Agilent Technologies E5071C Manual page 543

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
-400
Query error
-410
Query
INTERRUPTED
-420
Query
UNTERMINATED
-440
Query
UNTERMINATED
after indefinite
response
-350
Queue overflow
R
No.
Message
105
Recall failed
S
No.
Message
770
A comprehensive query error has
occurred showing that this instrument
cannot detect a more detailed error.
This code simply shows that a query
error defined in 11.5.1.1.7 and 6.3,
IEEE488.2 has occurred.
The state that generates a
"INTERRUPTED" Query error (see
6.3.2.3, IEEE488.1). This error occurs
when data bytes (DAB) or GET are
received before the transmission of the
response after a query has not been
completed, for example.
The state that generates an
"UNTERMINATED" Query error (see
6.3.2, IEEE488.2). This error occurs
when this instrument is designated as
the talker and an incomplete program
message is received, for example.
After a query asking for an indefinite
response has been run, another query
is received in the same program
message (See 6.5.7.5.7, IEEE488.2).
The queue contains a specific code in
place of the code which caused this
error. The code indicates that the error
occurred because of no space available
in the queue, but the error is not
recorded.
Description
This error occurs when reading an
instrument status file (State01.sta,
etc.) (MMEM:LOAD:STAT command)
fails.