Sweep Function

The list (segment) sweep function enables you to perform measurements
corresponding to two or more sweep conditions in one sweep operation
and is supported both on the 8753ES and E5071C. While the 8753ES
allows only up to 30 segments per table to be defined, the E5071C allows
up to 201 segments to be defined. Furthermore, while two or more
commands are needed to create a table using GPIB (SCPI) commands on
the 8753ES, only one command does the same function on the E5071C.
The 8753ES supports an IF bandwidth up to 6 kHz, but the E5071C
supports an IF bandwidth up to 500 kHz, enabling faster sweep operations
than with the 8753ES.
While the 8753ES supports automatically selecting the output power range,
the E5071C supports only manual selection.
The 8753ES allows you to measure frequency conversion device such as a
mixer; since it supports the frequency offset sweep. The E5071C also
supports this function when the option 008 is used.
Calibration
The types of calibration kits supported by the 8753ES and E5071C are
shown in the following table.
Supported Calibration Kits
902
15 dBm to 8
dBm
(Options 014
and 075)
- 25 dBm to 0
dBm
- 35 dBm to -
10 dBm
- 45 dBm to -
20 dBm
- 55 dBm to -
30 dBm
- 65 dBm to -
40 dBm
- 75 dBm to -
50 dBm
- 85 dBm to -
60 dBm