test A11 Source Module A3/A4 Receiver Module Internal Level Monitor Test A11 Source Module RF Output Level Range Test...
Function Specific Troubleshooting Function Specific Troubleshooting If the E5061B exhibits a failure symptom that is related to a specific function or control such as a front panel key control, display, data storage, remote control interface, printer interface, external trigger, external keyboard or mouse, isolate the trouble using the Function Specific Troubleshooting procedures described below.
Function Specific Troubleshooting To Check the Device Driver Make sure first whether the E5061B device drivers are installed properly or not by the following procedure, if a function of specific device in the E5061B doesn't work. Step 1. Exit from the E5061B measurement view in accordance with the procedure described in “How to exit from the E5061B Measurement View”...
Troubleshooting Function Specific Troubleshooting Figure 2-19 System Properties Window (Hardware) Click the icon with the right button and click Property to show the detail of the status. as shown in Figure 2-20. Figure 2-20 Opening Device Driver Property To Check the Front Panel Procedure Randomly press the front panel keys and rotate the knob to verify that they work normally.
Troubleshooting Function Specific Troubleshooting Step 1. Press key. Step 2. Click and, then, in the softkeys. Service Menu Test Menu Step 3. Click in the test menu. This opens "Front Panel Test" dialog box as shown in Front Panel Figure 2-21. Figure 2-21 Front Panel Test dialog box Step 4.
White and Black every 2 seconds and returns to the measurement view. If the color test screen doesn't appear correctly, perform step 4. Step 4. Connect an external VGA monitor to the VIDEO output port on the E5061B rear panel. •...
To Check the Video output Procedure Step 1. Connect an external VGA color monitor to the Video output port on the E5061B rear panel. Step 2. Turn the external monitor on. Step 3. Verify that the monitor screen view is the same as the display on the LCD. If the monitor screen view is abnormal, a failure seems in the CPU module.
This section describes the adjustment and troubleshooting procedures used when the E5061B fails the performance tests. If the performance of the instrument is critical for the test limits and seems to be adjustable, perform first the adjustment(s) related to the failed test.
Page 44
Troubleshooting Performance Test failure Troubleshooting Table 2-6 Recommended adjustment for performance test failure (Option 3L5) First failed test Recommended adjustment √ √ √ LF output level accuracy& flatness test (Opt 3L5) √ √ √ LF output level linearity test (Opt 3L5) √...
Troubleshooting Performance Test failure Troubleshooting Table 2-7 Recommended adjustment for performance test failure (option 115/117/135/137/215/217/235/237) First failed test Recommended adjustment √ Frequency accuracy test √ √ √ √ RF output level accuracy test RF output level linearity test √ Crosstalk & system dynamic range test √...
Page 46
Troubleshooting Performance Test failure Troubleshooting assembly and others are tight or loose. Also check for possible disconnection (impairment) of the cables and connectors. Table 2-8 Performance Test failure Troubleshooting information (Option 3L5) First failed test Probable faulty board assembly or parts A1/A11 A9 LF...
Page 47
Troubleshooting Performance Test failure Troubleshooting Table 2-9 Performance Test failure Troubleshooting information (Option 115/117/135/137/215/217/235/237) First failed test Probable faulty board assembly or parts A11 Source A3/A4 Module Receiver Module Frequency accuracy test RF output level accuracy test RF output level linearity test Crosstalk &...
Replaceable Parts Ordering Information Ordering Information To order a part in the replaceable parts lists, quote the Agilent Technologies part number, indicate the quantity required, and address the order to the nearest Agilent Technologies Sales Office. To order a part not listed in the replaceable parts lists, include the instrument model number, the description and function of the part, and the quantity of parts required.
Replaceable Parts Replaceable Parts List Replaceable Parts List The latest replaceable parts list is shown on the Agilent web site. See the http://www.agilent.com/find/ena_support/ Chapter 3...
Page 51
Replaceable Parts Replaceable Parts List Chapter 3...
Replacement Procedure Replacing an Assembly Replacing an Assembly The following steps show the sequence for replacing an assembly in a E5061B Network Analyzer. 1. Identify the faulty group. Refer to Chapter 2, “Troubleshooting.” 2. Order a replacement assembly. Refer to Chapter 3, “Replaceable Parts.”...
Replacement Procedure Required Tools Required Tools The following tools are required for repair of E5061B. Table 4-1 Required Tools Assembly torque screwdriver flat edge hex key box torque box torque open torque screwdriver driver wrench wrench 1.5 mm 5.5 mm 9/16 in.
Removal Procedure Refer to Figure 4-1 for this procedure. Step 1. Disconnect the power cable from the E5061B. Step 2. Remove the four bottom feet (item 1). Step 3. Remove the two TORX T20 screws (item 2) fastening the side strap handles.
Replacement Procedure Front Panel Replacement Front Panel Replacement Tools required • torque screwdriver T20 • torque screwdriver T15 (set to 14 in-lb) • flat edge screwdriver Removal Procedure Refer to Figure 4-2 for this procedure. Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54. Step 2.
Page 58
Replacement Procedure Front Panel Replacement Figure 4-2 Front Panel Removal Chapter 4...
Replacement Procedure Hard Disk Assembly Replacement (Option 019) Hard Disk Assembly Replacement (Option 019) Tools Required • torque screwdriver, T15, and T20 • torque screwdriver, T10 (set to 4.4 in-lb, and 9 in-lb) Removal Procedure Refer to Figure 4-3 for this procedure. Step 1.
Page 60
Replacement Procedure Hard Disk Assembly Replacement (Option 019) Figure 4-3 Hard Disk Assembly Removal Chapter 4...
Replacement Procedure Hard Disk Assembly Replacement (Option 020) Hard Disk Assembly Replacement (Option 020) Tools Required • torque screwdriver, T15, and T20 • torque screwdriver, T10 (set to 4.4 in-lb, and 9 in-lb) Removal Procedure Refer to Figure 4-3 for this procedure. Step 1.
Page 62
Replacement Procedure Hard Disk Assembly Replacement (Option 020) Figure 4-4 Hard Disk Assembly Removal (option 020) Chapter 4...
Replacement Procedure A50 CPU Module Replacement A50 CPU Module Replacement Tools Required • torque screwdriver, T15, and T20 • torque screwdriver, T10 (set to 9 in-lb) Removal Procedure Refer to Figure 4-6 for this procedure. Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54. Step 2.
Page 66
Replacement Procedure A50 CPU Module Replacement Figure 4-6 A50 CPU Module Removal Chapter 4...
Replacement Procedure A60 CPU Module Replacement A60 CPU Module Replacement Tools Required • torque screwdriver, T15, and T20 • torque screwdriver, T10 (set to 9 in-lb) Removal Procedure Refer to Figure 4-7 for this procedure. Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54. Step 2.
Page 68
Replacement Procedure A60 CPU Module Replacement Figure 4-7 A60 CPU Module Removal Chapter 4...
Replacement Procedure Fan Assembly Replacement Fan Assembly Replacement Tools Required • torque screwdriver, T15 and T20 • torque screwdriver, T10 (set to 9 in-lb) • box torque wrench, 9/16 inch (set to 30.4 in-lb) Removal Procedure Refer to Figure 4-17 for this procedure. Step 1.
Page 88
Replacement Procedure Fan Assembly Replacement Figure 4-17 Fan Assembly Removal Chapter 4...
Replacement Procedure Front Frame Replacement Front Frame Replacement Tools required • torque screwdriver T20 (set to 21 in-lb) • torque screwdriver T15 (set to 14 in-lb) • flat edge screwdriver Removal Procedure Refer to Figure 4-24 for this procedure. Step 1. Remove the Outer Cover as described in “Outer Cover Replacement” on page 54. Step 2.
Page 100
Replacement Procedure Front Frame Replacement Figure 4-24 Front Frame Removal Chapter 4...
Page 104
Post-Repair Procedures This chapter lists the procedures required to verify the E5061B operation after an assembly is replaced with a new one.
Page 105
Post-Repair Procedures Post-Repair Procedures Post-Repair Procedures Table 5-1 Post Repair Procedures lists the required procedures that must be performed after the replacement of an assembly. These are the recommended minimum procedures to ensure that the replacement is successfully completed. Table 5-1 Post-Repair Procedures (Option 3L5) Replaced Required Adjustments...
Page 106
Post-Repair Procedures Post-Repair Procedures Table 5-1 Post-Repair Procedures (Option 3L5) Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part A2 Receiver Perform the following required “To Execute the Diagnostic Test” on page 31 Module adjustments using “A2 Receiver Module” in Spot Adjustment of the program.
Page 107
Post-Repair Procedures Post-Repair Procedures Table 5-1 Post-Repair Procedures (Option 3L5) Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part A50 CPU No adjustment needed Inspect the Booting Process module DSP Card Perform the following required Inspect the Booting Process adjustments using “A51 DSP board”...
Page 108
Post-Repair Procedures Post-Repair Procedures Table 5-2 Post-Repair Procedures Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part A11 Source Perform the following required “To Execute the Diagnostic Test” on page 31 Module adjustments using “A11 Source Module” Frequency Accuracy Test in Spot Adjustment of the program.
Page 109
Post-Repair Procedures Post-Repair Procedures Table 5-2 Post-Repair Procedures Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part Perform “HDD” (Hard Disk Drive) in Spot Inspect the Booting Process Adjustment of the program. Calibration of the Touch Screen (for opt. 016) Power Supply No adjustment needed...
This appendix contains the information required to adapt this manual to versions or configurations of the E5061B manufactured earlier than the current printing date of this manual. The information in this manual applies directly to E5061B units with the serial number that is printed on the title page of this manual.
Manual Changes Manual Changes Manual Changes To adapt this manual to your E5061B, refer to Table A-1 and Table A-2. Table A-1 Manual Changes by Serial Number Serial Prefix or Number Make Manual Changes Table A-2 Manual Changes by Firmware Version...
System Recovery System Recovery By executing system recovery, you can return the system of the E5061B (the Windows operating system and the firmware) to the factory state (at the time of purchase The procedure of system recovery is described in Installation Guide. Please refer to Agilent website below.
Firmware Update Required Equipment • E5061B (with keyboard and mouse). • USB mass storage. How to update the E5061B firmware The firmware update procedure is described in E5061B WebHelp. Please refer to Agilent website as below. http://ena.tm.agilent.com/e5061b/manuals/webhelp/eng/index.htm#product_information/ maintenance/updating_firmware.htm Appendix C...
47 to 63 Hz Maximum power consumption 300 VA *1. Switched automatically by the E5061B in conformity to the voltage. Power Cable In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame.
Page 118
Power Requirement Preparation for Power Supply Figure D-1 Power cable options Appendix D...
If the power supply is cut off directly by disconnecting the power cable plug or by disconnecting the power supply to the AC outlet, the shutdown process will not be carried out, and there is a risk of damage to the software or hardware of the E5061B. Appendix D...
This appendix explains what these messages mean. They are listed in alphabetical order. Messages showing the status of the E5061B are displayed in the lower-left area of the E5061B LCD screen. These messages include error messages that occur during the...
Page 121
Messages execution of GPIB commands and others that indicate the internal status of the equipment. Error messages are indicated following the character string “[Err]” and can be read out by a GPIB command. Other kinds of messages are indicated without the “[Err]” character string and cannot be read out by a GPIB command.
GPIB devices in IEEE488.2 Error messages are decribed in WebHelp. Please kindly refer to Agilent Website below. http://ena.tm.agilent.com/e5061b/manuals/webhelp/eng/index.htm#product_information/e rror_messages/error_messages.htm Warning Message A warning message is displayed in the instrument message/Warning area in the lower left part of the display against a gray background.
Need help?
Do you have a question about the E5061B and is the answer not in the manual?
Questions and answers